Trial Outcomes & Findings for Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis (NCT NCT02531646)
NCT ID: NCT02531646
Last Updated: 2017-03-09
Results Overview
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
COMPLETED
NA
66 participants
3 months
2017-03-09
Participant Flow
Participant milestones
| Measure |
Predicate & Invest. DE - Human Subjects
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Human Subjects
Radiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Phantom Images
Radiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|---|---|
|
Overall Study
STARTED
|
37
|
18
|
11
|
|
Overall Study
COMPLETED
|
32
|
17
|
11
|
|
Overall Study
NOT COMPLETED
|
5
|
1
|
0
|
Reasons for withdrawal
| Measure |
Predicate & Invest. DE - Human Subjects
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Human Subjects
Radiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Phantom Images
Radiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|---|---|
|
Overall Study
equip. image processing issue
|
5
|
1
|
0
|
Baseline Characteristics
Clinical Evaluation of CARESTREAM Dual Energy and Digital Tomosynthesis
Baseline characteristics by cohort
| Measure |
Predicate & Invest. DE - Human Subjects
n=37 Participants
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Human Subjects
n=18 Participants
Radiation - Fifteen to twenty (15-20) patients will receive a DR standard of care chest exam using the DRX Plus detector, and a DT exam. Each DT patient exam includes a scout image (chest PA) and a DT scan using the investigational DT SW. The DT scan is used by the DT console software to generate tomographic images.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Predicate & Invest. DT - Phantom Images
n=11 Participants
Radiation - Eleven (11) phantoms of various anatomy will be imaged with linear tomography (LT) as predicate and DT for investigational.
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
Total
n=66 Participants
Total of all reporting groups
|
|---|---|---|---|---|
|
Age, Customized
> 18 years
|
37 years
n=93 Participants
|
18 years
n=4 Participants
|
11 years
n=27 Participants
|
66 years
n=483 Participants
|
|
Sex/Gender, Customized
unknown
|
37 participants
n=93 Participants
|
18 participants
n=4 Participants
|
11 participants
n=27 Participants
|
66 participants
n=483 Participants
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 224 image ratings from 7 radiologist readers (32 images rated x 7 readers) for Predicate \& Invest. DE - Human Subjects arm above.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=224 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DE Predicate PA Chest
|
3.4540 units on a scale
Standard Error 0.0183
|
PRIMARY outcome
Timeframe: 3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=224 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Composite
|
3.3214 units on a scale
Standard Error 0.0242
|
PRIMARY outcome
Timeframe: 3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=224 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DE Investigational High Energy
|
3.2911 units on a scale
Standard Error 0.0178
|
PRIMARY outcome
Timeframe: 3 months1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=224 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DE Investigational Low Energy
|
3.5504 units on a scale
Standard Error 0.0163
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for Predicate \& Invest. DT Reference 1 - PA Chest arm above.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=119 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DT Reference 1- PA Chest
|
3.4832 units on a scale
Standard Error 0.0228
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for Predicate \& Invest. DT Reference 2 - PA and LAT Chest arm above.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=119 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DT Reference 2 - PA and LAT Chest
|
3.5546 units on a scale
Standard Error 0.0190
|
PRIMARY outcome
Timeframe: 3 monthPopulation: 119 image ratings from 7 radiologist readers (17 images rated x 7 readers) for DT Investigational - Scout \& DT Volume.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=119 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DT Investigational - Scout & DT Volume
|
3.6277 units on a scale
Standard Error 0.0230
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 77 phantom image ratings from 7 radiologist readers (11 images rated x 7 readers) for Predicate \& Invest. DT - Phantom Images arm above.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=77 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - LT Predicate Phantom Images
|
3.3494 units on a scale
Standard Error 0.0433
|
PRIMARY outcome
Timeframe: 3 monthsPopulation: 77 phantom image ratings from 7 radiologist readers (11 images rated x 7 readers) for Predicate \& Invest. DT - Phantom Images arm above.
1-1.9-Non-diagnostic Unacceptable for diagnostic purposes. Little or no clinically usable diagnostic information (e.g., gross underexposure, system failure or extensive motion artifact). Almost all such imaging should be repeated. 2-2.9-Limited Acceptable, with some technical defect (motion artifact, body habitus/poor x-ray penetration, or patient positioning may limit visualization of some body-regions but still adequate for diagnostic purposes). Not as much diagnostic information as is typical for an examination of this type, but likely sufficient. 3-3.9-Diagnostic Image quality that would be expected routinely when imaging cooperative patients. 4-Exemplary Good, most adequate for diagnostic purposes. Image quality that can serve as an example that should be emulated.
Outcome measures
| Measure |
Predicate & Invest. DE - Human Subjects
n=77 images
Radiation -Thirty to forty (30-40) patients will receive a DR standard of care chest exam using the DRX Plus detector and a DE exam. Each DE patient exam includes high energy and low energy image exposures using the investigational device. These images are used by the DE console software to generate additional DE images (e.g. bone and soft tissue).
Radiation: Radiation - Each human subject will receive one standard of care x-ray and one Duel Energy exposure or one standard of care x-ray and one Digital Tomosynthesis exposure.
|
|---|---|
|
Radlex Scale for Diagnostic Quality Ratings - DT Investigational Phantom Images
|
3.7156 units on a scale
Standard Error 0.0268
|
Adverse Events
Predicate & Invest. DE - Human Subjects
Predicate & Invest. DT - Human Subjects
Predicate & Invest. DT - Phantom Images
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Susan Pate, RN, MS, RAC, Senior Clinical Manager
Carestream Health, Inc.
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place