Trial Outcomes & Findings for SentiMag® Intraoperative Comparison in Breast Cancer (NCT NCT02336737)
NCT ID: NCT02336737
Last Updated: 2020-11-23
Results Overview
The proportion of lymph nodes detected intraoperatively by SentiMag and SiennaXP in relation the proportion of lymph nodes detected by the combination of Technetium Sulfur Colloid and Isosulfan blue dye
COMPLETED
NA
160 participants
During surgical procedure <1 hour
2020-11-23
Participant Flow
Participant milestones
| Measure |
SiennaXP Injection
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Overall Study
STARTED
|
160
|
|
Overall Study
COMPLETED
|
147
|
|
Overall Study
NOT COMPLETED
|
13
|
Reasons for withdrawal
| Measure |
SiennaXP Injection
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Overall Study
Withdrawal by Subject
|
5
|
|
Overall Study
Physician Decision
|
8
|
Baseline Characteristics
SentiMag® Intraoperative Comparison in Breast Cancer
Baseline characteristics by cohort
| Measure |
SiennaXP Injection
n=369 Sentinel Lymph Nodes (SLN)
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
90 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
57 Participants
n=5 Participants
|
|
Age, Continuous
|
61.1 Years
STANDARD_DEVIATION 12.3 • n=5 Participants
|
|
Sex: Female, Male
Female
|
147 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
0 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
17 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
130 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
7 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
11 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
121 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
8 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
147 subjects
n=5 Participants
|
PRIMARY outcome
Timeframe: During surgical procedure <1 hourPopulation: 146/147 were evaluated for the primary objective. One (1) subject had missing data and therefore was not included in the primary endpoint analysis
The proportion of lymph nodes detected intraoperatively by SentiMag and SiennaXP in relation the proportion of lymph nodes detected by the combination of Technetium Sulfur Colloid and Isosulfan blue dye
Outcome measures
| Measure |
SiennaXP Injection
n=146 Participants
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Number of Participants With Detected Lymph Nodes
|
145 Participants
|
PRIMARY outcome
Timeframe: 3-6 weeksPopulation: Number of subjects experiencing Sienna XP injection-related adverse events from enrollment through subject exit or completion
Number of Participants with Adverse Events relating to Sienna XP Injection
Outcome measures
| Measure |
SiennaXP Injection
n=160 Participants
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Safety of SiennaXP and SentiMag® as Indicated by Adverse Events and Serious Adverse Events and Their Relatedness to the Detection Method or Procedure.
|
20 Participants
|
SECONDARY outcome
Timeframe: Intraoperative <1 hourThe number of nodes identified by both SiennaXP and Control out of nodes identified by Control
Outcome measures
| Measure |
SiennaXP Injection
n=147 Participants
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Nodal Concordance: Number of Nodes Identified by Both Test and Control Out of Nodes Identified by Control
|
94.5 percentage of concordance
Interval 92.1 to 96.9
|
SECONDARY outcome
Timeframe: Intraoperative <1 hourThe per patient detection rate for a combination of all methods (magnetic, combined radioisotope and blue dye; radioisotope alone; blue dye alone)
Outcome measures
| Measure |
SiennaXP Injection
n=146 Participants
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Number of Participants With Lymph Nodes Detected by Combined Radioisotope,Blue Dye and Magnetic Technique
|
145 Participants
|
Adverse Events
SiennaXP Injection
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
SiennaXP Injection
n=147 participants at risk
Single injection of SiennaXP in addition to comparator single dose of radioisotope (Technetium Tc99m Sulfur Colloid) and single dose of isosulfan blue dye.
Lymph node localization using the SentiMag handheld intraoperative localization system in addition to localization with standard of care handheld gamma probe.
SiennaXP: Sub-cutaneous injection of SiennaXP magnetic marker, followed by lymph node localization using the SentiMag handheld magnetic probe
Technetium Tc99m Sulfur Colloid: Injection of a single dose of radioisotope (Technetium Tc99m Sulfur Colloid)
Isosulfan blue dye: Injection of a single dose of isosulfan blue dye
|
|---|---|
|
Skin and subcutaneous tissue disorders
Breast discolouration/hyperpigmentation
|
15.6%
23/147 • Number of events 23
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place