Trial Outcomes & Findings for Comparative Effectiveness Study of Bubble CPAP Devices in the NICU (NCT NCT02392806)
NCT ID: NCT02392806
Last Updated: 2018-03-09
Results Overview
Bubble CPAP failure (re-intubation or use of non-invasive positive pressure ventilation) within 72 hours following extubation
Recruitment status
COMPLETED
Study phase
NA
Target enrollment
43 participants
Primary outcome timeframe
Within 72 hours of extubation
Results posted on
2018-03-09
Participant Flow
Participant milestones
| Measure |
Bubble CPAP- BabiPlus, Respiralogics
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device
Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device
Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
|---|---|---|
|
Overall Study
STARTED
|
22
|
21
|
|
Overall Study
COMPLETED
|
22
|
21
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Comparative Effectiveness Study of Bubble CPAP Devices in the NICU
Baseline characteristics by cohort
| Measure |
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device
Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device
Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
Total
n=43 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
26.0 Weeks Gestational Age
STANDARD_DEVIATION 2.0 • n=5 Participants
|
25.4 Weeks Gestational Age
STANDARD_DEVIATION 1.4 • n=7 Participants
|
25.7 Weeks Gestational Age
STANDARD_DEVIATION 1.7 • n=5 Participants
|
|
Sex: Female, Male
Female
|
9 Participants
n=5 Participants
|
9 Participants
n=7 Participants
|
18 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
13 Participants
n=5 Participants
|
12 Participants
n=7 Participants
|
25 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: Within 72 hours of extubationBubble CPAP failure (re-intubation or use of non-invasive positive pressure ventilation) within 72 hours following extubation
Outcome measures
| Measure |
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device
Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device
Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
|---|---|---|
|
Number of Participants With Extubation Failure
|
9 Participants
|
7 Participants
|
SECONDARY outcome
Timeframe: 96 hoursNumber of infants that reach 21% inspired oxygen during initial study period of 72 hours plus the crossover period of 24 hours
Outcome measures
| Measure |
Bubble CPAP- BabiPlus, Respiralogics
n=22 Participants
Infants randomized to the BabiPlus device will be extubated to BabiPlus Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the B\&B Bubbler device
Bubble CPAP- BabiPlus, Respiralogics: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
n=21 Participants
Infants randomized to the B\&B Bubbler device will be extubated to B\&B Bubbler Bubble CPAP device and after 72 hours, if the infant has remained extubated, will be crossed-over to the BabiPlus device
Bubble CPAP- B\&B Bubbler, B\&B Medical devices: Infant will be randomized to BabiPlus verses B\&B Bubbler at time of extubation
|
|---|---|---|
|
Oxygenation - Oxygen Saturation Via Pulse Oximetry Recorded Hourly
|
5 Participants
|
5 Participants
|
Adverse Events
Bubble CPAP- BabiPlus, Respiralogics
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Bubble CPAP - B&B Bubbler, B&B Medical Technologies
Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place