Alarms Due to Loss of Signal Integrity in Partially Wireless Versus Traditional Pulse Oximetry in Pediatric Patients
NCT ID: NCT03187665
Last Updated: 2017-11-24
Study Results
The study team has not published outcome measurements, participant flow, or safety data for this trial yet. Check back later for updates.
Basic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
36 participants
OBSERVATIONAL
2017-07-26
2017-11-01
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
This study will help the investigators learn more about whether a partially wireless, more portable pulse oximeter that connects to a small device worn on the child's arm or leg will give the investigators a more reliable signal/reading while letting children move more easily. The investigators will compare this device with the traditional wall-connected unit. Movement of the cable or a child moving may give a false oxygen reading. The investigators will ask the child to do activities that create movement and will look at the readings when the child moves. The investigators think the partially wireless pulse oximeter will be more reliable during movement than the traditional wall-connected unit.
Subjects will have two continuous pulse oximeter probes placed on them. These soft probes will go on a finger, toe, foot or hand and will be attached to two different pulse oximeter monitors. The child will then be asked to do common childhood activities based on their age for about 20 minutes. These activities will be play activities the child already does such as grabbing a toy, drawing with crayons or kicking a ball. Continuous pulse oximetry data will be recorded during the testing and will be stored in a way that it cannot be linked to the subject after the testing is complete.
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
HYPOTHESIS Continuously monitoring pulse oximetry with a partially wireless monitor without physical connection to a wall unit will decrease episodes of loss of signal integrity generated by motion artifact in healthy pediatric subjects completing age appropriate activities.
BACKGROUND, SIGNIFICANCE, AND RATIONALE Pulse oximetry is used in both inpatient and outpatient environments to measure the oxygen saturation in a person's blood and the pulse, or heart rate. Pulse oximeters use a painless Band-Aid-like probe that attaches to a fingertip, toe, hand or foot and shines a light through the skin. The monitor then determines how much oxygen is in the blood and the pulse rate based on the way the light passes through the skin. The probe is connected to the monitor by a cable of varying lengths. Movement can cause incorrect pulse oximeter measurements.
Continuous pulse oximetry is considered a standard of care in pediatric intensive care units; however pediatric patients have a high incidence of false alarms in part due to motion artifact. These false alarms contribute to the 94% false alarm incidence in pediatric ICU patients. (Lawless) Motion artifact is potentially exacerbated by the long cables connecting the patients to the continuous monitors. Currently the FDA has not approved any completely cable-free pulse oximeters for continuous use; however Masimo has produced a device (the Radius 7) that overcomes the previously required cable connection to a wall unit via wireless transmission from a patient-worn continuous monitor. (Masimo) There are currently no studies looking at this patient worn monitor in relation to the motion artifacts that frequently cause false alarms in pediatric patients.
The Joint Commission has listed alarm fatigue as a top patient safety concern and as of January 1, 2016 has required hospitals to establish policies and procedures for managing alarm fatigue. Reduction in false alarms is necessary to reduce alarm fatigue. This pilot study will use clinical data to measure the effect of eliminating this cable-to-wall connection on motion artifact. Additionally, early mobility in critically ill patients has been shown to improve outcomes. Freeing patients from one of the cables currently tying them to the room may decrease the false alarms from motion artifact while simultaneously improving mobility, therefore improving patient outcomes and safety.
DESIGN AND METHODS Subjects: Healthy pediatric subjects ranging from 6 months to 17 years of age. Pediatric subjects will be recruited via fliers placed in pediatric critical care areas, in the Translational Research Control Unit (TRU) and on the MCW research bulletin board outside the MCW office of research. Subjects may include those responding to the fliers directly as well as the healthy siblings of patients being evaluated or treated in the Children's Hospital of Wisconsin Emergency Room (EDTC), Clinics or Hospital.
Setting: The Children's Hospital of Wisconsin (CHW): primarily in the TRU designed to mimic a CHW outpatient clinic room; and also including available sites near the EDTC or clinic to facilitate making testing more convenient for the subject's family if the subject is a healthy sibling of a patient being evaluated in the EDTC or clinic.
