Trial Outcomes & Findings for Noninvasive Hemoglobin Testing, Prospective Blood Donors (NCT NCT04450381)

NCT ID: NCT04450381

Last Updated: 2021-07-27

Results Overview

Sensitivity of the the Rad-67 will be determined for its use in a blood donation center

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

131 participants

Primary outcome timeframe

Up to one hour per subject

Results posted on

2021-07-27

Participant Flow

Participant milestones

Participant milestones
Measure
Test Subjects
All subjects are enrolled and receive the Pulse oximeter \& sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Overall Study
STARTED
131
Overall Study
COMPLETED
125
Overall Study
NOT COMPLETED
6

Reasons for withdrawal

Reasons for withdrawal
Measure
Test Subjects
All subjects are enrolled and receive the Pulse oximeter \& sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Overall Study
Protocol Violation
4
Overall Study
Wrong device setting
2

Baseline Characteristics

Noninvasive Hemoglobin Testing, Prospective Blood Donors

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Test Subjects
n=125 Participants
All subjects are enrolled and receive Rad-67 Pulse oximeter \& DCI Mini sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI Mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Age, Categorical
<=18 years
1 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
106 Participants
n=5 Participants
Age, Categorical
>=65 years
18 Participants
n=5 Participants
Sex: Female, Male
Female
76 Participants
n=5 Participants
Sex: Female, Male
Male
49 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · American Indian or Alaska Native
1 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Asian
16 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Native Hawaiian or Other Pacific Islander
2 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Black or African American
2 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · White
76 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · More Than One Race
6 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Unknown or Not Reported
3 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Other: American
1 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Other: Assyrian
1 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Other: Mexican
16 Participants
n=5 Participants
Race/Ethnicity, Customized
Race · Other: Native American
1 Participants
n=5 Participants
Region of Enrollment
United States
125 participants
n=5 Participants

PRIMARY outcome

Timeframe: Up to one hour per subject

Population: 2 subjects were excluded from analysis due to wrong device setting

Sensitivity of the the Rad-67 will be determined for its use in a blood donation center

Outcome measures

Outcome measures
Measure
Test Subjects
n=123 Participants
All subjects are enrolled and receive the Pulse oximeter \& sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Rad-67 Sensitivity
44 percentage of sensitivity

PRIMARY outcome

Timeframe: Up to one hour per subject

Population: 2 subjects were excluded from analysis due to wrong device setting

Specificity of the the Rad-67 will be determined for its use in a blood donation center

Outcome measures

Outcome measures
Measure
Test Subjects
n=123 Participants
All subjects are enrolled and receive the Pulse oximeter \& sensor for measurement of hemoglobin. Rad-67 Pulse oximeter \& DCI mini sensor: Noninvasive pulse oximeter that measures hemoglobin
Rad-67 Specificity
86.73 percentage of specificity

Adverse Events

Test Subjects

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Vikram Ramakanth

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: GT60