Trial Outcomes & Findings for Evaluation of a Portable Oxygen Concentrator During Ambulation for Patients Who Require Supplemental Oxygen (POC-LEAN) (NCT NCT04107493)

NCT ID: NCT04107493

Last Updated: 2021-05-26

Results Overview

Each subject's SpO2 will be measured when each subject takes six minutes walk test

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

33 participants

Primary outcome timeframe

Mean Sp02, measured during 6 minute walk test

Results posted on

2021-05-26

Participant Flow

Participants acted as their own controls. Participants were randomised as to the order they used the devices during the 6-minute walking test. There was a one hour wash out period between tests

Participant milestones

Participant milestones
Measure
Portable Oxygen Cylinder, Then Mobi
Participants acted as their own control in this study. Participants used two portable oxygen cylinders during a 6 minute walk test. One was a standard POC, and one was the trial device, the Mobi POC Portable Oxygen Cylinder: A POC is a medical device that is indicated for patients who require supplemental oxygen, including COPD patients
Mobi, Then Portable Oxygen Cylinder
Participants acted as their own control in this study. Participants used two portable oxygen cylinders during a 6 minute walk test. One was a standard POC, and one was the trial device, the Mobi POC Portable Oxygen Cylinder: A POC is a medical device that is indicated for patients who require supplemental oxygen, including COPD patients
First Intervention (1 Hour)
STARTED
16
17
First Intervention (1 Hour)
COMPLETED
16
17
First Intervention (1 Hour)
NOT COMPLETED
0
0
Washout (One Hour)
STARTED
16
17
Washout (One Hour)
COMPLETED
16
17
Washout (One Hour)
NOT COMPLETED
0
0
Second Intervention
STARTED
16
17
Second Intervention
COMPLETED
16
17
Second Intervention
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Evaluation of a Portable Oxygen Concentrator During Ambulation for Patients Who Require Supplemental Oxygen (POC-LEAN)

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
All Study Participants
n=33 Participants
Participants acted as their own control during this study. Participants used a portable oxygen cylinder whilst doing a 6 minute walking test. After a wash out period the participants completed the 6 minute walking test with the second portable oxygen cylinder. The order of the oxygen cylinders (mobi or standard) was randomised
Age, Continuous
69.86 years
STANDARD_DEVIATION 8.82 • n=5 Participants
Sex: Female, Male
Female
21 Participants
n=5 Participants
Sex: Female, Male
Male
12 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
3 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
30 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
n=5 Participants
Race (NIH/OMB)
Asian
0 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
1 Participants
n=5 Participants
Race (NIH/OMB)
White
29 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
2 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Mean Sp02, measured during 6 minute walk test

Each subject's SpO2 will be measured when each subject takes six minutes walk test

Outcome measures

Outcome measures
Measure
Portable Oxygen Cylinder
n=33 Participants
Continuous flow oxygen cylinders will be used as a comparison. Portable Oxygen Cylinder: A POC is a medical device that is indicated for patients who require supplemental oxygen, including COPD patients
Mobi™ Portable Oxygen Concentrator
n=33 Participants
Mobi™ is indicated for patients who require supplemental oxygen, including COPD patients. It provides supplemental, high oxygen concentration to these patients. It is available via prescription only and may be used in the home, institution, and hospital settings, as well as travel environments Mobi: ResMed's variant of POC
Mean SpO2 (Oxygen Saturation) During Six Minute Walk Tests (6MWTs)
91.2 percentage of oxygen saturation
Standard Error 0.7
90.0 percentage of oxygen saturation
Standard Error 0.7

Adverse Events

Standard Oxygen Cylinder

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Mobi

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

maureen crocker

ResMed

Phone: +1 8588366653

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place