Trial Outcomes & Findings for Oxygen Reserve Index Validation for RD Lite Sensors (NCT NCT03128892)

NCT ID: NCT03128892

Last Updated: 2021-03-01

Results Overview

A measure of the ORI parameter's ability to detect changes in dissolved oxygen levels in blood plasma. Specificity, as measured in this outcome, is the probability of ORI not dropping when there is no drop in blood oxygen levels (desaturation).

Recruitment status

COMPLETED

Target enrollment

34 participants

Primary outcome timeframe

1-5 hours per subject

Results posted on

2021-03-01

Participant Flow

Participant milestones

Participant milestones
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Overall Study
STARTED
34
Overall Study
COMPLETED
31
Overall Study
NOT COMPLETED
3

Reasons for withdrawal

Reasons for withdrawal
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Overall Study
Withdrawal by Subject
2
Overall Study
Physician Decision
1

Baseline Characteristics

Oxygen Reserve Index Validation for RD Lite Sensors

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
n=31 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
31 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Sex: Female, Male
Female
9 Participants
n=5 Participants
Sex: Female, Male
Male
22 Participants
n=5 Participants
Race/Ethnicity, Customized
African American
12 Participants
n=5 Participants
Race/Ethnicity, Customized
Asian or Pacific Islander
1 Participants
n=5 Participants
Race/Ethnicity, Customized
Caucasian
5 Participants
n=5 Participants
Race/Ethnicity, Customized
Hispanic
13 Participants
n=5 Participants
Region of Enrollment
United States
31 participants
n=5 Participants

PRIMARY outcome

Timeframe: 1-5 hours per subject

A measure of the ORI parameter's ability to detect changes in dissolved oxygen levels in blood plasma. Specificity, as measured in this outcome, is the probability of ORI not dropping when there is no drop in blood oxygen levels (desaturation).

Outcome measures

Outcome measures
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
n=31 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 110 mmHg
98.50 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 120 mmHg
96.68 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 130 mmHg
98.89 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 140 mmHg
96.95 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 150 mmHg
97.67 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 160 mmHg
94.75 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 170 mmHg
93.60 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 180 mmHg
96.23 percentage of true negative
Specificity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 190 mmHg
95.82 percentage of true negative

PRIMARY outcome

Timeframe: 1-5 hours

A measure of the ORI parameter's ability to detect changes in dissolved oxygen levels in blood plasma. Sensitivity, as measured in this outcome, is the probability of ORI dropping when there is a drop in blood oxygenation levels (desaturation).

Outcome measures

Outcome measures
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
n=31 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 190 mmHg
88.27 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 110 mmHg
84.96 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 120 mmHg
89.85 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 130 mmHg
85.51 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 140 mmHg
89.47 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 150 mmHg
86.17 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 160 mmHg
87.13 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 170 mmHg
83.96 percentage of true positives
Sensitivity of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 180 mmHg
85.40 percentage of true positives

PRIMARY outcome

Timeframe: 1-5 hours

A measure of the ORI parameter's ability to detect changes in dissolved oxygen levels in blood plasma. Concordance, as measured in this outcome, is the probability of ORI to respond accurately to changes (or lack of thereof) in underlying blood oxygenation levels. It is computed as a ratio of true classifications (true positives and true negatives) to the sample size. An ideal detector will have a concordance value of 1 (or 100%).

Outcome measures

Outcome measures
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
n=31 Participants
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 110 mmHg
88.83 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 120 mmHg
92.46 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 130 mmHg
91.47 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 140 mmHg
93.33 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 150 mmHg
93.00 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 160 mmHg
92.16 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 170 mmHg
90.80 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 180 mmHg
93.78 percentage of accuracy
Concordance of ORI to Detect Drops in Oxygenation
PaO2 Threshold: 190 mmHg
94.39 percentage of accuracy

Adverse Events

Oxygen Reserve Index (ORi) - RD Lite Sensors

Serious events: 0 serious events
Other events: 1 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Oxygen Reserve Index (ORi) - RD Lite Sensors
n=34 participants at risk
All subjects who are enrolled into the test group and participate in data collection receive the noninvasive Oxygen Reserve Index - RD Lite Sensors.
Surgical and medical procedures
Arterial line site sensitivity
2.9%
1/34 • 1 to 5 hours

Additional Information

Chelsea Frank

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place