Trial Outcomes & Findings for Accuracy of Non-invasive Non-oscillometric Blood Pressure Wristband (NCT NCT03629535)

NCT ID: NCT03629535

Last Updated: 2021-02-02

Results Overview

If the value obtained from the LiveMetric determination lied within the range of the reference BP, an error of 0 mm Hg (0kPa) was assigned to this determination. If the value obtained from the LiveMetric determination lies outside the range of the reference BP, the value of the determination from the adjacent limit of the range of the variation of BP was subtracted. That difference represents the error for this determination. The arithmetic mean of the error is calculated and its experimental standard deviation from the errors of each recording for each BP value (Systolic \& Diastolic) and determination for each patient.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

8 participants

Primary outcome timeframe

during intervention for 30 minutes

Results posted on

2021-02-02

Participant Flow

Participant milestones

Participant milestones
Measure
Patients in SICU
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects. Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet. Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
Overall Study
STARTED
8
Overall Study
COMPLETED
8
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Accuracy of Non-invasive Non-oscillometric Blood Pressure Wristband

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Patients in SICU
n=8 Participants
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects. Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet. Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
Age, Continuous
61.6 years
STANDARD_DEVIATION 16.4 • n=5 Participants
Sex: Female, Male
Female
5 Participants
n=5 Participants
Sex: Female, Male
Male
3 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
1 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
3 Participants
n=5 Participants
Race (NIH/OMB)
White
0 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
4 Participants
n=5 Participants
Region of Enrollment
United States
8 participants
n=5 Participants

PRIMARY outcome

Timeframe: during intervention for 30 minutes

If the value obtained from the LiveMetric determination lied within the range of the reference BP, an error of 0 mm Hg (0kPa) was assigned to this determination. If the value obtained from the LiveMetric determination lies outside the range of the reference BP, the value of the determination from the adjacent limit of the range of the variation of BP was subtracted. That difference represents the error for this determination. The arithmetic mean of the error is calculated and its experimental standard deviation from the errors of each recording for each BP value (Systolic \& Diastolic) and determination for each patient.

Outcome measures

Outcome measures
Measure
Patients in SICU
n=8 Participants
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects. Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet. Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
Average Error in Blood Pressure Readings
Systolic
-1.5 mmHg
Standard Deviation 6.3
Average Error in Blood Pressure Readings
Diastolic
3.2 mmHg
Standard Deviation 6.0

Adverse Events

Patients in SICU

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported