Trial Outcomes & Findings for Study of the Consumer Use of a New Home Test to Measure Sperm Concentration (NCT NCT02475395)

NCT ID: NCT02475395

Last Updated: 2017-10-18

Results Overview

Lay users obtained categorical sperm concentration result. Positive (for subfertility) results are less than or equal to 15 M/mL threshold. Gold standard reference (analysis by Computer-aided Semen Analysis \[CASA\]) result was measured and compared to Trak. True positive and true negative matched Reference category result and false negative, false positive did not match gold standard reference category result.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

272 participants

Primary outcome timeframe

Participants will be followed for one visit for up to 2 hours

Results posted on

2017-10-18

Participant Flow

Men were recruited at three medical clinic sites from 2 September 2015 to 4 December 2015.

Participant milestones

Participant milestones
Measure
Donor/Tester Subjects
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Overall Study
STARTED
258
14
Overall Study
COMPLETED
228
11
Overall Study
NOT COMPLETED
30
3

Reasons for withdrawal

Reasons for withdrawal
Measure
Donor/Tester Subjects
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Overall Study
Protocol Violation
30
3

Baseline Characteristics

Study of the Consumer Use of a New Home Test to Measure Sperm Concentration

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Donor/Tester Subjects
n=258 Participants
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
n=14 Participants
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Total
n=272 Participants
Total of all reporting groups
Age, Categorical
<=18 years
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
258 Participants
n=5 Participants
14 Participants
n=7 Participants
272 Participants
n=5 Participants
Age, Categorical
>=65 years
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
Age, Continuous
34 years
n=5 Participants
33 years
n=7 Participants
34 years
n=5 Participants
Sex: Female, Male
Female
0 Participants
n=5 Participants
14 Participants
n=7 Participants
14 Participants
n=5 Participants
Sex: Female, Male
Male
258 Participants
n=5 Participants
0 Participants
n=7 Participants
258 Participants
n=5 Participants
Region of Enrollment
United States
258 participants
n=5 Participants
14 participants
n=7 Participants
272 participants
n=5 Participants

PRIMARY outcome

Timeframe: Participants will be followed for one visit for up to 2 hours

Lay users obtained categorical sperm concentration result. Positive (for subfertility) results are less than or equal to 15 M/mL threshold. Gold standard reference (analysis by Computer-aided Semen Analysis \[CASA\]) result was measured and compared to Trak. True positive and true negative matched Reference category result and false negative, false positive did not match gold standard reference category result.

Outcome measures

Outcome measures
Measure
Donor/Tester Subjects
n=228 Participants
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
n=11 Participants
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Number of Untrained Lay Users That Obtained Accurate and Inaccurate Subfertility Results From the TRAK Device When Compared to Results Obtained From the Gold Standard
True Positive Trak Results
52 Participants
2 Participants
Number of Untrained Lay Users That Obtained Accurate and Inaccurate Subfertility Results From the TRAK Device When Compared to Results Obtained From the Gold Standard
True Negative Trak Results
160 Participants
7 Participants
Number of Untrained Lay Users That Obtained Accurate and Inaccurate Subfertility Results From the TRAK Device When Compared to Results Obtained From the Gold Standard
False Positive Trak Results
10 Participants
2 Participants
Number of Untrained Lay Users That Obtained Accurate and Inaccurate Subfertility Results From the TRAK Device When Compared to Results Obtained From the Gold Standard
False Negative Trak Results
6 Participants
0 Participants

SECONDARY outcome

Timeframe: Participants will be followed for one visit for up to 2 hours

Healthcare professions obtained a categorical sperm concentration result by observing completed assay outputs as performed by Lay Users. Positive (for subfertility) results are less than or equal to 15 M/mL threshold. Reference result using gold standard (CASA) was measured and compared to Trak. True positive and true negative matched Reference category result and false negative, false positive did not match reference category result.

Outcome measures

Outcome measures
Measure
Donor/Tester Subjects
n=228 Participants
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
n=11 Participants
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Number of Accurate and Inaccurate Subfertility Results as Obtained by Healthcare Professionals Observing Assays Result Performed by Untrained Lay Users.
True Positive Trak Results
56 Participants
1 Participants
Number of Accurate and Inaccurate Subfertility Results as Obtained by Healthcare Professionals Observing Assays Result Performed by Untrained Lay Users.
True Negative Trak Results
161 Participants
7 Participants
Number of Accurate and Inaccurate Subfertility Results as Obtained by Healthcare Professionals Observing Assays Result Performed by Untrained Lay Users.
False Positive Trak Results
9 Participants
2 Participants
Number of Accurate and Inaccurate Subfertility Results as Obtained by Healthcare Professionals Observing Assays Result Performed by Untrained Lay Users.
False Negative Trak Results
2 Participants
1 Participants

SECONDARY outcome

Timeframe: Participants will be followed for one visit for up to 2 hours

Population: One fewer was analyzed in this population because a sample did not have enough sample to allow the HCP to run the test.

Healthcare professions obtained a categorical sperm concentration result by performing assay on aliquot obtained from Lay User's sample. Positive (for subfertility) results are less than or equal to 15 M/mL threshold. Reference result using gold standard (CASA) was measured and compared to Trak. True positive and true negative matched Reference category result and false negative, false positive did not match reference category result.

Outcome measures

Outcome measures
Measure
Donor/Tester Subjects
n=227 Participants
Male subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Tester Only Subjects
n=11 Participants
Male or female subjects use the TRAK device to attain a measurement of sperm concentration from a semen specimen TRAK device: Use of TRAK to attain sperm concentration measurement
Number of Accurate and Inaccurate Results Obtained by Healthcare Professionals Performing Trak Assays on Subjects' Samples.
True Positive Trak Results
57 Participants
1 Participants
Number of Accurate and Inaccurate Results Obtained by Healthcare Professionals Performing Trak Assays on Subjects' Samples.
True Negative Trak Results
160 Participants
7 Participants
Number of Accurate and Inaccurate Results Obtained by Healthcare Professionals Performing Trak Assays on Subjects' Samples.
False Positive Trak Results
9 Participants
2 Participants
Number of Accurate and Inaccurate Results Obtained by Healthcare Professionals Performing Trak Assays on Subjects' Samples.
False Negative Trak Results
1 Participants
1 Participants

Adverse Events

Donor/Tester Subjects

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Tester Only Subjects

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Cheif Science Officer

Sandstone Diagnostics

Phone: 9253157246

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place