Digital Ankle Brachial Index (ABI) as a Screening Tool in Detecting Peripheral Arterial Disease
NCT ID: NCT03161327
Last Updated: 2018-12-11
Study Results
Outcome measurements, participant flow, baseline characteristics, and adverse events have been published for this study.
View full resultsBasic Information
Get a concise snapshot of the trial, including recruitment status, study phase, enrollment targets, and key timeline milestones.
COMPLETED
NA
238 participants
INTERVENTIONAL
2017-07-15
2017-09-14
Brief Summary
Review the sponsor-provided synopsis that highlights what the study is about and why it is being conducted.
QuantaFlo™ (Semler Scientific, Inc.) is a novel, noninvasive, 510K FDA approved digital device that is used as a screening tool to measure ABI of patients at risk of PAD. This single center prospective clinical trial will evaluate the sensitivity and specificity of digital ABI in detecting PAD using color Doppler ultrasound and 'gold standard' angiography as reference. Specifically in patient undergoing dialysis and who cannot undergo ABI using pressure cuffs dialysis grafts/ fistulae in the arms, we will evaluate the value of digital ABI in detecting PAD.
Detailed Description
Dive into the extended narrative that explains the scientific background, objectives, and procedures in greater depth.
Early detection is critical for mitigating PAD progression. Ankle-brachial index (ABI) testing is recommended by the US Preventative Services Task Force as an affordable and effective screening tool for evaluating PAD risk. ABI testing is especially useful in primary care settings allowing for earlier diagnosis and treatment. Several ABI methods are used; classically this test uses a device for measuring blood pressure with an inflatable cuff, and blood pressure measurements are taken at the upper arm and the ankle; healthcare providers have used a doppler probe to detect audible systolic pressure signals within the arteries - Doppler ABI. Barriers to implementation include; time required to conduct tests, training and required technical skills. Hence, Doppler-based ABI in primary care has been limited. Recently designed oscillometric (manual or digital with automatic electronic calculation of blood pressure) and photophlethysmographic devices are now available. However current guidelines do not endorse the use of these newer devices but recommend the hand-held doppler technique.
Digital ABI is noninvasive, automated ABI using pulsatile blood volume changes - plethysmography, irrespective of vessel compressibility. The QuantaFlo™ (Semler Scientific, Inc.) is an innovative, 510K FDA approved device that provides bilateral, non-invasive physiologic studies of upper and lower extremity arteries using volume plethysmography of the posterior tibial and anterior tibial/dorsalis pedis arterial distributions. QuantaFlo™ uses an optical sensor attached onto a digit in the hand and feet to assess extremity blood volume. It then displays a signal, directly related to blood volume, on a Volume Plethysmography Chart used to calculate PAD probability. This test is easy to use and portable, delivering fast, accurate results in \~5 minutes; an attractive option in primary care settings. While maintaining accuracy, results comparing digital methods to Doppler show concordance above 90%. The American Medical Association (AMA) concluded, "Evaluation of the digits in both legs with volume plethysmography is appropriate for this code provided that an ankle/brachial index (ABI) be performed." QuantaFlo™ meets both of these standards.
ABI ≤ 0.9 has been used as the criterion for diagnosing PAD based on ACC/AHA guidelines for the management of patients with PAD ; the cutoff value of ABPI ≤0.9 it has been reported that the sensitivity of diagnosing PAD was 83% to 85% and a specificity of 95% to 100% \[5\]. ABI has been validated against lower extremity contrast angiography obtaining high sensitivity (85%-95%) and specificity (90%-100%), for diagnosing PAD . Many studies have reported more than 95% of sensitivity and specificity of colour duplex ultrasound scan in detecting a significant stenosis of the arteries .
QuantaFlo™ (Semler Scientific, Inc.) is a novel, noninvasive, 510K FDA approved digital device that is used as a screening tool to measure ABI of patients at risk of PAD. This single center prospective clinical trial will evaluate the sensitivity and specificity of digital ABI in detecting PAD using color Doppler ultrasound and 'gold standard' angiography as reference. Specifically in patient undergoing dialysis and who cannot undergo ABI using pressure cuffs dialysis grafts/ fistulae in the arms, we will evaluate the value of digital ABI in detecting PAD.
