Trial Outcomes & Findings for Comparison of Depth of Sedation Performance Between SedLine and Comparator Device During General Anesthesia (NCT NCT03865316)

NCT ID: NCT03865316

Last Updated: 2022-04-11

Results Overview

The correlation was computed using 1 min average values between the PSi and the comparator's sedation index.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

108 participants

Primary outcome timeframe

Duration of surgery

Results posted on

2022-04-11

Participant Flow

Participant milestones

Participant milestones
Measure
SedLine Vs Comparator Test Group
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
Overall Study
STARTED
108
Overall Study
COMPLETED
100
Overall Study
NOT COMPLETED
8

Reasons for withdrawal

Reasons for withdrawal
Measure
SedLine Vs Comparator Test Group
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
Overall Study
Procedure Cancelled
4
Overall Study
Equipment failure
4

Baseline Characteristics

Comparison of Depth of Sedation Performance Between SedLine and Comparator Device During General Anesthesia

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
SedLine Vs Comparator Test Group
n=100 Participants
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
63 Participants
n=5 Participants
Age, Categorical
>=65 years
37 Participants
n=5 Participants
Sex: Female, Male
Female
38 Participants
n=5 Participants
Sex: Female, Male
Male
62 Participants
n=5 Participants
Race/Ethnicity, Customized
African American
5 participants
n=5 Participants
Race/Ethnicity, Customized
White
74 participants
n=5 Participants
Race/Ethnicity, Customized
American Indian
1 participants
n=5 Participants
Race/Ethnicity, Customized
Asian
1 participants
n=5 Participants
Race/Ethnicity, Customized
Other
5 participants
n=5 Participants
Race/Ethnicity, Customized
Not reported
14 participants
n=5 Participants
Region of Enrollment
United States
100 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Duration of surgery

Population: Due to the complexity of data collection, complete data for analysis was available from 61 subjects.

The correlation was computed using 1 min average values between the PSi and the comparator's sedation index.

Outcome measures

Outcome measures
Measure
SedLine Vs Comparator Test Group
n=61 Participants
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
Correlation of Sedation Indices Performance Across Anesthesia Range
0.8314 Pearson r
Interval 0.824 to 0.8385

Adverse Events

SedLine Vs Comparator Test Group

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
SedLine Vs Comparator Test Group
n=100 participants at risk
SedLine Vs Comparator Test Group: Subjects will receive electrodes that compare EEG signals between Masimo's SedLine system and the comparator's system.
Skin and subcutaneous tissue disorders
Petechial Rash
1.0%
1/100 • Number of events 1 • Approximately 8 hours.

Additional Information

Cristina Johnson

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: GT60