Trial Outcomes & Findings for Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions (NCT NCT02614768)

NCT ID: NCT02614768

Last Updated: 2023-05-06

Results Overview

The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

12 participants

Primary outcome timeframe

36 hours

Results posted on

2023-05-06

Participant Flow

Participant milestones

Participant milestones
Measure
Glucose Sensor
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
Overall Study
STARTED
12
Overall Study
COMPLETED
12
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Glucose Sensor
n=12 Participants
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
12 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
Sex: Female, Male
Female
5 Participants
n=5 Participants
Sex: Female, Male
Male
7 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
12 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Region of Enrollment
Europe
12 participants
n=5 Participants

PRIMARY outcome

Timeframe: 36 hours

The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.

Outcome measures

Outcome measures
Measure
Glucose Sensor
n=12 Participants
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Abbot
73.2 Percentage of sensor values
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Dexcom
56.1 Percentage of sensor values
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Medtronic
52.0 Percentage of sensor values

Adverse Events

Glucose Sensor

Serious events: 0 serious events
Other events: 3 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Other adverse events
Measure
Glucose Sensor
n=12 participants at risk
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
Skin and subcutaneous tissue disorders
Superficial excoriation
8.3%
1/12 • Number of events 1
Nervous system disorders
Headache
8.3%
1/12 • Number of events 1
Ear and labyrinth disorders
Vertigo
8.3%
1/12 • Number of events 1

Additional Information

Mader, Julia, Assoz. Dr.med.univ.

Medical University of Graz

Phone: 0043316385

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place