Trial Outcomes & Findings for Accuracy of Noninvasive Hemoglobin Measurement (SpHb) Using Rainbow DCI Pulse Oximeter Sensor (NCT NCT03124693)

NCT ID: NCT03124693

Last Updated: 2017-06-12

Results Overview

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

90 participants

Primary outcome timeframe

Up to 24 hours

Results posted on

2017-06-12

Participant Flow

Participant milestones

Participant milestones
Measure
Test Group
All subjects are enrolled into the test group and all subjects received the Rainbow DCI pulse oximeter sensor. Rainbow DCI pulse oximeter sensor
Overall Study
STARTED
90
Overall Study
COMPLETED
90
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Test Group
n=90 Participants
All subjects are enrolled into the test group and all subjects received the Rainbow DCI pulse oximeter sensor. Rainbow DCI pulse oximeter sensor
Age, Categorical
<=18 years
0 Participants
n=90 Participants
Age, Categorical
Between 18 and 65 years
90 Participants
n=90 Participants
Age, Categorical
>=65 years
0 Participants
n=90 Participants
Sex: Female, Male
Female
18 Participants
n=90 Participants
Sex: Female, Male
Male
72 Participants
n=90 Participants
Region of Enrollment
United States
90 participants
n=90 Participants

PRIMARY outcome

Timeframe: Up to 24 hours

Accuracy will be determined by comparing the noninvasive hemoglobin measurement of the pulse oximeter to the hemoglobin value obtained from a blood sample and calculating the arithmetic root mean square (Arms) error value.

Outcome measures

Outcome measures
Measure
Test Group
n=90 Participants
All subjects are enrolled into the test group and all subjects received the Rainbow DCI pulse oximeter sensor.
Accuracy of Sensor by Arms Calculation
0.9 g/dL

Adverse Events

Test Group

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Vikram Ramakanth

Masimo

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: LTE60