Trial Outcomes & Findings for Desaturation Validation of INVSENSOR00009 (NCT NCT03396835)
NCT ID: NCT03396835
Last Updated: 2019-08-28
Results Overview
Trending ARMs accuracy is defined as root mean squared relative error of all data points of all subjects. Accuracy of the sensors will be determined by comparing regional oxygen saturation (rSO2) readings from INVSENSOR00009 and control sensor and calculating the Arithmetic root mean square (ARMS) value.
COMPLETED
NA
16 participants
1-5 hours
2019-08-28
Participant Flow
Participant milestones
| Measure |
INVSENSOR00009 Sensor
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
|
|---|---|
|
Overall Study
STARTED
|
16
|
|
Overall Study
COMPLETED
|
14
|
|
Overall Study
NOT COMPLETED
|
2
|
Reasons for withdrawal
| Measure |
INVSENSOR00009 Sensor
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
|
|---|---|
|
Overall Study
Withdrawn prior to completion
|
1
|
|
Overall Study
Did not meet criteria for data analysis
|
1
|
Baseline Characteristics
One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Baseline characteristics by cohort
| Measure |
INVSENSOR00009 Sensor
n=16 Participants
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Age, Categorical
Between 18 and 65 years
|
14 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Age, Categorical
>=65 years
|
0 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Sex: Female, Male
Female
|
7 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Sex: Female, Male
Male
|
7 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Race/Ethnicity, Customized
White
|
4 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Race/Ethnicity, Customized
Hispanic
|
3 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Race/Ethnicity, Customized
Asian or Pacific Islander
|
1 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Race/Ethnicity, Customized
African American
|
6 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
|
Region of Enrollment
United States
|
14 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
|
PRIMARY outcome
Timeframe: 1-5 hoursTrending ARMs accuracy is defined as root mean squared relative error of all data points of all subjects. Accuracy of the sensors will be determined by comparing regional oxygen saturation (rSO2) readings from INVSENSOR00009 and control sensor and calculating the Arithmetic root mean square (ARMS) value.
Outcome measures
| Measure |
INVSENSOR00009 Sensor
n=14 Participants
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
|
|---|---|
|
Trending Regional Oxygen Saturation Accuracy of the INVSENSOR00009 Relative to the Control Sensor by Arms Calculation
|
1.3 % tissue oxygen saturation
|
Adverse Events
INVSENSOR00009 Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place