Trial Outcomes & Findings for Comparison of Noninvasive Hemoglobin Disposable Sensors (NCT NCT03125031)
NCT ID: NCT03125031
Last Updated: 2017-07-27
Results Overview
Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. The accuracy results from both sensors will be assessed for equivalence
COMPLETED
NA
44 participants
1-5 hours
2017-07-27
Participant Flow
Participant milestones
| Measure |
Rainbow Adhesive Adult/Pediatric Sensor
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/pediatric sensors.
|
Rainbow Adhesive Adult/Neonatal Sensor
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/neonatal sensors.
|
|---|---|---|
|
Overall Study
STARTED
|
44
|
44
|
|
Overall Study
COMPLETED
|
44
|
44
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Comparison of Noninvasive Hemoglobin Disposable Sensors
Baseline characteristics by cohort
| Measure |
Rainbow Adhesive Adult/Pediatric and Adult/Neonatal Sensors
n=44 Participants
All subjects are enrolled into the test group and receive the Rainbow adhesive sensors (adult/pediatric and adult/neonatal sensors).
|
|---|---|
|
Age, Customized
18-38 years
|
44 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
12 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
32 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1-5 hoursAccuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. The accuracy results from both sensors will be assessed for equivalence
Outcome measures
| Measure |
Group-Sensor 1
n=691 data points
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/pediatric sensors.
|
Group-Sensor 2
n=928 data points
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/neonatal sensors.
|
|---|---|---|
|
Accuracy of Noninvasive Sensors by Arms Calculation
|
1.0 g/dL
|
1.0 g/dL
|
Adverse Events
Rainbow Adhesive Adult/Pediatric and Adult/Neonatal Sensors
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place