Trial Outcomes & Findings for Study of a Hand-held Device Plus Mobile App Versus Standard Eye Care Devices in Measuring Refractive Error of the Eye (NCT NCT03929588)
NCT ID: NCT03929588
Last Updated: 2022-03-24
Results Overview
Best-corrected visual acuity (BVCA) utilizing refraction results using the handheld device with supporting application will be compared to standard manual refraction methods (using a phoropter) for the age stratum 45 through 65 years.
COMPLETED
NA
230 participants
Through study completion, an average of 5 months
2022-03-24
Participant Flow
Participant milestones
| Measure |
All Participants
BCVA with handheld device with app.
Phoropter: Manual refraction and ETDRS chart
Autorefractor: Automated refraction
|
|---|---|
|
Overall Study
STARTED
|
230
|
|
Overall Study
COMPLETED
|
143
|
|
Overall Study
NOT COMPLETED
|
87
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Refraction With a Hand-held Device Supported by Mobile App.
n=230 Participants
BCVA with handheld device with app.
Phoropter: Manual refraction and ETDRS chart
Autorefractor: Automated refraction
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=230 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
230 Participants
n=230 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=230 Participants
|
|
Sex: Female, Male
Female
|
115 Participants
n=230 Participants
|
|
Sex: Female, Male
Male
|
115 Participants
n=230 Participants
|
PRIMARY outcome
Timeframe: Through study completion, an average of 5 monthsBest-corrected visual acuity (BVCA) utilizing refraction results using the handheld device with supporting application will be compared to standard manual refraction methods (using a phoropter) for the age stratum 45 through 65 years.
Outcome measures
| Measure |
All Participants
n=143 Participants
BCVA with handheld device with app.
Phoropter: Manual refraction and ETDRS chart
Autorefractor: Automated refraction
|
|---|---|
|
BVCA Utilizing Refraction Results From Handheld Device Compared to Manual Refraction Methods for the Age Stratum 45 Through 65 Years
BCVA handheld device
|
0.06 logMAR
Interval 0.04 to 0.08
|
|
BVCA Utilizing Refraction Results From Handheld Device Compared to Manual Refraction Methods for the Age Stratum 45 Through 65 Years
BCVA manual refraction
|
0.02 logMAR
Interval 0.0 to 0.04
|
|
BVCA Utilizing Refraction Results From Handheld Device Compared to Manual Refraction Methods for the Age Stratum 45 Through 65 Years
BCVA automated refraction
|
0.03 logMAR
Interval 0.01 to 0.05
|
Adverse Events
Refraction With a Hand-held Device Supported by Mobile App.
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place