Trial Outcomes & Findings for Comparison of Noninvasive Hemoglobin Disposable Sensors (NCT NCT03125031)

NCT ID: NCT03125031

Last Updated: 2017-07-27

Results Overview

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. The accuracy results from both sensors will be assessed for equivalence

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

44 participants

Primary outcome timeframe

1-5 hours

Results posted on

2017-07-27

Participant Flow

Participant milestones

Participant milestones
Measure
Rainbow Adhesive Adult/Pediatric Sensor
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/pediatric sensors.
Rainbow Adhesive Adult/Neonatal Sensor
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/neonatal sensors.
Overall Study
STARTED
44
44
Overall Study
COMPLETED
44
44
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Comparison of Noninvasive Hemoglobin Disposable Sensors

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Rainbow Adhesive Adult/Pediatric and Adult/Neonatal Sensors
n=44 Participants
All subjects are enrolled into the test group and receive the Rainbow adhesive sensors (adult/pediatric and adult/neonatal sensors).
Age, Customized
18-38 years
44 Participants
n=5 Participants
Sex: Female, Male
Female
12 Participants
n=5 Participants
Sex: Female, Male
Male
32 Participants
n=5 Participants

PRIMARY outcome

Timeframe: 1-5 hours

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value. The accuracy results from both sensors will be assessed for equivalence

Outcome measures

Outcome measures
Measure
Group-Sensor 1
n=691 data points
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/pediatric sensors.
Group-Sensor 2
n=928 data points
All subjects are enrolled into the test group and receive the Rainbow adhesive adult/neonatal sensors.
Accuracy of Noninvasive Sensors by Arms Calculation
1.0 g/dL
1.0 g/dL

Adverse Events

Rainbow Adhesive Adult/Pediatric and Adult/Neonatal Sensors

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Tala Harake

Masimo

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place