Trial Outcomes & Findings for Oxygen Reserve Index: Utility as Early Warning for Desaturation in Morbidly Obese Patients (NCT NCT03021551)

NCT ID: NCT03021551

Last Updated: 2022-04-19

Results Overview

Evaluate the clinical utility of the change in the oxygen reserve index as an early warning of impending arterial oxygen desaturation in Obese patients (30 \< BMI \< 40 kg m-2) and Normal BMI patients (19 \< BMI \< 25 kg m-2).

Recruitment status

COMPLETED

Target enrollment

80 participants

Primary outcome timeframe

1 year

Results posted on

2022-04-19

Participant Flow

Participant milestones

Participant milestones
Measure
Rainbow Sensor
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
Overall Study
STARTED
80
Overall Study
COMPLETED
77
Overall Study
NOT COMPLETED
3

Reasons for withdrawal

Reasons for withdrawal
Measure
Rainbow Sensor
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
Overall Study
Equipment Issues
2
Overall Study
Protocol Violation
1

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Rainbow Sensor
n=77 Participants
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
Age, Categorical
<=18 years
0 Participants
n=77 Participants
Age, Categorical
Between 18 and 65 years
53 Participants
n=77 Participants
Age, Categorical
>=65 years
24 Participants
n=77 Participants
Sex: Female, Male
Female
51 Participants
n=77 Participants
Sex: Female, Male
Male
26 Participants
n=77 Participants
Region of Enrollment
United States
77 participants
n=77 Participants

PRIMARY outcome

Timeframe: 1 year

Population: In 4 subjects, the device failed initial calibration. In another subject, clinical care constraints precluded completion of the data collection.

Evaluate the clinical utility of the change in the oxygen reserve index as an early warning of impending arterial oxygen desaturation in Obese patients (30 \< BMI \< 40 kg m-2) and Normal BMI patients (19 \< BMI \< 25 kg m-2).

Outcome measures

Outcome measures
Measure
Rainbow Sensor
n=72 Participants
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
Utility of ORi Providing Added Warning Time
Added warning time in Obese
46.5 seconds
Utility of ORi Providing Added Warning Time
Added warning time in Normal BMI
87.0 seconds

Adverse Events

Rainbow Sensor

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Cristina Johnson

Masimo Corporation

Phone: 949-297-7000

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place

Restriction type: GT60