Trial Outcomes & Findings for Measurement of Refractive SE Following Implantation of LS313 MF30 IOL (NCT NCT03292614)

NCT ID: NCT03292614

Last Updated: 2019-09-26

Results Overview

After the implantation of the LS 313 MF30 (a refractive, asymmetrical, multifocal intraocular lens), the most reliable refraction method so far has been the subjective determination of far and near refraction. A faster and widespread method is the measurement with auto-refractometers (KR-1W (TOPCON), KR-800s (TOPCON), ARK-560 (NIDEK), ARK-760A (NIDEK)) or manual refractometers (PR60/50 (Rodenstock)). KR-1W, KR-800s, ARK-560A, ARK-760A only allow a central measurement without separate measurement of the two optical zones. The manual refractometer PR60/50 (Rodenstock) is able to separately measure the different focal points of the multifocal lens using a light cone with 3 mm diameter. The test device iTrace (TraceyTechnologies): is also able to separately measure the different focal points of the multifocal lens, but uses a light cone with a diameter of about 1 mm. The results of all refractometers are compared with the subjective refraction.

Recruitment status

COMPLETED

Target enrollment

25 participants

Primary outcome timeframe

one year

Results posted on

2019-09-26

Participant Flow

Participant milestones

Participant milestones
Measure
Multifocal Intraocular Lens Implantation
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
Overall Study
STARTED
25
Overall Study
COMPLETED
25
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Multifocal Intraocular Lens Implantation
n=25 Participants
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
Age, Categorical
<=18 years
0 Participants
n=25 Participants
Age, Categorical
Between 18 and 65 years
5 Participants
n=25 Participants
Age, Categorical
>=65 years
20 Participants
n=25 Participants
Age, Continuous
69.84 years
STANDARD_DEVIATION 8.19 • n=25 Participants
Sex: Female, Male
Female
11 Participants
n=25 Participants
Sex: Female, Male
Male
14 Participants
n=25 Participants
Region of Enrollment
Germany
25 participants
n=25 Participants

PRIMARY outcome

Timeframe: one year

After the implantation of the LS 313 MF30 (a refractive, asymmetrical, multifocal intraocular lens), the most reliable refraction method so far has been the subjective determination of far and near refraction. A faster and widespread method is the measurement with auto-refractometers (KR-1W (TOPCON), KR-800s (TOPCON), ARK-560 (NIDEK), ARK-760A (NIDEK)) or manual refractometers (PR60/50 (Rodenstock)). KR-1W, KR-800s, ARK-560A, ARK-760A only allow a central measurement without separate measurement of the two optical zones. The manual refractometer PR60/50 (Rodenstock) is able to separately measure the different focal points of the multifocal lens using a light cone with 3 mm diameter. The test device iTrace (TraceyTechnologies): is also able to separately measure the different focal points of the multifocal lens, but uses a light cone with a diameter of about 1 mm. The results of all refractometers are compared with the subjective refraction.

Outcome measures

Outcome measures
Measure
Multifocal Intraocular Lens Implantation
n=25 Participants
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
subjective Refraction
0.36 diopter
Standard Deviation 0.65
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
iTrace
0.27 diopter
Standard Deviation 0.64
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
KR-W1
-0.81 diopter
Standard Deviation 0.59
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
KR-800s
-0.78 diopter
Standard Deviation 0.73
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
ARK-760A
-0.73 diopter
Standard Deviation 0.59
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
ARK-560A
-0.76 diopter
Standard Deviation 0.63
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
PR60/50
-0.18 diopter
Standard Deviation 0.43

Adverse Events

Multifocal Intraocular Lens Implantation

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported