Trial Outcomes & Findings for Desaturation Validation of INVSENSOR00009 (NCT NCT03396835)

NCT ID: NCT03396835

Last Updated: 2019-08-28

Results Overview

Trending ARMs accuracy is defined as root mean squared relative error of all data points of all subjects. Accuracy of the sensors will be determined by comparing regional oxygen saturation (rSO2) readings from INVSENSOR00009 and control sensor and calculating the Arithmetic root mean square (ARMS) value.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

16 participants

Primary outcome timeframe

1-5 hours

Results posted on

2019-08-28

Participant Flow

Participant milestones

Participant milestones
Measure
INVSENSOR00009 Sensor
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
Overall Study
STARTED
16
Overall Study
COMPLETED
14
Overall Study
NOT COMPLETED
2

Reasons for withdrawal

Reasons for withdrawal
Measure
INVSENSOR00009 Sensor
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
Overall Study
Withdrawn prior to completion
1
Overall Study
Did not meet criteria for data analysis
1

Baseline Characteristics

One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
INVSENSOR00009 Sensor
n=16 Participants
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
Age, Categorical
<=18 years
0 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Age, Categorical
Between 18 and 65 years
14 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Age, Categorical
>=65 years
0 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Sex: Female, Male
Female
7 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Sex: Female, Male
Male
7 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Race/Ethnicity, Customized
White
4 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Race/Ethnicity, Customized
Hispanic
3 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Race/Ethnicity, Customized
Asian or Pacific Islander
1 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Race/Ethnicity, Customized
African American
6 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.
Region of Enrollment
United States
14 Participants
n=14 Participants • One subject was withdrawn prior to completing study procedures. One subject did not meet criteria for data analysis.

PRIMARY outcome

Timeframe: 1-5 hours

Trending ARMs accuracy is defined as root mean squared relative error of all data points of all subjects. Accuracy of the sensors will be determined by comparing regional oxygen saturation (rSO2) readings from INVSENSOR00009 and control sensor and calculating the Arithmetic root mean square (ARMS) value.

Outcome measures

Outcome measures
Measure
INVSENSOR00009 Sensor
n=14 Participants
All subjects who are enrolled into the test group and participate in data collection receive both the INVSENSOR00009 and the control sensor simultaneously on the forehead.
Trending Regional Oxygen Saturation Accuracy of the INVSENSOR00009 Relative to the Control Sensor by Arms Calculation
1.3 % tissue oxygen saturation

Adverse Events

INVSENSOR00009 Sensor

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Tala Harake

Masimo Corporation

Phone: 949-297-7137

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place