Trial Outcomes & Findings for ZIPS Study - Zip Incision aPproximation vs. Suture (NCT NCT02213510)
NCT ID: NCT02213510
Last Updated: 2016-05-30
Results Overview
Duration of time starting when suture needle (control) or Zip device touches the skin until final suture knot is cut or Zip device application is complete (e.g., top liner is removed.)
COMPLETED
NA
40 participants
2 weeks
2016-05-30
Participant Flow
Participant milestones
| Measure |
Zip Surgical Skin Closure Device
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Overall Study
STARTED
|
21
|
19
|
|
Overall Study
COMPLETED
|
20
|
18
|
|
Overall Study
NOT COMPLETED
|
1
|
1
|
Reasons for withdrawal
| Measure |
Zip Surgical Skin Closure Device
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Overall Study
Death
|
0
|
1
|
|
Overall Study
Lost to Follow-up
|
1
|
0
|
Baseline Characteristics
ZIPS Study - Zip Incision aPproximation vs. Suture
Baseline characteristics by cohort
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
Total
n=40 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
71 years
n=5 Participants
|
70 years
n=7 Participants
|
70 years
n=5 Participants
|
|
Sex: Female, Male
Female
|
8 Participants
n=5 Participants
|
7 Participants
n=7 Participants
|
15 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
13 Participants
n=5 Participants
|
12 Participants
n=7 Participants
|
25 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
1 Participants
n=5 Participants
|
1 Participants
n=7 Participants
|
2 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
1 Participants
n=5 Participants
|
1 Participants
n=7 Participants
|
2 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
19 Participants
n=5 Participants
|
17 Participants
n=7 Participants
|
36 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
21 participants
n=5 Participants
|
19 participants
n=7 Participants
|
40 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
1 - Always burns, never tans
|
2 participants
n=5 Participants
|
0 participants
n=7 Participants
|
2 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
2 - Usually burns, tans with difficulty
|
1 participants
n=5 Participants
|
1 participants
n=7 Participants
|
2 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
3 - Somtimes mild burn, gradually tans
|
7 participants
n=5 Participants
|
6 participants
n=7 Participants
|
13 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
4 - Rarely burns, tans with ease
|
6 participants
n=5 Participants
|
7 participants
n=7 Participants
|
13 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
5 - Very rarely burns, tans very easily
|
4 participants
n=5 Participants
|
3 participants
n=7 Participants
|
7 participants
n=5 Participants
|
|
Fitzpatrick Skin Scale
6 - Never burns, tans very easily
|
1 participants
n=5 Participants
|
2 participants
n=7 Participants
|
3 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 2 weeksDuration of time starting when suture needle (control) or Zip device touches the skin until final suture knot is cut or Zip device application is complete (e.g., top liner is removed.)
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Overall Closure Time
|
78 seconds
Standard Deviation 30.36
|
216 seconds
Standard Deviation 93.71
|
PRIMARY outcome
Timeframe: 3 MonthsPopulation: Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Based on photographs taken of scars taken at 3 months following CIED procedure. The CVAS scale is measured from 0mm (representing best scar) to 100 mm (representing worst scar).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Wound Healing as Determined by the CVAS (Cosmetic Visual Analogue Scale)
|
35.8 mm
Standard Deviation 40.6
|
40.6 mm
Standard Deviation 21.7
|
SECONDARY outcome
Timeframe: 2 weeksAt 2 weeks post-procedure, the Wound Evaluation Scale was measured. The scale is based from 0 (representing normal skin) to 100 (representing a poor scar).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Surgeon Wound Evaluation Scale (WES)
|
5.82 units on a scale
Standard Deviation 0.50
|
5.78 units on a scale
Standard Deviation 0.55
|
SECONDARY outcome
Timeframe: 3 MonthsThe surgeon will complete an assessment of the following: 1. Closure Method Satisfaction 2. Scar Satisfaction 3. Wound Healing as judged by Wound Evaluation Scale Scale rated on 1 (least favorable) to 6 (most favorable) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Surgeon Evaluation Based on the Wound Evaluation Scale (WES)
|
6 units on a scale
Standard Deviation 0
|
6 units on a scale
Standard Deviation 0
|
SECONDARY outcome
Timeframe: 3 monthsAt 3 months post-procedure, the surgeon evaluated their satisfaction on a scale from 1 to 5, 5 being the least favorable. Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Surgeon Satisfaction With Scar
|
1 units on a scale
Standard Deviation 0
|
1.06 units on a scale
Standard Deviation 0.24
|
SECONDARY outcome
Timeframe: 2 weeksAt 2 weeks post-procedure, patient comfort was evaluated with a scale 1 (most favorable) to 5 (least favorable).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Comfort
|
1.77 units on a scale
Standard Deviation 1.15
|
1.94 units on a scale
Standard Deviation 1.16
|
SECONDARY outcome
Timeframe: 3 monthsPatient will complete questionnaires that includes assessments of the following: 1. Pain 2. Closure Method Comfort 3. Closure Method Satisfaction 4. Scar Satisfaction Scar Satisfaction was measured on a scale from 1 (most satisfied) to 5 (least satisfied) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Satisfaction With Scar
|
1.32 units on a scale
Standard Deviation 0.48
|
1.22 units on a scale
Standard Deviation 0.43
|
SECONDARY outcome
Timeframe: 3 monthsAt 3 months post-procedure, patients were asked to rate their scar. Scale was measured from 0 (best scar) - 10 (worst scar) Please note that only 18 patients were evaluated (of 19) in the Standard Suture Closure due to one patient withdrawing from the trial.
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Rating of Scar
|
0.87 units on a scale
Standard Deviation 0.76
|
1.04 units on a scale
Standard Deviation 1.39
|
SECONDARY outcome
Timeframe: 1 day (discharge)At the discharge visit, patients evaluated their level of pain at the incision site based on a scale 1 (least pain) through 10 (worst pain).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Incision Pain
|
1.77 units on a scale
Standard Deviation 2.48
|
2.15 units on a scale
Standard Deviation 1.83
|
SECONDARY outcome
Timeframe: 2 weeksAt 2 week post-procedure, incision pain was evaluated on a scale 0 to 10 (worst pain).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=19 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Incision Pain
|
0.64 units on a scale
Standard Deviation 0.92
|
0.71 units on a scale
Standard Deviation 1.41
|
SECONDARY outcome
Timeframe: 3 monthsIncision pain experienced by the patient was evaluated at 3 months post-procedure. This was evaluated on a scale of 0 (least pain) to 10 (worst pain).
Outcome measures
| Measure |
Zip Surgical Skin Closure Device
n=21 Participants
The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
Zip Surgical Skin Closure Device: The Zip Surgical Skin Closure device is a single use, sterile medical device that replaces sutures, staples, and glue for closure of the skin layer for surgical incisions or laceration repair. The device will applied by the surgeon at the end of the CIED procedure and be worn until the two week post-operative wound check.
|
Standard Suture Closure
n=18 Participants
The surgeon will perform standard suture closure for the skin layer following CIED procedure.
Standard Suture Closure: The surgeon will perform standard suture closure for the skin layer following CIED procedure.
|
|---|---|---|
|
Patient Incision Pain
|
0.27 units on a scale
Standard Deviation 0.41
|
0.27 units on a scale
Standard Deviation 0.73
|
Adverse Events
Zip Surgical Skin Closure Device
Standard Suture Closure
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported