Trial Outcomes & Findings for Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions (NCT NCT02614768)
NCT ID: NCT02614768
Last Updated: 2023-05-06
Results Overview
The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.
COMPLETED
NA
12 participants
36 hours
2023-05-06
Participant Flow
Participant milestones
| Measure |
Glucose Sensor
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose).
Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
|
|---|---|
|
Overall Study
STARTED
|
12
|
|
Overall Study
COMPLETED
|
12
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions
Baseline characteristics by cohort
| Measure |
Glucose Sensor
n=12 Participants
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose).
Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
12 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Female
|
5 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
7 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
0 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
12 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
Europe
|
12 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 36 hoursThe primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations \<5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.
Outcome measures
| Measure |
Glucose Sensor
n=12 Participants
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose).
Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
|
|---|---|
|
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Abbot
|
73.2 Percentage of sensor values
|
|
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Dexcom
|
56.1 Percentage of sensor values
|
|
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Medtronic
|
52.0 Percentage of sensor values
|
Adverse Events
Glucose Sensor
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Glucose Sensor
n=12 participants at risk
Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose).
Glucose Sensor: The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
|
|---|---|
|
Skin and subcutaneous tissue disorders
Superficial excoriation
|
8.3%
1/12 • Number of events 1
|
|
Nervous system disorders
Headache
|
8.3%
1/12 • Number of events 1
|
|
Ear and labyrinth disorders
Vertigo
|
8.3%
1/12 • Number of events 1
|
Additional Information
Mader, Julia, Assoz. Dr.med.univ.
Medical University of Graz
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place