Trial Outcomes & Findings for Study of Low Level Laser Therapy to Treat Diabetic Peripheral Neuropathy Foot Pain (NCT NCT02461225)
NCT ID: NCT02461225
Last Updated: 2022-03-15
Results Overview
The Visual Analog Scale (VAS) from baseline to endpoint (6 weeks) measurement is calculated for each subject. Individual subject success is defined as a 30% or greater change in VAS pain scores across the evaluation period. A negative (-) percent change indicates a decrease in pain level and is positive for individual subject success. A lower Visual Analog Scale pain score from baseline to study endpoint represents a better outcome, whereas a higher score from baseline to study endpoint represents a worse outcome. For the primary study outcome, the overall study success is defined as a 35% or greater difference between the proportion (number) of individual successes in each treatment group.
COMPLETED
NA
30 participants
Baseline and 6 weeks
2022-03-15
Participant Flow
Participant milestones
| Measure |
Erchonia® FX-635™
The Erchonia® FX-635™ is made up of 3 independent 17 milliWatts (mW), 635 nanometers (nm) red laser diodes mounted in scanner devices with flexible arms positioned equidistant from each other.
Erchonia® FX-635™: The Erchonia® FX-635™ is made up of 3 independent 17 mW, 635 nm red laser diodes that are applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
Placebo Laser
The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light.
Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light. It is applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
|---|---|---|
|
Overall Study
STARTED
|
19
|
11
|
|
Overall Study
COMPLETED
|
19
|
11
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Study of Low Level Laser Therapy to Treat Diabetic Peripheral Neuropathy Foot Pain
Baseline characteristics by cohort
| Measure |
Erchonia® FX-635™
n=19 Participants
The Erchonia® FX-635™ is made up of 3 independent 17 milliWatts (mW), 635 nanometers (nm) red laser diodes mounted in scanner devices with flexible arms positioned equidistant from each other.
Erchonia® FX-635™: The Erchonia® FX-635™ is made up of 3 independent 17 mW, 635 nm red laser diodes that are applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
Placebo Laser
n=11 Participants
The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light.
Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light. It is applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
Total
n=30 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
49.83 years
STANDARD_DEVIATION 13.19 • n=5 Participants
|
53.45 years
STANDARD_DEVIATION 11.26 • n=7 Participants
|
51.24 years
STANDARD_DEVIATION 12.56 • n=5 Participants
|
|
Sex: Female, Male
Female
|
12 Participants
n=5 Participants
|
5 Participants
n=7 Participants
|
17 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
7 Participants
n=5 Participants
|
6 Participants
n=7 Participants
|
13 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
4 Participants
n=5 Participants
|
2 Participants
n=7 Participants
|
6 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
2 Participants
n=5 Participants
|
2 Participants
n=7 Participants
|
4 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
13 Participants
n=5 Participants
|
7 Participants
n=7 Participants
|
20 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
Ireland
|
19 Participants
n=5 Participants
|
11 Participants
n=7 Participants
|
30 Participants
n=5 Participants
|
|
Visual Analog Scale Pain Rating
|
68.89 units on a scale
STANDARD_DEVIATION 12.49 • n=5 Participants
|
65.91 units on a scale
STANDARD_DEVIATION 10.69 • n=7 Participants
|
67.88 units on a scale
STANDARD_DEVIATION 11.99 • n=5 Participants
|
PRIMARY outcome
Timeframe: Baseline and 6 weeksThe Visual Analog Scale (VAS) from baseline to endpoint (6 weeks) measurement is calculated for each subject. Individual subject success is defined as a 30% or greater change in VAS pain scores across the evaluation period. A negative (-) percent change indicates a decrease in pain level and is positive for individual subject success. A lower Visual Analog Scale pain score from baseline to study endpoint represents a better outcome, whereas a higher score from baseline to study endpoint represents a worse outcome. For the primary study outcome, the overall study success is defined as a 35% or greater difference between the proportion (number) of individual successes in each treatment group.
Outcome measures
| Measure |
Erchonia® FX-635™
n=19 Participants
The Erchonia® FX-635™ is made up of 3 independent 17 milliWatts (mW), 635 nanometers (nm) red laser diodes mounted in scanner devices with flexible arms positioned equidistant from each other.
Erchonia® FX-635™: The Erchonia® FX-635™ is made up of 3 independent 17 mW, 635 nm red laser diodes that are applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
Placebo Laser
n=11 Participants
The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light.
Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light. It is applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
|---|---|---|
|
Number of Participants With a 30% or Greater Change in Pain Rating on the Visual Analog Scale (VAS)
|
18 participants
|
4 participants
|
SECONDARY outcome
Timeframe: Baseline and 6 weeksThe Visual Analog Scale (VAS) assesses level or degree of pain. It is a horizontal line anchored on the left by the label '0: no pain at all' and on the right by the label '100: worst pain imaginable'. The subject marks a location on the 0-100 line that appears to represent any level of pain he or she is experiencing at that time. This marking is measured with a 0 to 100 mm ruler and the number recorded. A lower Visual Analog Scale pain score from baseline to study endpoint represents a better outcome, whereas a higher score from baseline to study endpoint represents a worse outcome.
Outcome measures
| Measure |
Erchonia® FX-635™
n=19 Participants
The Erchonia® FX-635™ is made up of 3 independent 17 milliWatts (mW), 635 nanometers (nm) red laser diodes mounted in scanner devices with flexible arms positioned equidistant from each other.
Erchonia® FX-635™: The Erchonia® FX-635™ is made up of 3 independent 17 mW, 635 nm red laser diodes that are applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
Placebo Laser
n=11 Participants
The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light.
Placebo Laser: The Placebo Laser has the same appearance as the Erchonia® FX-635™ but does not emit any therapeutic light. It is applied for 15 minutes to each foot. The diodes are positioned around the foot such that each of the 3 laser lights is positioned between 3 and 4 inches away from, but directed toward: a) the top of the foot (dorsal aspect); b) the bottom of the foot (plantar aspect); and c) the posterior tibial nerve within the tarsal tunnel. The laser diodes laze each of these 3 areas for 15 minutes simultaneously. Each subject receives 12 total procedure administrations with the Erchonia® FX-635™ across the consecutive 6-week procedure administration phase, 2 procedure administrations per week, each procedure administration approximately evenly spaced.
|
|---|---|---|
|
Change in Pain Rating on the Visual Analog Scale (VAS)
|
-60.97 score on a scale
Standard Deviation 23.28
|
-9.09 score on a scale
Standard Deviation 34.06
|
Adverse Events
Erchonia® FX-635™
Placebo Laser
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place