Trial Outcomes & Findings for Clinical Study of the Optic Disc Parameters (NCT NCT02277119)
NCT ID: NCT02277119
Last Updated: 2022-06-28
Results Overview
Reporting of the Optic Disc Size difference between the Maestro and iVue
COMPLETED
NA
109 participants
1 Hour
2022-06-28
Participant Flow
Participant milestones
| Measure |
Normal Eyes
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Overall Study
STARTED
|
58
|
24
|
27
|
|
Overall Study
COMPLETED
|
55
|
20
|
19
|
|
Overall Study
NOT COMPLETED
|
3
|
4
|
8
|
Reasons for withdrawal
| Measure |
Normal Eyes
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Overall Study
Lack of Efficacy
|
1
|
1
|
1
|
|
Overall Study
Protocol Violation
|
2
|
3
|
7
|
Baseline Characteristics
Clinical Study of the Optic Disc Parameters
Baseline characteristics by cohort
| Measure |
Normal Eyes
n=58 Participants
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
n=24 Participants
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
n=27 Participants
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Total
n=109 Participants
Total of all reporting groups
|
|---|---|---|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
0 Participants
n=4 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
55 Participants
n=5 Participants
|
16 Participants
n=7 Participants
|
19 Participants
n=5 Participants
|
90 Participants
n=4 Participants
|
|
Age, Categorical
>=65 years
|
3 Participants
n=5 Participants
|
8 Participants
n=7 Participants
|
8 Participants
n=5 Participants
|
19 Participants
n=4 Participants
|
|
Age, Continuous
|
39.2 years
STANDARD_DEVIATION 13.3 • n=5 Participants
|
57.8 years
STANDARD_DEVIATION 20.2 • n=7 Participants
|
60.5 years
STANDARD_DEVIATION 13.4 • n=5 Participants
|
48.7 years
STANDARD_DEVIATION 18.1 • n=4 Participants
|
|
Sex: Female, Male
Female
|
32 Participants
n=5 Participants
|
10 Participants
n=7 Participants
|
13 Participants
n=5 Participants
|
55 Participants
n=4 Participants
|
|
Sex: Female, Male
Male
|
26 Participants
n=5 Participants
|
14 Participants
n=7 Participants
|
14 Participants
n=5 Participants
|
54 Participants
n=4 Participants
|
|
Region of Enrollment
United States
|
58 participants
n=5 Participants
|
24 participants
n=7 Participants
|
27 participants
n=5 Participants
|
109 participants
n=4 Participants
|
|
Optic Disc Area Difference
|
0.197 Microns Squared
STANDARD_DEVIATION 0.228 • n=5 Participants
|
0.201 Microns Squared
STANDARD_DEVIATION 0.262 • n=7 Participants
|
0.097 Microns Squared
STANDARD_DEVIATION 0.216 • n=5 Participants
|
0.165 Microns Squared
STANDARD_DEVIATION 0.235 • n=4 Participants
|
PRIMARY outcome
Timeframe: 1 HourReporting of the Optic Disc Size difference between the Maestro and iVue
Outcome measures
| Measure |
Normal Eyes
n=58 Participants
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
n=24 Participants
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
n=27 Participants
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Optic Disc Measurements (Optic Disc Size)
|
0.197 Mircrons squared
Standard Deviation .228
|
0.097 Mircrons squared
Standard Deviation 0.216
|
0.201 Mircrons squared
Standard Deviation 0.262
|
PRIMARY outcome
Timeframe: 1 HourRNFL thickness measured
Outcome measures
| Measure |
Normal Eyes
n=58 Participants
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
n=24 Participants
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
n=27 Participants
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Retinal Nerve Fiber Layer (RNFL) Thickness Measurements
|
4.682 Microns
Standard Deviation 4.333
|
2.526 Microns
Standard Deviation 6.367
|
4.103 Microns
Standard Deviation 8.789
|
PRIMARY outcome
Timeframe: 1 HourPopulation: Glaucomatous eyes were not scanned and analyzed in the Full Retina Thickness portion of the study. Since the imaging is done in a different area of the eye compare to Retinal Nerve Fiber Layer participants analyzed will be different between these measurement areas.
Full Retinal Thicknesses Measurement
Outcome measures
| Measure |
Normal Eyes
n=19 Participants
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
n=18 Participants
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Full Retinal Thickness Measurement
|
22.845 Microns
Standard Deviation 11.482
|
—
|
21.299 Microns
Standard Deviation 24.524
|
PRIMARY outcome
Timeframe: 1 HourReporting of the Cup size difference between the Maestro and iVue
Outcome measures
| Measure |
Normal Eyes
n=58 Participants
Subjects with no known ocular diseases will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Glaucomatous Eyes
n=24 Participants
Subjects presenting with different stages of glaucoma will be scanned with the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
Eyes With Retinal Diseases
n=27 Participants
Subjects presenting with Retinal pathological eyes will be scanned on the iVue and Maestro device
Maestro: OCT machines used for diagnostic purposes
iVue: OCT machines used for diagnostic purposes
|
|---|---|---|---|
|
Optic Disc Measurement (Cup Size)
|
0 Microns cubed
Standard Deviation 0.072
|
-0.003 Microns cubed
Standard Deviation 0.124
|
0.016 Microns cubed
Standard Deviation 0.065
|
Adverse Events
Normal Eyes
Glaucomatous Eyes
Eyes With Retinal Diseases
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60