Trial Outcomes & Findings for Accuracy of Commercially Available Heart Rate Monitors II (NCT NCT02818244)

NCT ID: NCT02818244

Last Updated: 2017-07-02

Results Overview

The primary outcome measure is the accuracy of each heart rate monitor compared to ECG. This will be expressed by the correlation coefficient and will also be depicted by Bland-Altman plots.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

50 participants

Primary outcome timeframe

24 minutes

Results posted on

2017-07-02

Participant Flow

All subjects who met inclusion criteria and had no exclusions were enrolled. All subjects who were consented, participated and completed the study wearing a device on each wrist.

Unit of analysis: Heart Rate Monitors

Participant milestones

Participant milestones
Measure
Fit Bit Blaze
Fit Bit Blaze Heart Rate Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Garmin Forerunner 235
Garmin Forerunner 235 Heart Rate Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Tom Tom Spark Cardio
Tom Tom Spark Cardio Heart Rate Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Apple Watch
Apple Watch Heart Rate Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Overall Study
STARTED
25 25
25 25
25 25
25 25
Overall Study
COMPLETED
25 25
25 25
25 25
25 25
Overall Study
NOT COMPLETED
0 0
0 0
0 0
0 0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Race and Ethnicity were not collected from any participant.

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Fit Bit Blaze
n=25 Heart Rate
Fit Bit Blaze Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Garmin Forerunner 235
n=25 Heart Rate
Garmin Forerunner 235 Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Tom Tom Spark Cardio
n=25 Heart Rate
Tom Tom Spark Cardio Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Apple Watch
n=25 Heart Rate
Apple Watch Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Total
n=100 Heart Rate
Total of all reporting groups
Age, Categorical
<=18 years
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=100 Heart Rate
Age, Categorical
Between 18 and 65 years
25 Heart Rate
n=25 Heart Rate
25 Heart Rate
n=25 Heart Rate
25 Heart Rate
n=25 Heart Rate
25 Heart Rate
n=25 Heart Rate
100 Heart Rate
n=100 Heart Rate
Age, Categorical
>=65 years
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=25 Heart Rate
0 Heart Rate
n=100 Heart Rate
Sex: Female, Male
Female
14 Heart Rate
n=25 Heart Rate
12 Heart Rate
n=25 Heart Rate
15 Heart Rate
n=25 Heart Rate
13 Heart Rate
n=25 Heart Rate
54 Heart Rate
n=100 Heart Rate
Sex: Female, Male
Male
11 Heart Rate
n=25 Heart Rate
13 Heart Rate
n=25 Heart Rate
10 Heart Rate
n=25 Heart Rate
12 Heart Rate
n=25 Heart Rate
46 Heart Rate
n=100 Heart Rate
Race and Ethnicity Not Collected
0 Participants
Race and Ethnicity were not collected from any participant.

PRIMARY outcome

Timeframe: 24 minutes

The primary outcome measure is the accuracy of each heart rate monitor compared to ECG. This will be expressed by the correlation coefficient and will also be depicted by Bland-Altman plots.

Outcome measures

Outcome measures
Measure
Fit Bit Blaze
n=25 Participants
Fit Bit Blaze Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Garmin Forerunner 235
n=25 Participants
Garmin Forerunner 235 Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Tom Tom Spark Cardio
n=25 Participants
Tom Tom Spark Cardio Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Apple Watch
n=25 Participants
Apple Watch Heart Rate Monitoring Device Monitoring Device Standard ECG Polar Chest strap Scosche Rhythm+
Primary Outcome Measure: Heart Rate Monitor Accuracy Compared to ECG Expressed as Correlation Coefficient.
.67 Concordance Correlation Coefficient wECG
Interval 0.615 to 0.719
.81 Concordance Correlation Coefficient wECG
Interval 0.776 to 0.843
.83 Concordance Correlation Coefficient wECG
Interval 0.799 to 0.86
.92 Concordance Correlation Coefficient wECG
Interval 0.903 to 0.934

Adverse Events

Fit Bit Blaze

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Garmin Forerunner 235

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Tom Tom Spark Cardio

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Apple Watch

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Marc Gillinov MD

Cleveland Clinic

Phone: 216 445-8841

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place