Trial Outcomes & Findings for Potential Effect of Dialyzer Leaching of BPA From the Fresenius Optiflux 160NR Compared to the Nipro ELISIO-15H (NCT NCT02627118)
NCT ID: NCT02627118
Last Updated: 2018-06-11
Results Overview
blood sampling of BPA (Bisphenol-A) in ng/mL following 2 months of dialysis with the Fresenius 160NR or the Nipro Elisio 15-H
COMPLETED
NA
10 participants
2 MONTHS OF TREATMENT
2018-06-11
Participant Flow
Participant milestones
| Measure |
Dialyzer Comparison
2 MONTHS OF DIALYSIS WITH THE FRESENIUS 160NR followed by 2 MONTHS OF DIALYSIS WITH THE NIPRO ELISIO-15H
|
|---|---|
|
Overall Study
STARTED
|
10
|
|
Overall Study
COMPLETED
|
10
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Potential Effect of Dialyzer Leaching of BPA From the Fresenius Optiflux 160NR Compared to the Nipro ELISIO-15H
Baseline characteristics by cohort
| Measure |
Dialyzer Comparison
n=10 Participants
2 MONTHS OF DIALYSIS WITH THE FRESENIUS 160NR followed by 2 MONTHS OF DIALYSIS WITH THE NIPRO ELISIO-15H
|
|---|---|
|
Age, Continuous
|
56.1 years
STANDARD_DEVIATION 16.6 • n=5 Participants
|
|
Sex: Female, Male
Female
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
10 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
White Non-Hispanic/Latino
|
3 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Hispanic
|
5 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
African American
|
1 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Other
|
1 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 2 MONTHS OF TREATMENTPopulation: When data was collected, BPA level was not measureable in one patient (lab reported the presence of unidentified "interfering substance") and that patient's data was removed from final analysis.
blood sampling of BPA (Bisphenol-A) in ng/mL following 2 months of dialysis with the Fresenius 160NR or the Nipro Elisio 15-H
Outcome measures
| Measure |
Dialyzer Comparison
n=9 Participants
2 MONTHS OF DIALYSIS WITH THE FRESENIUS 160NR followed by 2 MONTHS OF DIALYSIS WITH THE NIPRO ELISIO-15H
|
|---|---|
|
BPA Level Following 2 Months of Dialysis
following Fresenius 160NR dialyzer
|
9.76 ng/mL
Standard Error 14.65
|
|
BPA Level Following 2 Months of Dialysis
following 2 months of Nipro Elisio 15-H dialyzer
|
1.55 ng/mL
Standard Error 1.90
|
Adverse Events
FRESENIUS OPTIFLUX 160NR
NIPRO ELISIO-15H
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place