Trial Outcomes & Findings for Pivot Breath Sensor Human Factors and Usability Study (NCT NCT03835260)
NCT ID: NCT03835260
Last Updated: 2022-11-08
Results Overview
User performance of participants in intended use scenarios will be assessed at a single visit. Participants will be observed if they can set up the device, use the device, and interpret results. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. 'Successful' = user was able to complete the task safely and effectively. Unsuccessful' = user was unable unable to complete the task independently or did not complete the task safely and effectively. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 60 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute and assessments will be made over the study session which can last up to 60 minutes.
COMPLETED
NA
15 participants
Each study session was completed on a single day, for up to 60 minutes. Device set up was assessed at about 5 minutes in. Device use was assessed at about 10 minutes in. Interpretation of results was assessed at about 15 and 20 minutes in.
2022-11-08
Participant Flow
Participant milestones
| Measure |
Pivot Breath Sensor (User Group)
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Overall Study
STARTED
|
15
|
|
Overall Study
COMPLETED
|
15
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Pivot Breath Sensor Human Factors and Usability Study
Baseline characteristics by cohort
| Measure |
Pivot Breath Sensor (User Group)
n=15 Participants
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Age, Customized
Age · Age 18-49 years
|
13 Participants
n=5 Participants
|
|
Age, Customized
Age · Age 49-50 years
|
2 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
7 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
8 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Caucasian
|
10 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · African American
|
1 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Middle Eastern
|
2 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Native American
|
1 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Ethnicity · Other
|
1 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
15 participants
n=5 Participants
|
|
Daily Smoking Frequency (cigarettes per day)
2-5 cpd
|
2 Participants
n=5 Participants
|
|
Daily Smoking Frequency (cigarettes per day)
6-10 cpd
|
4 Participants
n=5 Participants
|
|
Daily Smoking Frequency (cigarettes per day)
11-15 cpd
|
4 Participants
n=5 Participants
|
|
Daily Smoking Frequency (cigarettes per day)
16+ cpd
|
5 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: Each study session was completed on a single day, for up to 60 minutes. Device set up was assessed at about 5 minutes in. Device use was assessed at about 10 minutes in. Interpretation of results was assessed at about 15 and 20 minutes in.User performance of participants in intended use scenarios will be assessed at a single visit. Participants will be observed if they can set up the device, use the device, and interpret results. Observer ratings include 'Successful' and 'Unsuccessful'. Specific attention will be given to assess if any harms can occur from observed use errors. 'Successful' = user was able to complete the task safely and effectively. Unsuccessful' = user was unable unable to complete the task independently or did not complete the task safely and effectively. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 60 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute and assessments will be made over the study session which can last up to 60 minutes.
Outcome measures
| Measure |
Pivot Breath Sensor (User Group)
n=15 Participants
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Set up device - open packaging · Successful
|
15 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Set up device - open packaging · Unsuccessful
|
0 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Use device · Successful
|
14 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Use device · Unsuccessful
|
1 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Interpret results · Successful
|
15 Participants
|
|
Human Factors - User Performance in Use Scenarios Assessment With Observer Ratings
Interpret results · Unsuccessful
|
0 Participants
|
PRIMARY outcome
Timeframe: Each study session was completed on a single day, for up to 60 minutes. The user documentation assessment occurred at approximately 30 minutes into the study session.Participant understanding of user documentation will be assessed at single visit. User documentation includes packaging information, quick start guide and user manual. Observer ratings include 'Successful' and 'Unsuccessful'. 'Successful' = user was able to complete the task safely and effectively. Unsuccessful' = user was unable unable to complete the task independently or did not complete the task safely and effectively. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 60 minutes. Observations are made throughout the study session. Assessments for individual tasks will take approximately 1 minute each. The User documentation assessment portion of the study session will last up to 30 minutes.
Outcome measures
| Measure |
Pivot Breath Sensor (User Group)
n=15 Participants
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the packaging · Successful
|
15 Participants
|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the packaging · Unsuccessful
|
0 Participants
|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the quick start guide · Successful
|
14 Participants
|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the quick start guide · Unsuccessful
|
1 Participants
|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the user manual · Successful
|
15 Participants
|
|
Human Factors - User Documentation Assessment With Observer Ratings
understanding of the user manual · Unsuccessful
|
0 Participants
|
PRIMARY outcome
Timeframe: Each study session was completed on a single day, for up to 60 minutes. The subjective feedback questionnaire was administered at approximately 40 minutes into the study session.Subjective feedback focusing on any concerns on using device will be assessed at single visit. Participants will answer with a Yes or No to the questions. For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 60 minutes. Subjective feedback focusing on any concerns prior to the conclusion of the study session. Subjective feedback questions will take approximately 1-3 minutes each. The Subjective feedback portion of the study session will last up to 5 to 15 minutes.
Outcome measures
| Measure |
Pivot Breath Sensor (User Group)
n=15 Participants
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Human Factors - Subjective Feedback Questionnaire
Dd you experience any moments of hesitation while using the product? · yes
|
10 Participants
|
|
Human Factors - Subjective Feedback Questionnaire
Dd you experience any moments of hesitation while using the product? · no
|
5 Participants
|
|
Human Factors - Subjective Feedback Questionnaire
Do you have any safety concerns regarding this product or its labeling? · yes
|
3 Participants
|
|
Human Factors - Subjective Feedback Questionnaire
Do you have any safety concerns regarding this product or its labeling? · no
|
12 Participants
|
PRIMARY outcome
Timeframe: Each study session was completed on a single day, for up to 60 minutes. Participant feedback using a rating scale was provided at approximately 50 minutes into the study session.Participant will be asked to provide ratings on the following: * How would you rate the ease or difficulty using the device? * How clear or unclear was the Quick Start Guide? * How clear or unclear was the Packaging? * How clear or unclear was the User Manual? * How easy or difficult was it to understand and interpret the test results? The participant will use the following rating scale (higher score = better outcome): 5 - Very Easy 4 - Easy 3 - Neither Easy nor Difficult 2 - Difficult 1 - Very difficult For each participant, the single visit includes the consent process, entry into the study, the study session, and conclusion of the study. The study session is completed on a single day requiring up to 60 minutes. Participant will provide the ratings prior to the conclusion of the study session. Participant assessment with questionnaire rating scales will take approximately 1-2 minutes each.
Outcome measures
| Measure |
Pivot Breath Sensor (User Group)
n=15 Participants
Self-reported daily smokers of 2 or more cigarettes per day
Pivot Breath Sensor: The Pivot Breath Sensor measures the level of carbon monoxide in exhaled breath and displays the carbon monoxide value to the user.
|
|---|---|
|
Human Factors - Participant Feedback Using Rating Scale
How would you rate the ease or difficulty of using the device? · very easy/very clear
|
5 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How would you rate the ease or difficulty of using the device? · easy/clear
|
8 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How would you rate the ease or difficulty of using the device? · neither easy/clear nor difficult/unclear
|
2 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How would you rate the ease or difficulty of using the device? · difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How would you rate the ease or difficulty of using the device? · very difficult/very unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? · very easy/very clear
|
8 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? · easy/clear
|
6 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? · neither easy/clear nor difficult/unclear
|
1 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? · difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the Quick Start Guide? · very difficult/very unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the packaging? · very easy/very clear
|
9 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the packaging? · easy/clear
|
5 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the packaging? · neither easy/clear nor difficult/unclear
|
1 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the packaging? · difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the packaging? · very difficult/very unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? · very easy/very clear
|
11 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? · easy/clear
|
4 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? · neither easy/clear nor difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? · difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How clear or unclear was the User Manual on the iPad? · very difficult/very unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? · very easy/very clear
|
5 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? · easy/clear
|
8 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? · neither easy/clear nor difficult/unclear
|
2 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? · difficult/unclear
|
0 Participants
|
|
Human Factors - Participant Feedback Using Rating Scale
How easy or difficult was it to understand and interpret the test results? · very difficult/very unclear
|
0 Participants
|
Adverse Events
Pivot Breath Sensor (User Group)
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Senior Director of Clinical Affairs and Data Science
Pivot Health Technologies Inc.
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place