Trial Outcomes & Findings for Pivot Breath Sensor Performance Study (NCT NCT03870022)

NCT ID: NCT03870022

Last Updated: 2024-07-01

Results Overview

Assess user performance of the study device at a single visit. Participants will be observed if they can set up the Pivot Breath Sensor, and whether they can use the device. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

70 participants

Primary outcome timeframe

Each study session lasted up to 60 minutes per participant. Use performance was assessed at approximately 10-15 minutes into the study session.

Results posted on

2024-07-01

Participant Flow

Participant milestones

Participant milestones
Measure
18-80 Year Old Smokers
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
Overall Study
STARTED
70
Overall Study
COMPLETED
70
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Pivot Breath Sensor Performance Study

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
Age, Customized
Age 18-49
34 Participants
n=93 Participants
Age, Customized
Age 50-80
36 Participants
n=93 Participants
Sex: Female, Male
Female
30 Participants
n=93 Participants
Sex: Female, Male
Male
40 Participants
n=93 Participants
Race/Ethnicity, Customized
White
44 Participants
n=93 Participants
Race/Ethnicity, Customized
Hispanic, Latino, or Spanish Origin
6 Participants
n=93 Participants
Race/Ethnicity, Customized
Black or African American
10 Participants
n=93 Participants
Race/Ethnicity, Customized
Asian
8 Participants
n=93 Participants
Race/Ethnicity, Customized
American Indian or Alaska Native
1 Participants
n=93 Participants
Race/Ethnicity, Customized
Hawaiian or other Pacific Islander
1 Participants
n=93 Participants
Region of Enrollment
United States
70 participants
n=93 Participants
years smoking
29.8 years
STANDARD_DEVIATION 14.8 • n=93 Participants

PRIMARY outcome

Timeframe: Each study session lasted up to 60 minutes per participant. Use performance was assessed at approximately 10-15 minutes into the study session.

Assess user performance of the study device at a single visit. Participants will be observed if they can set up the Pivot Breath Sensor, and whether they can use the device. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance.

Outcome measures

Outcome measures
Measure
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
User Performance of Pivot Breath Sensor
Self-training · Unsuccessful
2 Participants
User Performance of Pivot Breath Sensor
Self-training · Successful
68 Participants
User Performance of Pivot Breath Sensor
Performance breath sample use and/or with use of reference materials · Successful
70 Participants
User Performance of Pivot Breath Sensor
Performance breath sample use and/or with use of reference materials · Unsuccessful
0 Participants

PRIMARY outcome

Timeframe: Each study session lasted up to 60 minutes per participant.

Compare study device performance with a similar device at a single visit. Participants will then be asked to use a similar CO breath sensor device. The correlation between the measured CO levels (in parts per million, or ppm) of the study CO breath sensor (Carrot Sense) and the similar breath sensor will be assessed.

Outcome measures

Outcome measures
Measure
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
Compare Performance of Pivot Breath Sensor With a Similar Device.
0.9710 Correlation coefficient (R-value)

SECONDARY outcome

Timeframe: Each study session lasted up to 60 minutes per participant. The user documentation assessment with observer ratings was done between approximately 10 and 35 minutes into the study session.

Participant understanding of user documentation will be assessed at single visit through interpretation of three different CO test results: green, orange, and red. User documentation used for the interpretation of each result includes: packaging information, quick start guide and user manual. Observer ratings include 'Successful' and 'Unsuccessful'. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance. Participant will be asked to assess the user documentation throughout the study session and will take approximately 10 minutes.

Outcome measures

Outcome measures
Measure
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
User Documentation Assessment With Observer Ratings
Interpretation of Orange CO category after using reference materials · Successful
57 Participants
User Documentation Assessment With Observer Ratings
Interpretation of Green CO category after using reference materials · Successful
69 Participants
User Documentation Assessment With Observer Ratings
Interpretation of Green CO category after using reference materials · Unsuccessful
1 Participants
User Documentation Assessment With Observer Ratings
Interpretation of Orange CO category after using reference materials · Unsuccessful
13 Participants
User Documentation Assessment With Observer Ratings
Interpretation of Red CO category after using reference materials · Successful
70 Participants
User Documentation Assessment With Observer Ratings
Interpretation of Red CO category after using reference materials · Unsuccessful
0 Participants

SECONDARY outcome

Timeframe: Each study session lasted up to 60 minutes per participant. Use performance was assessed with rating scales at approximately 35 minutes into the study session.

Participant will be asked to provide ratings on the following: * How would you rate the ease or difficulty using the device? * How clear or unclear was the Quick Start Guide? Why? * How clear or unclear was the Packaging? Why? * How clear or unclear was the User Manual on the iPad? Why? * Overall, how clear or unclear were the provided reference materials? Why? * How easy or difficult was it to understand and interpret the test results? Why? The participant will use the following rating scale: 5 - Very Easy 4 - Easy 3 - Neither Easy nor Difficult 2 - Difficult 1 - Very Difficult Not Applicable Participant will provide the ratings prior to the conclusion of the study session. Participant assessment with questionnaire rating scales will take approximately 1-2 minutes each.

Outcome measures

Outcome measures
Measure
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales. Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales. Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Neither Easy nor Difficult
3 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Easy
16 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Difficult
2 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Very Difficult
1 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Very Easy
27 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Easy
32 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Neither Easy nor Difficult
9 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Difficult
2 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Very Difficult
0 Participants
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Not Applicable
0 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Very Easy
42 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Easy
16 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Neither Easy nor Difficult
6 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Difficult
1 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Very Difficult
0 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Not Applicable
5 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Very Easy
54 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Easy
12 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Difficult
1 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Very Difficult
0 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Not Applicable
0 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Very Easy
11 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Easy
8 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Neither Easy nor Difficult
2 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Difficult
2 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Very Difficult
1 Participants
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Not Applicable
46 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Very Easy
39 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Easy
27 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Neither Easy nor Difficult
3 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Difficult
1 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Very Difficult
0 Participants
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Not Applicable
0 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Very Easy
48 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Neither Easy nor Difficult
3 Participants
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Not Applicable
0 Participants

Adverse Events

18-80 Year Old Smokers

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

VP of Clinical and Medical Affairs

Pivot Health Technologies Inc.

Phone: 415-757-7696

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place