Trial Outcomes & Findings for Accuracy of Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI) (NCT NCT03124758)
NCT ID: NCT03124758
Last Updated: 2017-07-27
Results Overview
Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.
COMPLETED
NA
354 participants
1-5 hours
2017-07-27
Participant Flow
Participant milestones
| Measure |
Noninvasive Hemoglobin Sensor
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
|
|---|---|
|
Overall Study
STARTED
|
354
|
|
Overall Study
COMPLETED
|
354
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Accuracy of Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI)
Baseline characteristics by cohort
| Measure |
Noninvasive Hemoglobin Sensor
n=354 Participants
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
354 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
127 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
227 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1-5 hoursAccuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.
Outcome measures
| Measure |
Noninvasive Hemoglobin Sensor
n=1990 data points
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
|
|---|---|
|
Accuracy of Sensor by Arms Calculation
|
1.0 g/dL
|
Adverse Events
Noninvasive Hemoglobin Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place