Trial Outcomes & Findings for Feasibility and Performance Evaluation for INVSENSOR00024 (NCT NCT03570255)
NCT ID: NCT03570255
Last Updated: 2020-02-20
Results Overview
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the root mean square (RMS) error value. In order to obtain the RMS value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for each sample, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the RMS Error value.
COMPLETED
NA
28 participants
up to 8 hours
2020-02-20
Participant Flow
Participant milestones
| Measure |
RD SET Neo SpO2
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
RD SET Neo SpO2: All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.
|
|---|---|
|
Overall Study
STARTED
|
28
|
|
Overall Study
COMPLETED
|
28
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Feasibility and Performance Evaluation for INVSENSOR00024
Baseline characteristics by cohort
| Measure |
RD SET Neo SpO2
n=28 Participants
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
RD SET Neo SpO2: All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.
|
|---|---|
|
Age, Categorical
<=18 years
|
28 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
10 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
18 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
3 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
1 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
14 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
10 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
28 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: up to 8 hoursPopulation: Subject 7 \& 11: Protocol deviation Subject 19 \& 26: Outlier data points Subject 5: Removed per ISO standard for SpO2 equal to 100.
Accuracy will be determined by comparing the noninvasive blood oxygen saturation measurement of the pulse oximeter to that obtained from a blood sample and calculating the root mean square (RMS) error value. In order to obtain the RMS value, the blood oxygen saturation measurement form a laboratory pulse Co-Oximeter is subtracted from the pulse oximeter oxygen saturation measurement for each sample, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the RMS Error value.
Outcome measures
| Measure |
RD SET Neo SpO2
n=23 Participants
All subjects enrolled in the study will receive the investigational sensor (RD SET Neo SpO2) for evaluation of SpO2.
RD SET Neo SpO2: All subjects are enrolled in the experimental group and receive the RD SET Neo SpO2 sensor.
|
|---|---|
|
Accuracy of Oxygen Saturation (SpO2) Measurement by RMS Calculation
|
2.1 percent of oxygen saturated hemoglobin
|