Procedures:
While being continuously monitored on the current equipment used by CHW (the wall-connected Radical 7) and the partially wireless Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities as developed by CHW physical therapists, for approximately twenty minutes. These activities include common childhood activities such as kicking a ball or drawing with crayons. If a subject declines to participate in an activity, an alternate will be suggested. The investigators will analyze the pulse oximetry data for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%. (Seixas, Seixas and Pereira)
Endpoints: Enrollment of and data acquisition on 48 pediatric subjects: 12 in each of the following developmental age groups: 6 month-1year, 1-6 years, 6-10 years, and 11-17 years.
Data analysis:
All data will be collected in a de-identified manner in which subjects will be given a study case number and all data will be collected under that case number.
The number of episodes of loss of signal integrity leading to an alarm state generated by the wall-connected Radical 7 unit compared to the partially wireless Radius 7 unit during completion of the developmentally appropriate activities will be compared. Analysis will be performed by the Quantitative Health Sciences department at MCW and a 20% reduction in episodes of loss of signal integrity will be considered clinically significant.
TOTAL NUMBER OF HUMAN RESEARCH PARTICIPANTS PROPOSED FOR THIS STUDY. 48 pediatric subjects: 12 in each of the following developmental age groups: 6 months-1year, 1-6 years, 6-10 years, and 11-17 years. This number of participants will provide representative data for a pilot study investigating the most common age groups seen in the pediatric ICU in order to identify the age groups most likely to benefit from further study.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Keywords
Explore important study keywords that can help with search, categorization, and topic discovery.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
CASE_ONLY
PROSPECTIVE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
1-6 years
12 subjects in the age range of 1-6 years old.
Radical 7
The Radical 7 is the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a 12 foot cable connecting the patient to a wall unit. While being continuously monitored with the Radical 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Radius 7
The Radius 7 is a newer, partially wireless pulse oximeter which eliminates the cable connecting the patient to a wall unit by instead connecting to a small device worn on the patient's arm or leg. The small device then sends the continuous pulse oximeter data wirelessly to the central monitoring station. While being continuously monitored with the Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
6-10 years
12 subjects in the age range of 6-10 years old.
Radical 7
The Radical 7 is the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a 12 foot cable connecting the patient to a wall unit. While being continuously monitored with the Radical 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Radius 7
The Radius 7 is a newer, partially wireless pulse oximeter which eliminates the cable connecting the patient to a wall unit by instead connecting to a small device worn on the patient's arm or leg. The small device then sends the continuous pulse oximeter data wirelessly to the central monitoring station. While being continuously monitored with the Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
11-17 years
12 subjects in the age range of 11-17 years old.
Radical 7
The Radical 7 is the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a 12 foot cable connecting the patient to a wall unit. While being continuously monitored with the Radical 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Radius 7
The Radius 7 is a newer, partially wireless pulse oximeter which eliminates the cable connecting the patient to a wall unit by instead connecting to a small device worn on the patient's arm or leg. The small device then sends the continuous pulse oximeter data wirelessly to the central monitoring station. While being continuously monitored with the Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
Radical 7
The Radical 7 is the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a 12 foot cable connecting the patient to a wall unit. While being continuously monitored with the Radical 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Radius 7
The Radius 7 is a newer, partially wireless pulse oximeter which eliminates the cable connecting the patient to a wall unit by instead connecting to a small device worn on the patient's arm or leg. The small device then sends the continuous pulse oximeter data wirelessly to the central monitoring station. While being continuously monitored with the Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.
Other Intervention Names
Discover alternative or legacy names that may be used to describe the listed interventions across different sources.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
Exclusion Criteria
6 Months
17 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
Pamela Petersen
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Pamela Petersen
Pediatric Critical Care Fellow
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Pamela C Petersen, M.D.
Role: PRINCIPAL_INVESTIGATOR
Medical College of Wisconsin Affiliated Hospitals, Inc.
Sheila Hanson, M.D.
Role: PRINCIPAL_INVESTIGATOR
Medical College of Wisconsin Affiliated Hospitals, Inc.
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Children's Hospital of Wisconsin
Milwaukee, Wisconsin, United States
Countries
Review the countries where the study has at least one active or historical site.
References
Explore related publications, articles, or registry entries linked to this study.
Lawless ST. Crying wolf: false alarms in a pediatric intensive care unit. Crit Care Med. 1994 Jun;22(6):981-5.
Seixas DM, Seixas DM, Pereira MC, Moreira MM, Paschoal IA. Oxygen desaturation in healthy subjects undergoing the incremental shuttle walk test. J Bras Pneumol. 2013 Jun-Aug;39(4):440-6. doi: 10.1590/S1806-37132013000400007.