Trial Outcomes & Findings for Comparison of Two Gingival Displacement Procedures (NCT NCT02716727)
NCT ID: NCT02716727
Last Updated: 2020-08-18
Results Overview
Impressions will be assessed for technique errors
COMPLETED
NA
16 participants
Through impression completion, an average of 10-20 minutes or when impression material is completely set
2020-08-18
Participant Flow
Participant milestones
| Measure |
Aquasil Ultra Cordless, Dentsply
Gingival displacement procedure that uses a pneumatic hand piece to inject the light body impression material into the gingival sulcus.
Aquasil Ultra Cordless, Dentsply: Cordless gingival displacement procedure
|
Ultrapak, Ultradent Cord
The knitted cord is used for physical displacement of gingival tissue. Cord is placed in the gingival sulcus and left in place for at least 4 minutes to achieve gingival displacement.
Ultrapak, Ultradent cord: Cordless gingival displacement procedure
|
|---|---|---|
|
Overall Study
STARTED
|
7
|
9
|
|
Overall Study
COMPLETED
|
7
|
9
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Race and Ethnicity were not collected from any participant.
Baseline characteristics by cohort
| Measure |
Aquasil Ultra Cordless, Dentsply
n=7 Participants
Gingival displacement procedure that uses a pneumatic hand piece to inject the light body impression material into the gingival sulcus.
Aquasil Ultra Cordless, Dentsply: Cordless gingival displacement procedure
|
Ultrapak, Ultradent Cord
n=9 Participants
The knitted cord is used for physical displacement of gingival tissue. Cord is placed in the gingival sulcus and left in place for at least 4 minutes to achieve gingival displacement.
Ultrapak, Ultradent cord: Cordless gingival displacement procedure
|
Total
n=16 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
56.2 years
STANDARD_DEVIATION 15.01 • n=7 Participants
|
65.0 years
STANDARD_DEVIATION 9.03 • n=9 Participants
|
60.5 years
STANDARD_DEVIATION 4.2 • n=16 Participants
|
|
Sex: Female, Male
Female
|
2 Participants
n=7 Participants
|
6 Participants
n=9 Participants
|
8 Participants
n=16 Participants
|
|
Sex: Female, Male
Male
|
5 Participants
n=7 Participants
|
3 Participants
n=9 Participants
|
8 Participants
n=16 Participants
|
|
Race and Ethnicity Not Collected
|
—
|
—
|
0 Participants
Race and Ethnicity were not collected from any participant.
|
|
Region of Enrollment
United States
|
7 Participants
n=7 Participants
|
9 Participants
n=9 Participants
|
16 Participants
n=16 Participants
|
PRIMARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setImpressions will be assessed for technique errors
Outcome measures
| Measure |
Aquasil Ultra Cordless, Dentsply
n=7 Participants
Gingival displacement procedure that uses a pneumatic hand piece to inject the light body impression material into the gingival sulcus.
Aquasil Ultra Cordless, Dentsply: Cordless gingival displacement procedure
|
Ultrapak, Ultradent Cord
n=9 Participants
The knitted cord is used for physical displacement of gingival tissue. Cord is placed in the gingival sulcus and left in place for at least 4 minutes to achieve gingival displacement.
Ultrapak, Ultradent cord: Cordless gingival displacement procedure
|
|---|---|---|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Pressure of Tray on Soft Tissue
|
0 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Void on Preparation
|
85.7 percentage of impressions
|
88.8 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Show-through of Occlusal/Incisal Edges
|
0 percentage of impressions
|
22.2 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Tissue Over Finish Line
|
42.9 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Voids/Bubbles
|
85.7 percentage of impressions
|
33.3 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Lack of Wash Material
|
28.6 percentage of impressions
|
44.4 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Tearing
|
71.4 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Cotton Roll Attached
|
0 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Inadequate Fusion Between Wash & Tray Material
|
57.1 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Lack of Polymerization
|
14.3 percentage of impressions
|
0 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Inadequate Retention
|
0 percentage of impressions
|
0 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Retraction Cord Attached
|
0 percentage of impressions
|
11.1 percentage of impressions
|
|
Percentage of Errors in Impressions Made for Indirect Restorations
Blood on the Impression
|
0 percentage of impressions
|
33.3 percentage of impressions
|
PRIMARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setThe time required to perform the impression procedures from the start of the gingival displacement procedure to completion of impression.
Outcome measures
| Measure |
Aquasil Ultra Cordless, Dentsply
n=7 Participants
Gingival displacement procedure that uses a pneumatic hand piece to inject the light body impression material into the gingival sulcus.
