Trial Outcomes & Findings for Personalised HTO Versus Generic HTO Virtual Clinical Trial (NCT NCT03419598)
NCT ID: NCT03419598
Last Updated: 2020-12-07
Results Overview
Mechanical stress (Von Mises) calculated for functional loading using finite element analysis
COMPLETED
30 participants
6 weeks simulated post-operation
2020-12-07
Participant Flow
The virtual patients were created based on CT data sets obtained for 30 patients with knee osteoarthritis. The virtual patients were duplicated, and then enrolled into both arms.
Participant milestones
| Measure |
All Study Participants
Opening wedge high tibial osteotomy (HTO).
Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia.
Since this is a virtual trial, the virtual patients were duplicated and had the osteotomy stabilised by a generic HTO plate in one copy of the cohort and by a personalised HTO plate in the other copy.
|
|---|---|
|
Overall Study
STARTED
|
30
|
|
Overall Study
Subject Specific Tibia Models Created and Virtual Osteotomy Performed
|
28
|
|
Overall Study
Virtual Cohort Duplicated Into A. Personalised and B. Generic Copies
|
28
|
|
Overall Study
Finite Element (FE) Model Created for Cohort Copy Stabilised by Personalised HTO Plate
|
28
|
|
Overall Study
FE Models Created for Cohort Copy Stabilised by Generic HTO Plate
|
28
|
|
Overall Study
FE Models Solved for Personalised HTO Plate Stabilised Cohort Copy
|
28
|
|
Overall Study
FE Models Solved for Generic HTO Plate Stabilised Cohort Copy
|
28
|
|
Overall Study
COMPLETED
|
28
|
|
Overall Study
NOT COMPLETED
|
2
|
Reasons for withdrawal
| Measure |
All Study Participants
Opening wedge high tibial osteotomy (HTO).
Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia.
Since this is a virtual trial, the virtual patients were duplicated and had the osteotomy stabilised by a generic HTO plate in one copy of the cohort and by a personalised HTO plate in the other copy.
|
|---|---|
|
Overall Study
Poor quality CT scans could not be adequately segmented to create virtual patient tibial geometry
|
2
|
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
All Study Participants
n=28 Participants
Opening wedge high tibial osteotomy (HTO): Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia
Cohort duplicated into 2 copies. Copy A. Stabilised by Personalised HTO plate. Copy B. Stabilised by Generic HTO plate.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=28 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
10 Participants
n=28 Participants
|
|
Age, Categorical
>=65 years
|
18 Participants
n=28 Participants
|
|
Age, Continuous
|
67.7 years
STANDARD_DEVIATION 9.0 • n=28 Participants
|
|
Sex: Female, Male
Female
|
15 Participants
n=28 Participants
|
|
Sex: Female, Male
Male
|
13 Participants
n=28 Participants
|
|
Region of Enrollment
United Kingdom
|
28 participants
n=28 Participants
|
|
Height (m)
|
1.686 metres
STANDARD_DEVIATION 0.124 • n=28 Participants
|
|
Weight (kg)
|
90.1 kg
STANDARD_DEVIATION 16.4 • n=28 Participants
|
|
Kellgren & Lawrence OA Grade
|
2.6 units on a scale
STANDARD_DEVIATION 0.9 • n=28 Participants
|
PRIMARY outcome
Timeframe: 6 weeks simulated post-operationMechanical stress (Von Mises) calculated for functional loading using finite element analysis
Outcome measures
| Measure |
Personalised HTO
n=28 Participants
Opening wedge high tibial osteotomy
Personalised subject specific custom HTO plate
Opening wedge high tibial osteotomy: Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia
Stabilised by personalised HTO plate, contoured to subject's tibial geometry
|
Generic HTO
n=28 Participants
Opening wedge high tibial osteotomy
Generic HTO plate
Opening wedge high tibial osteotomy: Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia
Generic HTO plate: Generic plate for stabilizing an open wedge HTO
|
|---|---|---|
|
Mechanical Stress in Plate During Gait
|
108.3 MPa
Standard Error 3.81
|
90.6 MPa
Standard Error 3.56
|
SECONDARY outcome
Timeframe: 6 weeks simulated post-operationMechanical strain calculated for functional loading using finite element analysis, note strain is dimensionless and hence has no units
Outcome measures
| Measure |
Personalised HTO
n=28 Participants
Opening wedge high tibial osteotomy
Personalised subject specific custom HTO plate
Opening wedge high tibial osteotomy: Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia
Stabilised by personalised HTO plate, contoured to subject's tibial geometry
|
Generic HTO
n=28 Participants
Opening wedge high tibial osteotomy
Generic HTO plate
Opening wedge high tibial osteotomy: Realignment of knee by creating an opening wedge osteotomy in the upper part of the tibia
Generic HTO plate: Generic plate for stabilizing an open wedge HTO
|
|---|---|---|
|
Mechanical Strain in Bone Around Screws During Gait
|
0.011 unitless
Standard Error 0.00056
|
0.011 unitless
Standard Error 0.00065
|
Adverse Events
All Study Participants
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place