Trial Outcomes & Findings for Evaluation of the Peel Force of New Adhesives From the Skin (NCT NCT03200444)

NCT ID: NCT03200444

Last Updated: 2025-04-25

Results Overview

The peel force (N) needed to peel off adhesive strips from the skin

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

6 participants

Primary outcome timeframe

30 min

Results posted on

2025-04-25

Participant Flow

Unit of analysis: Number of strips

Participant milestones

Participant milestones
Measure
Standard Adhesive 1
This strip consists of a standard hydrocolloid adhesive found in ostomy products
Standard Adhesive 2
This strip consists of a standard hydrocolloid adhesive found in ostomy products
Standard Adhesive 3
This strip consists of a standard hydrocolloid adhesive found in ostomy products
P-4 Adhesive
Newly developed adhesive
P-15 Adhesive
Newly developed adhesive
P-16 Adhesive
Newly developed adhesive
Overall Study
STARTED
6 6
6 6
6 6
6 6
6 6
6 6
Overall Study
COMPLETED
6 6
6 6
6 6
6 6
6 6
6 6
Overall Study
NOT COMPLETED
0 0
0 0
0 0
0 0
0 0
0 0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Evaluation of the Peel Force of New Adhesives From the Skin

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Testing of 6 Adhesive Strips
n=6 Participants
Each subject tests six adhesive strips on pre-striped skin. 1. Standard adhesive 1 2. standard adhesive 2 3. Standard adhesive 3 4. P-4 5. P-15 6. P-16 The six strips are applied on abdominal skin. The order to which the adhesive strips are located on the skin is randomized. The subjects will change the adhesive strips at home and the adhesion of the adhesive strips will be measured at 5 visits. Standard adhesive 1: This strip consists of a standard hydrocolloid adhesive found in ostomy products Standard adhesive 2: This strip consists of a standard hydrocolloid adhesive found in ostomy products Standard adhesive 3: This strip consists of a standard hydrocolloid adhesive found in ostomy products P-4: Newly developed adhesive P-4 P-15: Newly developed adhesive P-14 P-16: Newly developed adhesive P-8
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
6 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Sex: Female, Male
Female
3 Participants
n=5 Participants
Sex: Female, Male
Male
3 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
6 Participants
n=5 Participants
Region of Enrollment
Denmark
6 participants
n=5 Participants

PRIMARY outcome

Timeframe: 30 min

Population: All participants were analysed

The peel force (N) needed to peel off adhesive strips from the skin

Outcome measures

Outcome measures
Measure
Standard Adhesive 1
n=6 Participants
This strip consists of a standard hydrocolloid adhesive found in ostomy products
Standard Adhesive 2
n=6 Participants
This strip consists of a standard hydrocolloid adhesive found in ostomy products
Standard Adhesive 3
n=6 Participants
This strip consists of a standard hydrocolloid adhesive found in ostomy products
P-4 Adhesive
n=6 Participants
Newly developed adhesive
P-15 Adhesive
n=6 Participants
Newly developed adhesive
P-16 Adhesive
n=6 Participants
Newly developed adhesive
Peel Force
5.2 Newton
Interval 3.3 to 5.4
3 Newton
Interval 2.3 to 4.8
2.1 Newton
Interval 1.6 to 4.1
3.3 Newton
Interval 3.3 to 4.5
3.7 Newton
Interval 2.2 to 5.0
4.4 Newton
Interval 2.1 to 5.1

Adverse Events

Standard Adhesive 1

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Standard Adhesive 2

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Standard Adhesive 3

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

P-4 Adhesive

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

P-15 Adhesive

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

P-16 Adhesive

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Director of Pre-Clinical Research

Coloplast A/S

Phone: 49111111

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place