Trial Outcomes & Findings for Accuracy of Methemoglobin Measurement for Rainbow Universal Pulse Oximeter Sensor (NCT NCT03125005)
NCT ID: NCT03125005
Last Updated: 2017-07-27
Results Overview
Accuracy willl be determined by comparing the noninvasive blood methemoglobin measurement (expressed as a percentage of total hemoglobin) of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.
COMPLETED
NA
51 participants
5 hours
2017-07-27
Participant Flow
Participant milestones
| Measure |
Rainbow Universal Pulse Oximeter Sensor
All subjects will be enrolled in the test group and will receive Rainbow Universal Pulse Oximeter Sensor.
Rainbow Universal Pulse Oximeter Sensor
|
|---|---|
|
Overall Study
STARTED
|
51
|
|
Overall Study
COMPLETED
|
51
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy of Methemoglobin Measurement for Rainbow Universal Pulse Oximeter Sensor
Baseline characteristics by cohort
| Measure |
Rainbow Universal Pulse Oximeter Sensor
n=51 Participants
All subjects will be enrolled in the test group and will receive Rainbow Universal Pulse Oximeter Sensor.
Rainbow Universal Pulse Oximeter Sensor
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
51 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
29 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
22 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
51 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 5 hoursAccuracy willl be determined by comparing the noninvasive blood methemoglobin measurement (expressed as a percentage of total hemoglobin) of the pulse oximeter to that obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.
Outcome measures
| Measure |
Rainbow Universal Pulse Oximeter Sensor
n=51 Participants
All subjects will be enrolled in the test group and will receive Rainbow Universal Pulse Oximeter Sensor.
Rainbow Universal Pulse Oximeter Sensor
|
|---|---|
|
Accuracy of Sensor by Arms Calculation
|
0.63 Percentage of total hemoglobin
|
Adverse Events
Rainbow Universal Pulse Oximeter Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: LTE60