Trial Outcomes & Findings for A Comparison of Capnography Sampling Lines (NCT NCT03554629)

NCT ID: NCT03554629

Last Updated: 2021-01-06

Results Overview

Comparison of measured end-tidal CO2 (EtCO2), Rebreathing as measured by fractional inspiratory CO2 (FiCO2) as a function the activity for 8 different CCSF designs tested on 23 subjects with supplemental O2 at 5lmp for four activities. Device data collected a 20Hz (20 data points/second) for the 3 hours of enrollment. Participants lateral lying with head on a pillow with supplemental O2 during each nasal only or open mouth breathing at respiration rate of 6 and 24.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

30 participants

Primary outcome timeframe

Three hours

Results posted on

2021-01-06

Participant Flow

Execution of CIP v 1 enrollment at 6; data not included in results since data inadequate. Execution of CIP V2 enrollment was 23 One subject withdrawn before participation due to failure to meet failure to meet criteria of enrollment on the study day (runny nose)

Unit of analysis: Nasal/Oral-Nasal Sampling

Participant milestones

Participant milestones
Measure
Eight Capnography CO2 Sampling Filterline Performance
Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities for data analysis. All subjects worn all 8 CCSF devices under test for the four study activities. 1 and 2. Closed mouth breathing with respiration rate of 6 and 24 3 and 4. Open mouth breathing with respiration rate at 6 and 24
Overall Study
STARTED
29 232
Overall Study
COMPLETED
23 184
Overall Study
NOT COMPLETED
6 48

Reasons for withdrawal

Reasons for withdrawal
Measure
Eight Capnography CO2 Sampling Filterline Performance
Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities for data analysis. All subjects worn all 8 CCSF devices under test for the four study activities. 1 and 2. Closed mouth breathing with respiration rate of 6 and 24 3 and 4. Open mouth breathing with respiration rate at 6 and 24
Overall Study
Inadequate data
6

Baseline Characteristics

A Comparison of Capnography Sampling Lines

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Capnography CO2 Sampling Filterline
n=23 Participants
Capnography CO2 Sampling Filterline: Use of up to 8 Capnography CO2 Sampling FIlterlines by blinded code name during scripted activities
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
18 Participants
n=5 Participants
Age, Categorical
>=65 years
5 Participants
n=5 Participants
Age, Continuous
48.5 years
STANDARD_DEVIATION 17.1 • n=5 Participants
Sex: Female, Male
Female
14 Participants
n=5 Participants
Sex: Female, Male
Male
9 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
2 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
21 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
1 Participants
n=5 Participants
Race (NIH/OMB)
Asian
0 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
3 Participants
n=5 Participants
Race (NIH/OMB)
White
19 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Region of Enrollment
United States
23 Participants
n=5 Participants
Body Mass Index (BMI)
27.5 kg/m^2
STANDARD_DEVIATION 6.6 • n=5 Participants

PRIMARY outcome

Timeframe: Three hours

Population: Performance of 8 CO2 cannula sampling filterlines (CCSF): 4 filterlines provided nasal only sampling ports and 4 filterlines provided both nasal and oral sampling ports. Each of the 8 CCSF was assigned a letter by the investigation site to maintain blinding in data analysis. Side-stream capnography measurement via the Capnostream 35

Comparison of measured end-tidal CO2 (EtCO2), Rebreathing as measured by fractional inspiratory CO2 (FiCO2) as a function the activity for 8 different CCSF designs tested on 23 subjects with supplemental O2 at 5lmp for four activities. Device data collected a 20Hz (20 data points/second) for the 3 hours of enrollment. Participants lateral lying with head on a pillow with supplemental O2 during each nasal only or open mouth breathing at respiration rate of 6 and 24.

