Trial Outcomes & Findings for Spectrophon LTD Glucometry Monitor Accuracy (NCT NCT03359629)

NCT ID: NCT03359629

Last Updated: 2019-03-07

Results Overview

Difference in measurements of blood glucose level obtained using Spectrophon glucometry monitor and commercially available glucose analyser YSI 2300

Recruitment status

COMPLETED

Target enrollment

210 participants

Primary outcome timeframe

1-3 hours

Results posted on

2019-03-07

Participant Flow

Participant milestones

Participant milestones
Measure
Blood Collection (Healthy Individuals)
Participants's blood was collected and measured for glucose level before and after food intake
Blood Collection (Diabetic Patients)
Participants's blood was collected and measured for glucose level
Overall Study
STARTED
200
10
Overall Study
COMPLETED
200
10
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Blood Collection (Healthy Individuals)
n=200 Participants
Participants's blood was collected and measured for glucose level before and after food intake
Blood Collection (Diabetic Patients)
n=10 Participants
Participants's blood was collected and measured for glucose level
Total
n=210 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=200 Participants
0 Participants
n=10 Participants
0 Participants
n=210 Participants
Age, Categorical
Between 18 and 65 years
188 Participants
n=200 Participants
7 Participants
n=10 Participants
195 Participants
n=210 Participants
Age, Categorical
>=65 years
12 Participants
n=200 Participants
3 Participants
n=10 Participants
15 Participants
n=210 Participants
Sex: Female, Male
Female
114 Participants
n=200 Participants
7 Participants
n=10 Participants
121 Participants
n=210 Participants
Sex: Female, Male
Male
86 Participants
n=200 Participants
3 Participants
n=10 Participants
89 Participants
n=210 Participants
Race and Ethnicity Not Collected
0 Participants
Race and Ethnicity were not collected from any participant.
Region of Enrollment
Israel
200 participants
n=200 Participants
10 participants
n=10 Participants
210 participants
n=210 Participants

PRIMARY outcome

Timeframe: 1-3 hours

Difference in measurements of blood glucose level obtained using Spectrophon glucometry monitor and commercially available glucose analyser YSI 2300

Outcome measures

Outcome measures
Measure
Blood Collection (Healthy Individuals)
n=200 Participants
Participants's blood was collected and measured for glucose level before and after food intake
Blood Collection (Diabetic Patients)
n=10 Participants
Participants's blood was collected and measured for glucose level
Percentage of Measurements With < 15% Error
Algorithm 1.7
95.25 Percentage of measures with error <15%
100 Percentage of measures with error <15%
Percentage of Measurements With < 15% Error
Algorithm 1.6
94.5 Percentage of measures with error <15%
100 Percentage of measures with error <15%

Adverse Events

Blood Collection (Healthy Individuals)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Blood Collection (Diabetic Patients)

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Anatoly Kreinin

Maale Carmel Mental Health Center

Phone: (+972) -4-8559315

Results disclosure agreements

  • Principal investigator is a sponsor employee