Trial Outcomes & Findings for Accuracy of Hemoglobin Measurement for Various Rainbow Pulse Oximeter Sensors (NCT NCT03124966)

NCT ID: NCT03124966

Last Updated: 2017-07-27

Results Overview

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

118 participants

Primary outcome timeframe

1-3 hours

Results posted on

2017-07-27

Participant Flow

Participant milestones

Participant milestones
Measure
Noninvasive Hemoglobin Sensor
All subjects are enrolled into the test group and all will receive the pulse oximeter sensor.
Overall Study
STARTED
118
Overall Study
COMPLETED
118
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Noninvasive Hemoglobin Sensor
n=118 Participants
All subjects are enrolled into the test group and all will receive the pulse oximeter sensor.
Age, Categorical
<=18 years
0 Participants
n=118 Participants
Age, Categorical
Between 18 and 65 years
118 Participants
n=118 Participants
Age, Categorical
>=65 years
0 Participants
n=118 Participants
Sex: Female, Male
Female
40 Participants
n=118 Participants
Sex: Female, Male
Male
78 Participants
n=118 Participants
Region of Enrollment
United States
118 participants
n=118 Participants

PRIMARY outcome

Timeframe: 1-3 hours

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.

Outcome measures

Outcome measures
Measure
Noninvasive Hemoglobin Sensor
n=118 Participants
All subjects are enrolled into the test group and all will receive the pulse oximeter sensor.
Accuracy of Sensor by Arms Calculation
0.96 g/dL

Adverse Events

Noninvasive Hemoglobin Sensor

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Vikram Ramakanth

Masimo

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: LTE60