Aquasil Ultra Cordless, Dentsply: Cordless gingival displacement procedure
|
Ultrapak, Ultradent Cord
n=9 Participants
The knitted cord is used for physical displacement of gingival tissue. Cord is placed in the gingival sulcus and left in place for at least 4 minutes to achieve gingival displacement.
Ultrapak, Ultradent cord: Cordless gingival displacement procedure
|
|---|---|---|
|
Mean Time Required to Make the Impression
|
5.92 minutes
Standard Deviation 8.57
|
9.81 minutes
Standard Deviation 8.57
|
PRIMARY outcome
Timeframe: From start of visit until the end of the visit, approximately 3 hoursPatient discomfort will be assessed using the Wong-Baker FACES Pain Rating Scale™.The FACES Pain Scale is a dimensionless 10-point Likert scale used to assess self-reported pain intensity on a scale from 0 (no pain) to 10 (most pain you can imagine). Greater pain scores are indicative of more severe pain. Patients self-reported pain using the FACES evaluation at the beginning of the appointment, immediately before start of impression procedure, immediately after the impression procedure completed, and at the end of the appointment.
Outcome measures
| Measure |
Aquasil Ultra Cordless, Dentsply
n=7 Participants
Gingival displacement procedure that uses a pneumatic hand piece to inject the light body impression material into the gingival sulcus.
Aquasil Ultra Cordless, Dentsply: Cordless gingival displacement procedure
|
Ultrapak, Ultradent Cord
n=9 Participants
The knitted cord is used for physical displacement of gingival tissue. Cord is placed in the gingival sulcus and left in place for at least 4 minutes to achieve gingival displacement.
Ultrapak, Ultradent cord: Cordless gingival displacement procedure
|
|---|---|---|
|
Mean Patient Discomfort Scores
Immediately Before Starting Impression
|
8.64 score on a scale
Standard Deviation 7.87
|
7.43 score on a scale
Standard Deviation 7.87
|
|
Mean Patient Discomfort Scores
Start of Visit
|
7.64 score on a scale
Standard Deviation 6.03
|
8.31 score on a scale
Standard Deviation 6.03
|
|
Mean Patient Discomfort Scores
Immediately After Impression Completed
|
1.85 score on a scale
Standard Deviation 1.57
|
2.12 score on a scale
Standard Deviation 1.45
|
|
Mean Patient Discomfort Scores
At End of Visit
|
8.00 score on a scale
Standard Deviation 7.58
|
8.00 score on a scale
Standard Deviation 7.58
|
SECONDARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setPopulation: Due to loss of key study personnel these data were not collected.
Accuracy of the impression assessed for technique errors based on the clinical experience of the provider.
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setPopulation: Due to loss of key study personnel these data were not collected.
The time required to perform the impression procedures from the start of the gingival displacement procedure to completion of impression based on the clinical experience of the provider.
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: From start of visit until the end of the visit, approximately 3 hoursPopulation: Due to loss of key study personnel these data were not collected.
Patient discomfort will be assessed based on the clinical experience of the provider using the Wong-Baker FACES Pain Rating Scale™.The FACES Pain Scale is a dimensionless 10-point Likert scale used to assess self-reported pain intensity on a scale from 0 (no pain) to 10 (most pain you can imagine). Greater pain scores are indicative of more severe pain. Patients self-reported pain using the FACES evaluation at the beginning of the appointment, immediately before start of impression procedure, immediately after the impression procedure completed, and at the end of the appointment.
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setPopulation: Due to loss of key study personnel these data were not collected.
Accuracy of the impression assessed for technique errors based on the location of the tooth in the mouth (maxillary and mandibular arch).
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Through impression completion, an average of 10-20 minutes or when impression material is completely setPopulation: Due to loss of key study personnel these data were not collected.
The time required to perform the impression procedures from the start of the gingival displacement procedure to completion of impression based on the location of the tooth in the mouth (maxillary and mandibular arch).
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Through impression completion,an average of 10-20 minutes or when impression material is completely setPopulation: Due to loss of key study personnel these data were not collected.
Patient discomfort will be assessed based on the location of the tooth in the mouth (maxillary and mandibular arch) using the Wong-Baker FACES Pain Rating Scale™.The FACES Pain Scale is a dimensionless 10-point Likert scale used to assess self-reported pain intensity on a scale from 0 (no pain) to 10 (most pain you can imagine). Greater pain scores are indicative of more severe pain. Patients self-reported pain using the FACES evaluation at the beginning of the appointment, immediately before start of impression procedure, immediately after the impression procedure completed, and at the end of the appointment.
Outcome measures
Outcome data not reported
Adverse Events
Aquasil Ultra Cordless, Dentsply
Ultrapak, Ultradent Cord
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Sumitha Ahmed, DDS
University of North Carolina at Chapel Hill
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place