Trial Outcomes & Findings for Reduce Radiation Exposure in Fluoroscopic Interventions Evaluation (NCT NCT04404257)
NCT ID: NCT04404257
Last Updated: 2023-01-11
Results Overview
Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
COMPLETED
NA
170 participants
During each cardiac catheterization, an average of 90 minutes
2023-01-11
Participant Flow
Participant milestones
| Measure |
ControlRad System
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
Without ControlRad System
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
|---|---|---|
|
Overall Study
STARTED
|
85
|
85
|
|
Overall Study
COMPLETED
|
85
|
85
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
ControlRad System
n=85 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
Without ControlRad System
n=85 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
Total
n=170 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
70.3 years
STANDARD_DEVIATION 4.6 • n=85 Participants
|
57.1 years
STANDARD_DEVIATION 9.9 • n=85 Participants
|
63.7 years
STANDARD_DEVIATION 1.2 • n=170 Participants
|
|
Sex: Female, Male
Female
|
35 Participants
n=85 Participants
|
36 Participants
n=85 Participants
|
71 Participants
n=170 Participants
|
|
Sex: Female, Male
Male
|
50 Participants
n=85 Participants
|
49 Participants
n=85 Participants
|
99 Participants
n=170 Participants
|
|
Race and Ethnicity Not Collected
|
—
|
—
|
0 Participants
Race and Ethnicity were not collected from any participant.
|
|
Region of Enrollment
United States
|
85 participants
n=85 Participants
|
85 participants
n=85 Participants
|
170 participants
n=170 Participants
|
PRIMARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad
Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Primary Operator (RaySafe), Cardiac Catheterization
|
39 microsievert
Interval 19.0 to 61.0
|
18 microsievert
Interval 7.0 to 35.0
|
PRIMARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Primary Operator (Luxel Dosimeter), Cardiac Catheterization
|
41 microsievert
|
18 microsievert
|
PRIMARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure. A single Luxel dosimeter was and a cumulative dose was recorded.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Primary Operator, EP Procedure (Luxel)
|
58.1 microsieverts
|
21.7 microsieverts
|
PRIMARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutes durationPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group.
Median effective radiation dose in microsievert (μSv) to the primary operator at the end of each EP implant procedure using a RaySafe dosimeter. This is measured after each case.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Primary Operator, EP Procedure (RaySafe)
|
61 microsieverts
Interval 25.5 to 149.0
|
25 microsieverts
Interval 15.0 to 69.5
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad
Median effective radiation dose (μSv) to the secondary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Secondary Operator, Cardiac Catheterization (RaySafe)
|
21 microsievert
Interval 13.0 to 49.0
|
12 microsievert
Interval 6.0 to 22.0
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean effective radiation dose in microsievert (μSv) to the secondary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Secondary Operator (Luxel Dosimeter), Cardiac Catheterization
|
26 microsievert
|
13 microsievert
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad
Median effective radiation dose in microsievert (μSv) to the Nurse 0r Anesthetist at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (RaySafe)
|
NA microsievert
Dosage data unreliable as it is below lower limits of detection
|
NA microsievert
Dosage data unreliable as it is below lower limits of detection
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad
Mean effective radiation dose in microsievert (μSv) to the Nurse or Anesthetist at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Nurse or Anesthetist, Cardiac Catheterization (Luxel)
|
NA microsievert
Dosage data unreliable as it is below lower limits of detection
|
NA microsievert
Dosage data unreliable as it is below lower limits of detection
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad
Median radiation dose-area product (DAP) to the patient at the end of each cardiac catheterization case using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated micro-gray μGy\*m\^2.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Patient, Cardiac Catheterization
|
3487 μGy*m^2
Interval 2257.0 to 5977.0
|
2276 μGy*m^2
Interval 1179.0 to 3321.0
|
SECONDARY outcome
Timeframe: During each cardiac catheterization, an average of 90 minutesPopulation: This outcome measured only the cardiac catheterization procedures, 42 participants in the ControlRad group and 38 in the group without ControlRad. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean radiation dose microsieverts (μSv) at the end of each cardiac catheterization case at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=38 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=42 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose in Cath Lab Room #5, Cardiac Catheterization
|
25 microsieverts
|
15 microsieverts
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group.
Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Secondary Operator, EP Procedure (RaySafe)
|
5 microsieverts
Interval 3.0 to 16.0
|
3 microsieverts
Interval 1.5 to 5.0
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean effective radiation dose (μSv) to the secondary operator at the at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Secondary Operator, EP Procedure (Luxel)
|
9 microsieverts
|
1.6 microsieverts
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group.
Median effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a RaySafe dosimeter to assess the radiation dose in each case.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Nurse or Anesthetist, EP Procedure (RaySafe)
|
2 microsieverts
Interval 0.0 to 10.0
|
0 microsieverts
Interval 0.0 to 2.0
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean effective radiation dose in microsievert (μSv) to the primary operator at the thyroid. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Nurse or Anesthetist, EP Procedure (Luxel)
|
3.8 microsieverts
|
0 microsieverts
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 30-45 minutes of active radiationPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group.
Mean radiation dose-area product (DAP) to the patient at the end of each EP implant procedure using data from a fluoroscopy machine. DAP is defined as the absorbed dose multiplied by the area irradiated (μGy\*m\^2)
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose to Patient, EP Procedure
|
72.5 μGy*m^2
Interval 22.7 to 131.5
|
43.7 μGy*m^2
Interval 26.0 to 95.9
|
SECONDARY outcome
Timeframe: During each EP implant procedure, an average of 90 minutesPopulation: This outcome measured only the EP procedures, 43 in the ControlRad group and 47 in the Without ControlRad group. A single badge was used for all cases, the result is provided as a number per arm. The number is cumulative of the arm with no measure of dispersion.
Mean radiation dose (μSv) at the end of each EP implant procedure at a standard location on the wall of the catheterization laboratory. This was measured using a single Luxel dosimeter over all cases to assess the cumulative dose in the study arm.
Outcome measures
| Measure |
Without ControlRad System
n=47 Participants
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
ControlRad System
n=43 Participants
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
|---|---|---|
|
Radiation Dose in Cath Lab Room #5, EP Procedure
|
0 microsieverts
|
0 microsieverts
|
Adverse Events
ControlRad System
Without ControlRad System
Serious adverse events
| Measure |
ControlRad System
n=85 participants at risk
Participant will undergo Cardiac catheterization or electrophysiology implant procedures with the ControlRad system installed in Cath lab room 5
ControlRad System: Cardiac Catheterization or electrophysiology implant procedures with Control Rad system device installed in the cath lab room #5.
|
Without ControlRad System
n=85 participants at risk
Participant will undergo the cardiac catheterization or electrophysiology implant procedures per standard of care. Meaning, without the ControlRad system installed in Cath lab room 5.
Without ControlRad System: Cardiac Catheterization or electrophysiology implant procedures in cath lab room 5 without the ControlRad system device.
|
|---|---|---|
|
Injury, poisoning and procedural complications
Cardiac Perforation
|
0.00%
0/85 • During each cardiac catheterization/EP procedure, an average of 90 minutes
|
1.2%
1/85 • Number of events 1 • During each cardiac catheterization/EP procedure, an average of 90 minutes
|
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place