Trial Outcomes & Findings for Measurement of Refractive SE Following Implantation of LS313 MF30 IOL (NCT NCT03292614)
NCT ID: NCT03292614
Last Updated: 2019-09-26
Results Overview
After the implantation of the LS 313 MF30 (a refractive, asymmetrical, multifocal intraocular lens), the most reliable refraction method so far has been the subjective determination of far and near refraction. A faster and widespread method is the measurement with auto-refractometers (KR-1W (TOPCON), KR-800s (TOPCON), ARK-560 (NIDEK), ARK-760A (NIDEK)) or manual refractometers (PR60/50 (Rodenstock)). KR-1W, KR-800s, ARK-560A, ARK-760A only allow a central measurement without separate measurement of the two optical zones. The manual refractometer PR60/50 (Rodenstock) is able to separately measure the different focal points of the multifocal lens using a light cone with 3 mm diameter. The test device iTrace (TraceyTechnologies): is also able to separately measure the different focal points of the multifocal lens, but uses a light cone with a diameter of about 1 mm. The results of all refractometers are compared with the subjective refraction.
COMPLETED
25 participants
one year
2019-09-26
Participant Flow
Participant milestones
| Measure |
Multifocal Intraocular Lens Implantation
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
|
|---|---|
|
Overall Study
STARTED
|
25
|
|
Overall Study
COMPLETED
|
25
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Multifocal Intraocular Lens Implantation
n=25 Participants
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=25 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
5 Participants
n=25 Participants
|
|
Age, Categorical
>=65 years
|
20 Participants
n=25 Participants
|
|
Age, Continuous
|
69.84 years
STANDARD_DEVIATION 8.19 • n=25 Participants
|
|
Sex: Female, Male
Female
|
11 Participants
n=25 Participants
|
|
Sex: Female, Male
Male
|
14 Participants
n=25 Participants
|
|
Region of Enrollment
Germany
|
25 participants
n=25 Participants
|
PRIMARY outcome
Timeframe: one yearAfter the implantation of the LS 313 MF30 (a refractive, asymmetrical, multifocal intraocular lens), the most reliable refraction method so far has been the subjective determination of far and near refraction. A faster and widespread method is the measurement with auto-refractometers (KR-1W (TOPCON), KR-800s (TOPCON), ARK-560 (NIDEK), ARK-760A (NIDEK)) or manual refractometers (PR60/50 (Rodenstock)). KR-1W, KR-800s, ARK-560A, ARK-760A only allow a central measurement without separate measurement of the two optical zones. The manual refractometer PR60/50 (Rodenstock) is able to separately measure the different focal points of the multifocal lens using a light cone with 3 mm diameter. The test device iTrace (TraceyTechnologies): is also able to separately measure the different focal points of the multifocal lens, but uses a light cone with a diameter of about 1 mm. The results of all refractometers are compared with the subjective refraction.
Outcome measures
| Measure |
Multifocal Intraocular Lens Implantation
n=25 Participants
Patients who have undergone a cataract surgery with implantation of a multifocal intraocular lens LS313 MF30, Oculentis Berlin, will be invited for Measurements of the postoperative refraction
|
|---|---|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
subjective Refraction
|
0.36 diopter
Standard Deviation 0.65
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
iTrace
|
0.27 diopter
Standard Deviation 0.64
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
KR-W1
|
-0.81 diopter
Standard Deviation 0.59
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
KR-800s
|
-0.78 diopter
Standard Deviation 0.73
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
ARK-760A
|
-0.73 diopter
Standard Deviation 0.59
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
ARK-560A
|
-0.76 diopter
Standard Deviation 0.63
|
|
Evaluation of the Most Precise Method for Measuring Post-operative Refraction Following Implantation of the Asymmetric Multifocal LS-313 MF30
PR60/50
|
-0.18 diopter
Standard Deviation 0.43
|
Adverse Events
Multifocal Intraocular Lens Implantation
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place