Trial Outcomes & Findings for Accuracy of Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI) (NCT NCT03124758)

NCT ID: NCT03124758

Last Updated: 2017-07-27

Results Overview

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

354 participants

Primary outcome timeframe

1-5 hours

Results posted on

2017-07-27

Participant Flow

Participant milestones

Participant milestones
Measure
Noninvasive Hemoglobin Sensor
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
Overall Study
STARTED
354
Overall Study
COMPLETED
354
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Accuracy of Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI)

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Noninvasive Hemoglobin Sensor
n=354 Participants
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
354 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Sex: Female, Male
Female
127 Participants
n=5 Participants
Sex: Female, Male
Male
227 Participants
n=5 Participants

PRIMARY outcome

Timeframe: 1-5 hours

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square(Arms) error value.

Outcome measures

Outcome measures
Measure
Noninvasive Hemoglobin Sensor
n=1990 data points
All subjects are enrolled into the test group and all subjects received the Noninvasive Hemoglobin Sensor (Rainbow Reusable DCI, DCIP)
Accuracy of Sensor by Arms Calculation
1.0 g/dL

Adverse Events

Noninvasive Hemoglobin Sensor

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Tala Harake

Masimo

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place