Outcome measures

Outcome measures
Measure
EtCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to sample gas for EtCO2 capnography measurement under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to sample gas for end tidal CO2 (EtCO2)capnography measurement under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
EtCO2 During Mouth Open Breathing Rate of 6
n=23 Participants
Patient interface to sample gas for end tidal CO2 (EtCO2) capnography measurement under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Open Breathing Rate of 6
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 Mouth Open Breathing Rate of 24
n=23 Participants
Patient interface to sample gas for capnography measurement under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 Mouth Open Breathing Rate of 24
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
FlexiCare (Nasal)
30.6 mmHg
Standard Deviation 7.5
0.0 mmHg
Standard Deviation 0.2
26.6 mmHg
Standard Deviation 6.2
0.2 mmHg
Standard Deviation 0.8
20.1 mmHg
Standard Deviation 12.5
0.1 mmHg
Standard Deviation 0.2
19.1 mmHg
Standard Deviation 10.2
0.2 mmHg
Standard Deviation 0.6
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Hudson RCI /TeleFlex(Nasal)
13.5 mmHg
Standard Deviation 8.4
0.0 mmHg
Standard Deviation 0.0
12.3 mmHg
Standard Deviation 6.9
0.0 mmHg
Standard Deviation 0.1
7.6 mmHg
Standard Deviation 6.6
0.0 mmHg
Standard Deviation 0.0
7.8 mmHg
Standard Deviation 7.2
0.0 mmHg
Standard Deviation 0.0
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Salter Lab (Nasal)
30.1 mmHg
Standard Deviation 5.5
0.4 mmHg
Standard Deviation 0.8
24.1 mmHg
Standard Deviation 5.9
1.4 mmHg
Standard Deviation 1.2
22.3 mmHg
Standard Deviation 9.2
0.2 mmHg
Standard Deviation 0.3
19.9 mmHg
Standard Deviation 7.6
1.0 mmHg
Standard Deviation 1.1
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Medtronic Filterline (Nasal)
32.6 mmHg
Standard Deviation 5.4
0.0 mmHg
Standard Deviation 0.0
28.1 mmHg
Standard Deviation 5.2
0.0 mmHg
Standard Deviation 0.2
24.8 mmHg
Standard Deviation 8.3
0.0 mmHg
Standard Deviation 0.0
22.7 mmHg
Standard Deviation 7.6
0.0 mmHg
Standard Deviation 0.0
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
WestMed (Oral/Nasal)
33.9 mmHg
Standard Deviation 4.8
0.5 mmHg
Standard Deviation 0.4
29.2 mmHg
Standard Deviation 5.8
1.4 mmHg
Standard Deviation 1.5
28.0 mmHg
Standard Deviation 8.9
0.5 mmHg
Standard Deviation 0.6
24.1 mmHg
Standard Deviation 8.1
1.2 mmHg
Standard Deviation 1.4
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Dispo-Med (Oral/Nasal)
29.9 mmHg
Standard Deviation 7.5
0.1 mmHg
Standard Deviation 0.3
23.8 mmHg
Standard Deviation 8.2
1.4 mmHg
Standard Deviation 0.9
23.6 mmHg
Standard Deviation 6.9
0.1 mmHg
Standard Deviation 0.2
19.3 mmHg
Standard Deviation 7.0
1.4 mmHg
Standard Deviation 1.3
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Salter Lab Oral-Trac (Oral/Nasal)
29.3 mmHg
Standard Deviation 8.7
0.3 mmHg
Standard Deviation 0.5
25.0 mmHg
Standard Deviation 6.9
0.6 mmHg
Standard Deviation 0.9
26.8 mmHg
Standard Deviation 5.2
0.2 mmHg
Standard Deviation 0.3
23.4 mmHg
Standard Deviation 4.7
0.9 mmHg
Standard Deviation 1.1
Capnography Measured CO2 Waveform Data as a Function of CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity
Medtronic Smart Capnoline+(Oral/Nasal)
32.6 mmHg
Standard Deviation 5.4
0.0 mmHg
Standard Deviation 0.0
28.8 mmHg
Standard Deviation 4.8
0.0 mmHg
Standard Deviation 0.1
26.8 mmHg
Standard Deviation 7.4
0.0 mmHg
Standard Deviation 0.0
24.9 mmHg
Standard Deviation 6.2
0.0 mmHg
Standard Deviation 0.1

SECONDARY outcome

Timeframe: Three hours

Population: Participant actual Respiration Rate was coached using the Pace Breathing Android App: Version 2.1. RR = 6 : Inspiratory Time was 5.0 seconds and Expiratory time was 5.0 seconds No breath alarm is deployed when both RR and EtCO2 are equal to zero. RR = 24: Inspiratory Time was 1.2 seconds and Expiratory time was 1.3 seconds

Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.

