Trial Outcomes & Findings for Focused Cardiac Ultrasound (NCT NCT03887286)
NCT ID: NCT03887286
Last Updated: 2020-06-16
Results Overview
Comparison of diagnostic ability of portable hand-held cardiac ultrasound with standard transthoracic echocardiography
Recruitment status
COMPLETED
Study phase
NA
Target enrollment
47 participants
Primary outcome timeframe
Day 1
Results posted on
2020-06-16
Participant Flow
Participant milestones
| Measure |
Treatment Group
All participants to receive standard transthoracic echocardiogram and hand held echocardiogram
Philips EPIQ 7 ultrasound machine: Standard transthoracic echocardiography
Philips Lumify Broadband sector array transducer: Portable hand-held ultrasound device
|
|---|---|
|
Overall Study
STARTED
|
47
|
|
Overall Study
COMPLETED
|
47
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Treatment Group
n=47 Participants
All participants to receive standard transthoracic echocardiogram and hand held echocardiogram
Philips EPIQ 7 ultrasound machine: Standard transthoracic echocardiography
Philips Lumify Broadband sector array transducer: Portable hand-held ultrasound device
|
|---|---|
|
Age, Continuous
|
7.8 years
STANDARD_DEVIATION 8.6 • n=47 Participants
|
|
Sex: Female, Male
Female
|
19 Participants
n=47 Participants
|
|
Sex: Female, Male
Male
|
28 Participants
n=47 Participants
|
|
weight
|
27.24 kg
STANDARD_DEVIATION 23.41 • n=47 Participants
|
PRIMARY outcome
Timeframe: Day 1Comparison of diagnostic ability of portable hand-held cardiac ultrasound with standard transthoracic echocardiography
Outcome measures
| Measure |
Standard Transthoracic Echocardiogram Intervention
n=47 Participants
The transthoracic echocardiogram was performed and read via Intersocietal Commission Accreditation Echocardiography Laboratories (ICAEL) standard
|
Hand Held Echocardiography
n=47 Participants
The hand held echocardiogram was performed and read via Intersocietal Commission Accreditation Echocardiography Laboratories (ICAEL) standard
|
|---|---|---|
|
Number of Patients Detected With Ventricular Dysfunction, Pericardial Effusion, Valvular Disorders, Septal Disorders, and Coronary Anomalies.
minor cardiac abnormality
|
9 Participants
|
4 Participants
|
|
Number of Patients Detected With Ventricular Dysfunction, Pericardial Effusion, Valvular Disorders, Septal Disorders, and Coronary Anomalies.
normal anatomy
|
13 Participants
|
13 Participants
|
|
Number of Patients Detected With Ventricular Dysfunction, Pericardial Effusion, Valvular Disorders, Septal Disorders, and Coronary Anomalies.
major cardiac abnormality
|
25 Participants
|
23 Participants
|
Adverse Events
Treatment Group
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place