Trial Outcomes & Findings for Oxygen Reserve Index: Utility as Early Warning for Desaturation in Morbidly Obese Patients (NCT NCT03021551)
NCT ID: NCT03021551
Last Updated: 2022-04-19
Results Overview
Evaluate the clinical utility of the change in the oxygen reserve index as an early warning of impending arterial oxygen desaturation in Obese patients (30 \< BMI \< 40 kg m-2) and Normal BMI patients (19 \< BMI \< 25 kg m-2).
COMPLETED
80 participants
1 year
2022-04-19
Participant Flow
Participant milestones
| Measure |
Rainbow Sensor
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
|
|---|---|
|
Overall Study
STARTED
|
80
|
|
Overall Study
COMPLETED
|
77
|
|
Overall Study
NOT COMPLETED
|
3
|
Reasons for withdrawal
| Measure |
Rainbow Sensor
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
|
|---|---|
|
Overall Study
Equipment Issues
|
2
|
|
Overall Study
Protocol Violation
|
1
|
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Rainbow Sensor
n=77 Participants
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=77 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
53 Participants
n=77 Participants
|
|
Age, Categorical
>=65 years
|
24 Participants
n=77 Participants
|
|
Sex: Female, Male
Female
|
51 Participants
n=77 Participants
|
|
Sex: Female, Male
Male
|
26 Participants
n=77 Participants
|
|
Region of Enrollment
United States
|
77 participants
n=77 Participants
|
PRIMARY outcome
Timeframe: 1 yearPopulation: In 4 subjects, the device failed initial calibration. In another subject, clinical care constraints precluded completion of the data collection.
Evaluate the clinical utility of the change in the oxygen reserve index as an early warning of impending arterial oxygen desaturation in Obese patients (30 \< BMI \< 40 kg m-2) and Normal BMI patients (19 \< BMI \< 25 kg m-2).
Outcome measures
| Measure |
Rainbow Sensor
n=72 Participants
Rainbow sensor: Masimo Radical-7 and Root System with ORi parameter in all subjects are enrolled in the test group and receive an Rainbow sensor during their elective surgery.
|
|---|---|
|
Utility of ORi Providing Added Warning Time
Added warning time in Obese
|
46.5 seconds
|
|
Utility of ORi Providing Added Warning Time
Added warning time in Normal BMI
|
87.0 seconds
|
Adverse Events
Rainbow Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60