Outcome measures

Outcome measures
Measure
EtCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to sample gas for EtCO2 capnography measurement under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to sample gas for end tidal CO2 (EtCO2)capnography measurement under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
EtCO2 During Mouth Open Breathing Rate of 6
n=23 Participants
Patient interface to sample gas for end tidal CO2 (EtCO2) capnography measurement under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Open Breathing Rate of 6
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 Mouth Open Breathing Rate of 24
Patient interface to sample gas for capnography measurement under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 Mouth Open Breathing Rate of 24
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
FlexiCare (Nasal)
0.17 Alarms
Standard Deviation 0.53
0.13 Alarms
Standard Deviation 0.40
0.35 Alarms
Standard Deviation 0.57
0.30 Alarms
Standard Deviation 0.51
0.17 Alarms
Standard Deviation 0.38
0.15 Alarms
Standard Deviation 0.36
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Hudson/TeleFlex (Nasal)
0.20 Alarms
Standard Deviation 0.40
0.13 Alarms
Standard Deviation 0.34
0.74 Alarms
Standard Deviation 0.53
0.63 Alarms
Standard Deviation 0.57
0.52 Alarms
Standard Deviation 0.51
0.50 Alarms
Standard Deviation 0.51
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Salter Labs (Nasal)
0.02 Alarms
Standard Deviation 0.15
0.02 Alarms
Standard Deviation 0.15
0.37 Alarms
Standard Deviation 0.61
0.22 Alarms
Standard Deviation 0.47
0.13 Alarms
Standard Deviation 0.34
0.04 Alarms
Standard Deviation 0.21
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Medtronic CO2 (Nasal)
0.02 Alarms
Standard Deviation 0.15
0.02 Alarms
Standard Deviation 0.15
0.17 Alarms
Standard Deviation 0.38
0.11 Alarms
Standard Deviation 0.31
0.13 Alarms
Standard Deviation 0.40
0.07 Alarms
Standard Deviation 0.25
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
WestMed (Oral/Nasal)
0.04 Alarms
Standard Deviation 0.29
0.04 Alarms
Standard Deviation 0.29
0.13 Alarms
Standard Deviation 0.34
0.11 Alarms
Standard Deviation 0.31
0.09 Alarms
Standard Deviation 0.28
0.09 Alarms
Standard Deviation 0.28
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
DispoMed (Oral/Nasal)
0.04 Alarms
Standard Deviation 0.21
0.04 Alarms
Standard Deviation 0.21
0.37 Alarms
Standard Deviation 0.53
0.28 Alarms
Standard Deviation 0.46
0.20 Alarms
Standard Deviation 0.45
0.15 Alarms
Standard Deviation 0.36
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Salter Oral-Trac (Oral/Nasal)
0.02 Alarms
Standard Deviation 0.15
0.02 Alarms
Standard Deviation 0.15
0.22 Alarms
Standard Deviation 0.47
0.11 Alarms
Standard Deviation 0.31
0.13 Alarms
Standard Deviation 0.40
0.07 Alarms
Standard Deviation 0.25
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 24
Medtronic Smart CapnoLine Plus (Oral/Nasal)
0.00 Alarms
Standard Deviation 0.00
0.00 Alarms
Standard Deviation 0.00
0.09 Alarms
Standard Deviation 0.28
0.09 Alarms
Standard Deviation 0.28
0.04 Alarms
Standard Deviation 0.21
0.04 Alarms
Standard Deviation 0.21

SECONDARY outcome

Timeframe: Three hours

Population: Participant actual Respiration Rate was coached using the Pace Breathing Android App: Version 2.1. RR = 6 : Inspiratory Time was 5.0 seconds and Expiratory time was 5.0 seconds No breath alarm is deployed when both RR and EtCO2 are equal to zero. RR = 24: Inspiratory Time was 1.2 seconds and Expiratory time was 1.3 seconds

Gas sample provided by CCSF to the Capnostream 35 for measurement of the partial pressure of CO2 as a function of time. The Capnostream 35 Respiration rate algorithm counts the breaths per minute based upon the frequency of a CO2 validated breath waveform.

