Trial Outcomes & Findings for The Role of Pre-deployment Retraction in Decreasing Biopsy Clip Migration During Stereotactic Breast Biopsies (NCT NCT04398537)

NCT ID: NCT04398537

Last Updated: 2023-12-20

Results Overview

This number of participants will be counted if their clip migrated more than 10mm from the biopsy site.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

245 participants

Primary outcome timeframe

baseline through 1 hour (post biopsy mammogram/procedure)

Results posted on

2023-12-20

Participant Flow

Participant milestones

Participant milestones
Measure
5mm Retraction of Clip Deployment Apparatus
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Overall Study
STARTED
112
133
Overall Study
COMPLETED
102
118
Overall Study
NOT COMPLETED
10
15

Reasons for withdrawal

Reasons for withdrawal
Measure
5mm Retraction of Clip Deployment Apparatus
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Overall Study
Physician Decision
10
15

Baseline Characteristics

The Role of Pre-deployment Retraction in Decreasing Biopsy Clip Migration During Stereotactic Breast Biopsies

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
5mm Retraction of Clip Deployment Apparatus
n=102 Participants
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
n=117 Participants
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Total
n=219 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
60 Participants
n=5 Participants
71 Participants
n=7 Participants
131 Participants
n=5 Participants
Age, Categorical
>=65 years
42 Participants
n=5 Participants
46 Participants
n=7 Participants
88 Participants
n=5 Participants
Sex: Female, Male
Female
102 Participants
n=5 Participants
117 Participants
n=7 Participants
219 Participants
n=5 Participants
Sex: Female, Male
Male
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Race/Ethnicity, Customized
White Non-Hispanic
49 participants
n=5 Participants
59 participants
n=7 Participants
108 participants
n=5 Participants
Race/Ethnicity, Customized
White Hispanic
0 participants
n=5 Participants
1 participants
n=7 Participants
1 participants
n=5 Participants
Race/Ethnicity, Customized
White Ethnicity Not reported
3 participants
n=5 Participants
2 participants
n=7 Participants
5 participants
n=5 Participants
Race/Ethnicity, Customized
White Ethnicity Unknown
0 participants
n=5 Participants
2 participants
n=7 Participants
2 participants
n=5 Participants
Race/Ethnicity, Customized
Black Non-Hispanic
33 participants
n=5 Participants
33 participants
n=7 Participants
66 participants
n=5 Participants
Race/Ethnicity, Customized
Black Ethnicity Not Reported
1 participants
n=5 Participants
1 participants
n=7 Participants
2 participants
n=5 Participants
Race/Ethnicity, Customized
Asian Non-Hispanic
7 participants
n=5 Participants
10 participants
n=7 Participants
17 participants
n=5 Participants
Race/Ethnicity, Customized
Asian Ethnicity Unknown
1 participants
n=5 Participants
0 participants
n=7 Participants
1 participants
n=5 Participants
Race/Ethnicity, Customized
Hispanic
1 participants
n=5 Participants
0 participants
n=7 Participants
1 participants
n=5 Participants
Race/Ethnicity, Customized
Hispanic Non-Hispanic
0 participants
n=5 Participants
1 participants
n=7 Participants
1 participants
n=5 Participants
Race/Ethnicity, Customized
Decline Race Non-Hispanic
4 participants
n=5 Participants
6 participants
n=7 Participants
10 participants
n=5 Participants
Race/Ethnicity, Customized
Decline Race Ethnicity Not Reported
3 participants
n=5 Participants
2 participants
n=7 Participants
5 participants
n=5 Participants

PRIMARY outcome

Timeframe: baseline through 1 hour (post biopsy mammogram/procedure)

This number of participants will be counted if their clip migrated more than 10mm from the biopsy site.

Outcome measures

Outcome measures
Measure
5mm Retraction of Clip Deployment Apparatus
n=104 Participants
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
n=119 Participants
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Number of Breast Biopsy Patients Whose Clip Migrated Greater Than 10mm From the Biopsy Site.
46 Participants
44 Participants

PRIMARY outcome

Timeframe: baseline through 1 hour (post biopsy mammogram/procedure)

Outcome measures

Outcome measures
Measure
5mm Retraction of Clip Deployment Apparatus
n=104 Participants
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
n=119 Participants
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Percentage of Breast Biopsy Patients Whose Clip Migrated Greater Than 10mm From the Biopsy Site.
46 Participants
44 Participants

PRIMARY outcome

Timeframe: baseline through 1 hour (post biopsy mammogram/procedure)

Population: This was only measured on the 5mm retraction group

Outcome measures

Outcome measures
Measure
5mm Retraction of Clip Deployment Apparatus
n=46 Participants
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Average Distance of Clip Migration for the Arm That Received 5mm Retraction.
12.1 mm
Standard Deviation 14.9

PRIMARY outcome

Timeframe: baseline through 1 hour (post biopsy mammogram/procedure)

Population: This was only measured on the control group

Outcome measures

Outcome measures
Measure
5mm Retraction of Clip Deployment Apparatus
The participants in this group will have clip placement 5mm in front of the biopsy site site. 5mm retraction of clip deployment apparatus.: Advancing the clip to the biopsy site and retracting 5 mm and then deploying. This is one method of standard of care and involves depositing the clip after retraction of the deployment apparatus 5 mm.
no Retraction of Clip Deployment Apparatus
n=44 Participants
These participants will the clip delivered at the biopsy site. No retraction of clip deployment apparatus.: This is one method of standard of care and the clip will be deployed at the biopsy site.
Average Distance of Clip Migration for the Arm That Did Not Receive Clip Retraction.
9.8 mm
Standard Deviation 14.9

Adverse Events

5mm Retraction of Clip Deployment Apparatus

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

no Retraction of Clip Deployment Apparatus

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Stefanie Woodard

University of Alabama at Birmingham

Phone: 205-996-4132

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place