Trial Outcomes & Findings for Accuracy of Non-invasive Non-oscillometric Blood Pressure Wristband (NCT NCT03629535)
NCT ID: NCT03629535
Last Updated: 2021-02-02
Results Overview
If the value obtained from the LiveMetric determination lied within the range of the reference BP, an error of 0 mm Hg (0kPa) was assigned to this determination. If the value obtained from the LiveMetric determination lies outside the range of the reference BP, the value of the determination from the adjacent limit of the range of the variation of BP was subtracted. That difference represents the error for this determination. The arithmetic mean of the error is calculated and its experimental standard deviation from the errors of each recording for each BP value (Systolic \& Diastolic) and determination for each patient.
COMPLETED
NA
8 participants
during intervention for 30 minutes
2021-02-02
Participant Flow
Participant milestones
| Measure |
Patients in SICU
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects.
Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet.
Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
|
|---|---|
|
Overall Study
STARTED
|
8
|
|
Overall Study
COMPLETED
|
8
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy of Non-invasive Non-oscillometric Blood Pressure Wristband
Baseline characteristics by cohort
| Measure |
Patients in SICU
n=8 Participants
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects.
Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet.
Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
|
|---|---|
|
Age, Continuous
|
61.6 years
STANDARD_DEVIATION 16.4 • n=5 Participants
|
|
Sex: Female, Male
Female
|
5 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
3 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
1 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
3 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
4 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
8 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: during intervention for 30 minutesIf the value obtained from the LiveMetric determination lied within the range of the reference BP, an error of 0 mm Hg (0kPa) was assigned to this determination. If the value obtained from the LiveMetric determination lies outside the range of the reference BP, the value of the determination from the adjacent limit of the range of the variation of BP was subtracted. That difference represents the error for this determination. The arithmetic mean of the error is calculated and its experimental standard deviation from the errors of each recording for each BP value (Systolic \& Diastolic) and determination for each patient.
Outcome measures
| Measure |
Patients in SICU
n=8 Participants
Patients in the SICU with an intra-arterial blood pressure monitor already in place will be considered as subjects.
Non-invasive non-oscillometric blood pressure wristband device: The device is a band intended to be worn on the wrist for intermittent blood pressure measurements. The device applies no inflatable mechanics or moving parts. The wristband is watertight and can be worn as any type of bracelet.
Invasive intra-arterial blood pressure monitors.: Because the subjects also have an intra-arterial blood pressure monitor in place, the subjects will serve as their own control
|
|---|---|
|
Average Error in Blood Pressure Readings
Systolic
|
-1.5 mmHg
Standard Deviation 6.3
|
|
Average Error in Blood Pressure Readings
Diastolic
|
3.2 mmHg
Standard Deviation 6.0
|
Adverse Events
Patients in SICU
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place