Trial Outcomes & Findings for Intramuscular Heating Rates of a Chattanooga Intelect XT Therapeutic Ultrasound (NCT NCT03816709)
NCT ID: NCT03816709
Last Updated: 2021-05-27
Results Overview
A thermocouple was inserted at the 1 cm depth to determine the tissue temperatures.
COMPLETED
NA
25 participants
Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 6 minutes)
2021-05-27
Participant Flow
Twenty-five subjects were recruited through the Institution's Listserve from January-February, 2019.
Participant milestones
| Measure |
Therapeutic Ultrasound Treatment
All subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Therapeutic Ultrasound: A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes.
|
|---|---|
|
Overall Study
STARTED
|
25
|
|
Overall Study
COMPLETED
|
25
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Therapeutic Ultrasound Treatment
n=25 Participants
All subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Therapeutic Ultrasound: A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=25 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
25 Participants
n=25 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=25 Participants
|
|
Sex: Female, Male
Female
|
14 Participants
n=25 Participants
|
|
Sex: Female, Male
Male
|
11 Participants
n=25 Participants
|
|
Region of Enrollment
United States
|
25 participants
n=25 Participants
|
PRIMARY outcome
Timeframe: Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 6 minutes)Population: Intent to treat population. All subjects received the Ultrasound treatment with a thermocouple recording tissue temperatures at the 1 cm depth.
A thermocouple was inserted at the 1 cm depth to determine the tissue temperatures.
Outcome measures
| Measure |
Therapeutic Ultrasound Treatment
n=25 Participants
All subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Therapeutic Ultrasound: A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes at the depths of 1.0, 1.75, and 2.5 cm.
|
|---|---|
|
Heating Rate at 1 cm Depth
|
6.49 degrees Celcius
Standard Deviation 1.24
|
PRIMARY outcome
Timeframe: Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which occurred at 8 minutes)Population: Intent to treat. All subject received the Ultrasound treatment with a thermocouple at the 1.75 cm depth to determine tissue temperature
A thermocouple was inserted at the 1.75 cm depth to determine the tissue temperatures.
Outcome measures
| Measure |
Therapeutic Ultrasound Treatment
n=25 Participants
All subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Therapeutic Ultrasound: A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes at the depths of 1.0, 1.75, and 2.5 cm.
|
|---|---|
|
Heating Rate at 1.75 cm Depth
|
5.50 degrees Celcius
Standard Deviation 1.34
|
PRIMARY outcome
Timeframe: Baseline to end of treatment (up to 15 minutes or until temperatures increased 4 degrees, which did not occur in the 15 minute treatment)Population: Intent to treat. All subjects received the Ultrasound treatment with a thermocouple at the 2.5 cm depth to determine tissue temperatures.
A thermocouple was inserted at the 2.5 cm depth to determine the tissue temperatures.
Outcome measures
| Measure |
Therapeutic Ultrasound Treatment
n=25 Participants
All subjects received a therapeutic ultrasound treatment with the Chattanooga Intelect Legend XT machine with the following parameters: 3 MHz, 1.0 W/cm2, 15 minute treatment time.
Therapeutic Ultrasound: A therapeutic ultrasound treatment was applied to the left medial gastrocnemius muscle with the following parameters: 3 MHz, 1.0 W/cm2 for 15 minutes at the depths of 1.0, 1.75, and 2.5 cm.
|
|---|---|
|
Heating Rate at 2.5 cm Depth
|
3.07 degrees Celcius
Standard Deviation 1.53
|
Adverse Events
Therapeutic Ultrasound Treatment
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported