Trial Outcomes & Findings for Pivot Breath Sensor Performance Study (NCT NCT03870022)
NCT ID: NCT03870022
Last Updated: 2024-07-01
Results Overview
Assess user performance of the study device at a single visit. Participants will be observed if they can set up the Pivot Breath Sensor, and whether they can use the device. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance.
COMPLETED
NA
70 participants
Each study session lasted up to 60 minutes per participant. Use performance was assessed at approximately 10-15 minutes into the study session.
2024-07-01
Participant Flow
Participant milestones
| Measure |
18-80 Year Old Smokers
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
Overall Study
STARTED
|
70
|
|
Overall Study
COMPLETED
|
70
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Pivot Breath Sensor Performance Study
Baseline characteristics by cohort
| Measure |
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
Age, Customized
Age 18-49
|
34 Participants
n=93 Participants
|
|
Age, Customized
Age 50-80
|
36 Participants
n=93 Participants
|
|
Sex: Female, Male
Female
|
30 Participants
n=93 Participants
|
|
Sex: Female, Male
Male
|
40 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
White
|
44 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Hispanic, Latino, or Spanish Origin
|
6 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Black or African American
|
10 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Asian
|
8 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
American Indian or Alaska Native
|
1 Participants
n=93 Participants
|
|
Race/Ethnicity, Customized
Hawaiian or other Pacific Islander
|
1 Participants
n=93 Participants
|
|
Region of Enrollment
United States
|
70 participants
n=93 Participants
|
|
years smoking
|
29.8 years
STANDARD_DEVIATION 14.8 • n=93 Participants
|
PRIMARY outcome
Timeframe: Each study session lasted up to 60 minutes per participant. Use performance was assessed at approximately 10-15 minutes into the study session.Assess user performance of the study device at a single visit. Participants will be observed if they can set up the Pivot Breath Sensor, and whether they can use the device. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance.
Outcome measures
| Measure |
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
User Performance of Pivot Breath Sensor
Self-training · Unsuccessful
|
2 Participants
|
|
User Performance of Pivot Breath Sensor
Self-training · Successful
|
68 Participants
|
|
User Performance of Pivot Breath Sensor
Performance breath sample use and/or with use of reference materials · Successful
|
70 Participants
|
|
User Performance of Pivot Breath Sensor
Performance breath sample use and/or with use of reference materials · Unsuccessful
|
0 Participants
|
PRIMARY outcome
Timeframe: Each study session lasted up to 60 minutes per participant.Compare study device performance with a similar device at a single visit. Participants will then be asked to use a similar CO breath sensor device. The correlation between the measured CO levels (in parts per million, or ppm) of the study CO breath sensor (Carrot Sense) and the similar breath sensor will be assessed.
Outcome measures
| Measure |
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
Compare Performance of Pivot Breath Sensor With a Similar Device.
|
0.9710 Correlation coefficient (R-value)
|
SECONDARY outcome
Timeframe: Each study session lasted up to 60 minutes per participant. The user documentation assessment with observer ratings was done between approximately 10 and 35 minutes into the study session.Participant understanding of user documentation will be assessed at single visit through interpretation of three different CO test results: green, orange, and red. User documentation used for the interpretation of each result includes: packaging information, quick start guide and user manual. Observer ratings include 'Successful' and 'Unsuccessful'. "Successful" = the participant was able to complete the task independently, safely, effectively, and without difficulty. "Unsuccessful" = the participant was unable to complete the task independently, safely, effectively, and/or without difficulty. Includes instances where the participant required study moderator assistance. Participant will be asked to assess the user documentation throughout the study session and will take approximately 10 minutes.
Outcome measures
| Measure |
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
User Documentation Assessment With Observer Ratings
Interpretation of Orange CO category after using reference materials · Successful
|
57 Participants
|
|
User Documentation Assessment With Observer Ratings
Interpretation of Green CO category after using reference materials · Successful
|
69 Participants
|
|
User Documentation Assessment With Observer Ratings
Interpretation of Green CO category after using reference materials · Unsuccessful
|
1 Participants
|
|
User Documentation Assessment With Observer Ratings
Interpretation of Orange CO category after using reference materials · Unsuccessful
|
13 Participants
|
|
User Documentation Assessment With Observer Ratings
Interpretation of Red CO category after using reference materials · Successful
|
70 Participants
|
|
User Documentation Assessment With Observer Ratings
Interpretation of Red CO category after using reference materials · Unsuccessful
|
0 Participants
|
SECONDARY outcome
Timeframe: Each study session lasted up to 60 minutes per participant. Use performance was assessed with rating scales at approximately 35 minutes into the study session.Participant will be asked to provide ratings on the following: * How would you rate the ease or difficulty using the device? * How clear or unclear was the Quick Start Guide? Why? * How clear or unclear was the Packaging? Why? * How clear or unclear was the User Manual on the iPad? Why? * Overall, how clear or unclear were the provided reference materials? Why? * How easy or difficult was it to understand and interpret the test results? Why? The participant will use the following rating scale: 5 - Very Easy 4 - Easy 3 - Neither Easy nor Difficult 2 - Difficult 1 - Very Difficult Not Applicable Participant will provide the ratings prior to the conclusion of the study session. Participant assessment with questionnaire rating scales will take approximately 1-2 minutes each.
Outcome measures
| Measure |
18-80 Year Old Smokers
n=70 Participants
Smokers aged 18-80 - collection of 1 paired breath sample from each subject using two non-invasive breath sampling devices, as well as collecting participant responses to user documentation questions and rating scales.
Pivot Breath Sensor System: Collect breath sample from the Pivot Breath Sensor to measure the carbon monoxide in the participant's exhaled breath, along with recording responses to user documentation questions and rating scales.
Bedfont - Micro+™ Smokerlyzer®: Bedfont - Micro+™ Smokerlyzer® : Predicate Device to measure carbon monoxide levels in one's breath..
|
|---|---|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Neither Easy nor Difficult
|
3 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Easy
|
16 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Difficult
|
2 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Very Difficult
|
1 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Very Easy
|
27 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Easy
|
32 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Neither Easy nor Difficult
|
9 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Difficult
|
2 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Very Difficult
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How would you rate the ease or difficulty using the device? · Not Applicable
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Very Easy
|
42 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Easy
|
16 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Neither Easy nor Difficult
|
6 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Difficult
|
1 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Very Difficult
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? Why? · Not Applicable
|
5 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Very Easy
|
54 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Easy
|
12 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Difficult
|
1 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Very Difficult
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the Packaging? Why? · Not Applicable
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Very Easy
|
11 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Easy
|
8 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Neither Easy nor Difficult
|
2 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Difficult
|
2 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Very Difficult
|
1 Participants
|
|
Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? Why? · Not Applicable
|
46 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Very Easy
|
39 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Easy
|
27 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Neither Easy nor Difficult
|
3 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Difficult
|
1 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Very Difficult
|
0 Participants
|
|
Participant Feedback Using Rating Scale
Overall, how clear or unclear were the provided reference materials? Why? · Not Applicable
|
0 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Very Easy
|
48 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Neither Easy nor Difficult
|
3 Participants
|
|
Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? Why? · Not Applicable
|
0 Participants
|
Adverse Events
18-80 Year Old Smokers
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
VP of Clinical and Medical Affairs
Pivot Health Technologies Inc.
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place