Trial Outcomes & Findings for Comparative Effectiveness Study of Bubble CPAP Devices in the NICU (NCT NCT02392806)

NCT ID: NCT02392806

Last Updated: 2018-03-09

Results Overview

Bubble CPAP failure (re-intubation or use of non-invasive positive pressure ventilation) within 72 hours following extubation

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

43 participants

Primary outcome timeframe

Within 72 hours of extubation

Results posted on

2018-03-09

Participant Flow

Participant milestones

Participant milestones
Measure
Bubble CPAP- BabiPlus, Respiralogics
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Overall Study
STARTED
22
21
Overall Study
COMPLETED
22
21
Overall Study
NOT COMPLETED
0
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Comparative Effectiveness Study of Bubble CPAP Devices in the NICU

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Total
n=43 Participants
Total of all reporting groups
Age, Continuous
26.0 Weeks Gestational Age
STANDARD_DEVIATION 2.0 • n=5 Participants
25.4 Weeks Gestational Age
STANDARD_DEVIATION 1.4 • n=7 Participants
25.7 Weeks Gestational Age
STANDARD_DEVIATION 1.7 • n=5 Participants
Sex: Female, Male
Female
9 Participants
n=5 Participants
9 Participants
n=7 Participants
18 Participants
n=5 Participants
Sex: Female, Male
Male
13 Participants
n=5 Participants
12 Participants
n=7 Participants
25 Participants
n=5 Participants

PRIMARY outcome

Timeframe: Within 72 hours of extubation

Bubble CPAP failure (re-intubation or use of non-invasive positive pressure ventilation) within 72 hours following extubation

Outcome measures

Outcome measures
Measure
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Number of Participants With Extubation Failure
9 Participants
7 Participants

SECONDARY outcome

Timeframe: 96 hours

Number of infants that reach 21% inspired oxygen during initial study period of 72 hours plus the crossover period of 24 hours

Outcome measures

Outcome measures
Measure
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
Oxygenation - Oxygen Saturation Via Pulse Oximetry Recorded Hourly
5 Participants
5 Participants

Adverse Events

Bubble CPAP- BabiPlus, Respiralogics

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Bubble CPAP - B&B Bubbler, B&B Medical Technologies

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported

Additional Information

Noah Hillman

Saint Louis University

Phone: 3145779292

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place