Conditions
See the medical conditions and disease areas that this research is targeting or investigating.
Keywords
Explore important study keywords that can help with search, categorization, and topic discovery.
Study Design
Understand how the trial is structured, including allocation methods, masking strategies, primary purpose, and other design elements.
NA
SINGLE_GROUP
PREVENTION
NONE
Study Groups
Review each arm or cohort in the study, along with the interventions and objectives associated with them.
ABI
ABI will be performed in patient
ABI using QuantaFlo™
Perform Digital ABI:
* Will be performed by the nurse/medical assistant with training to perform this test
* The patient will be placed in the supine position, with the arms and legs at the same level as the heart, for a minimum of 5 minutes before measurement.
* The optical sensor (similar to a pulse oximeter) will be placed sequentially in the fingers of the hands and feet: Right and left hand fingers and then right and left feet toes.
* Usually each digit takes about 15 seconds to obtain a waveform
* At the end of 60 seconds, an automated digital ABI will be generated.
Clinical course: If based on Screening questionnaire and digital ABI, I. Patient is diagnosed with PAD- the referring physician and patient will receive a phone call about the same. He/ she will then be advised any further testing/ referral to IR clinic.
II. There is no evidence of PAD- no further action will be taken. The patient will be informed of the same.
Interventions
Learn about the drugs, procedures, or behavioral strategies being tested and how they are applied within this trial.
ABI using QuantaFlo™
Perform Digital ABI:
* Will be performed by the nurse/medical assistant with training to perform this test
* The patient will be placed in the supine position, with the arms and legs at the same level as the heart, for a minimum of 5 minutes before measurement.
* The optical sensor (similar to a pulse oximeter) will be placed sequentially in the fingers of the hands and feet: Right and left hand fingers and then right and left feet toes.
* Usually each digit takes about 15 seconds to obtain a waveform
* At the end of 60 seconds, an automated digital ABI will be generated.
Clinical course: If based on Screening questionnaire and digital ABI, I. Patient is diagnosed with PAD- the referring physician and patient will receive a phone call about the same. He/ she will then be advised any further testing/ referral to IR clinic.
II. There is no evidence of PAD- no further action will be taken. The patient will be informed of the same.
Eligibility Criteria
Check the participation requirements, including inclusion and exclusion rules, age limits, and whether healthy volunteers are accepted.
Inclusion Criteria
* Are you 65 years or older?
* Have you ever smoked?
* Do you have diabetes?
* Do you have high blood pressure or take medication for high blood pressure?
* Do you have high cholesterol, or take medication for high cholesterol?
* Have you previously had a stroke?
* Do you have heart disease?
* Do you experience any pain at rest in your lower legs or feet?
* Do you have discomfort (aching, fatigue, tingling, cramping or pain) when you walk, which is relieved by rest? Does the pain go away within 10 minutes after stopping?
* Are your toes or feet pale, discolored, or bluish?
* Do you have an infection, skin wound, or ulcer on your feet or toes?
Exclusion Criteria
* Previously screened with ABI in past 12 months
18 Years
90 Years
ALL
Yes
Sponsors
Meet the organizations funding or collaborating on the study and learn about their roles.
St. Louis University
OTHER
Responsible Party
Identify the individual or organization who holds primary responsibility for the study information submitted to regulators.
Keith Pereira, MD:
Assistant Professor
Principal Investigators
Learn about the lead researchers overseeing the trial and their institutional affiliations.
Keith Pereira, MD
Role: PRINCIPAL_INVESTIGATOR
Saint Louis University Hospital
Locations
Explore where the study is taking place and check the recruitment status at each participating site.
Saint Louis University
St Louis, Missouri, United States
Countries
Review the countries where the study has at least one active or historical site.
Provided Documents
Download supplemental materials such as informed consent forms, study protocols, or participant manuals.
Document Type: Study Protocol and Statistical Analysis Plan
Other Identifiers
Review additional registry numbers or institutional identifiers associated with this trial.
28225
Identifier Type: -
Identifier Source: org_study_id