Trial Outcomes & Findings for Accuracy of Noninvasive Rainbow Adhesive R1 Pulse Oximeter Sensor (NCT NCT03134313)
NCT ID: NCT03134313
Last Updated: 2017-07-02
Results Overview
Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
COMPLETED
NA
25 participants
1-5 hours per subject
2017-07-02
Participant Flow
Participant milestones
| Measure |
R1-25 Sensor
All subjects will be enrolled in the test group and will receive Rainbow Adhesive Noninvasive R1 Pulse Oximeter Sensor
Rainbow Adhesive Noninvasive R1 Pulse Oximeter Sensor
|
|---|---|
|
Overall Study
STARTED
|
25
|
|
Overall Study
COMPLETED
|
25
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy of Noninvasive Rainbow Adhesive R1 Pulse Oximeter Sensor
Baseline characteristics by cohort
| Measure |
R1-25 Sensor
n=25 Participants
All subjects will be enrolled in the test group and will receive Rainbow Adhesive Noninvasive R1 Pulse Oximeter Sensor
Rainbow Adhesive Noninvasive R1 Pulse Oximeter Sensor
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
25 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
7 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
18 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
25 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1-5 hours per subjectAccuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. In order to obtain the Arms value, the blood sample hemoglobin value is subtracted from the pulse oximeter hemoglobin value for a number of samples, the average of this difference is computed as the bias. The standard deviation of the differences is computed as the precision. The square root of the sum of the squares of bias and precision is computed as the Arms Error value.
Outcome measures
| Measure |
R1-25 Sensor
n=25 Participants
All subjects will be enrolled in the test group and will receive Rainbow Adhesive Noninvasive R1 Pulse Oximeter Sensor
|
|---|---|
|
Accuracy of SpHb on R1-25 by Arms Calculation
|
1.0 g/dL
|
Adverse Events
R1-25 Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: LTE60