Outcome measures

Outcome measures
Measure
EtCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to sample gas for EtCO2 capnography measurement under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 6
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to sample gas for end tidal CO2 (EtCO2)capnography measurement under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 During Mouth Closed Breathing Rate of 24
n=23 Participants
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of closed mouth and respiration rate of 24 with supplemental O2 at 5lpm.
EtCO2 During Mouth Open Breathing Rate of 6
Patient interface to sample gas for end tidal CO2 (EtCO2) capnography measurement under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
FiCO2 During Mouth Open Breathing Rate of 6
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 6 with supplemental O2 at 5lpm.
EtCO2 Mouth Open Breathing Rate of 24
Patient interface to sample gas for capnography measurement under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
FiCO2 Mouth Open Breathing Rate of 24
Patient interface to remove stale air when providing a gas sample for capnography measurement of Fractional Inspired CO2 (re-breathing) under condition of open mouth breathing and respiration rate of 24 with supplemental O2 at 5lpm.
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
FlexiCare (Nasal)
0.13 Alarms
Standard Deviation 0.34
0.09 Alarms
Standard Deviation 0.28
0.24 Alarms
Standard Deviation 0.43
0.22 Alarms
Standard Deviation 0.42
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Hudson/TeleFlex (Nasal)
0.30 Alarms
Standard Deviation 0.55
0.24 Alarms
Standard Deviation 0.48
0.54 Alarms
Standard Deviation 0.55
0.43 Alarms
Standard Deviation 0.50
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Salter Labs (Nasal)
0.11 Alarms
Standard Deviation 0.31
0.07 Alarms
Standard Deviation 0.25
0.11 Alarms
Standard Deviation 0.31
0.09 Alarms
Standard Deviation 0.28
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Medtronic CO2 (Nasal)
0.07 Alarms
Standard Deviation 0.25
0.07 Alarms
Standard Deviation 0.25
0.13 Alarms
Standard Deviation 0.34
0.09 Alarms
Standard Deviation 0.28
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
WestMed (Oral/Nasal)
0.02 Alarms
Standard Deviation 0.15
0.02 Alarms
Standard Deviation 0.15
0.07 Alarms
Standard Deviation 0.33
0.04 Alarms
Standard Deviation 0.21
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
DispoMed (Oral/Nasal)
0.04 Alarms
Standard Deviation 0.21
0.02 Alarms
Standard Deviation 0.15
0.11 Alarms
Standard Deviation 0.31
0.07 Alarms
Standard Deviation 0.25
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Salter Oral-Trac (Oral/Nasal)
0.04 Alarms
Standard Deviation 0.21
0.04 Alarms
Standard Deviation 0.21
0.11 Alarms
Standard Deviation 0.31
0.07 Alarms
Standard Deviation 0.25
Average Number of Alarms as a Function of Each Capnography CO2 Cannula Sampling Filterline (CCSF) Stratified by Activity With an Actual Respiration Rate of 6
Medtronic Smart CapnoLine Plus (Oral/Nasal)
0.0 Alarms
Standard Deviation 0.0
0.0 Alarms
Standard Deviation 0.0
0.07 Alarms
Standard Deviation 0.25
0.07 Alarms
Standard Deviation 0.25

Adverse Events

FlexiCare (Nasal)

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

Hudson RCI /TeleFlex(Nasal)

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Salter Lab (Nasal)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Medtronic Filterline (Nasal)

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

WestMed (Oral/Nasal) RCI /TeleFlex(Nasal)

Serious events: 0 serious events
Other events: 2 other events
Deaths: 0 deaths

Dispo-Med (Oral/Nasal)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Salter Lab Oral-Trac (Oral/Nasal)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Medtronic Smart Capnoline+(Oral/Nasal)

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
FlexiCare (Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Hudson RCI /TeleFlex(Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Salter Lab (Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Medtronic Filterline (Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
WestMed (Oral/Nasal) RCI /TeleFlex(Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Dispo-Med (Oral/Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Salter Lab Oral-Trac (Oral/Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Medtronic Smart Capnoline+(Oral/Nasal)
n=29 participants at risk
8 Capnography CO2 Sampling FIlterlines blinded by code name during scripted activities with supplemental oxygen at 5lpm for each participant. Adverse event Conditions that required a stop of the procedures until SpO2 \> 90% were the following
Respiratory, thoracic and mediastinal disorders
Hypoxemia
3.4%
1/29 • Number of events 1 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia
3.4%
1/29 • Number of events 1 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia
0.00%
0/29 • 3 hours
Hypoxemia

Additional Information

Kathleen H. Niebel RN BSN BSME

Clinical Research Program Manager

Phone: 978-549-9325

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place