Trial Outcomes & Findings for Kinematic Analysis: Posterior Stabilized, Fixed Bearing Total Knee Arthroplasty With Attune Knee System (NCT NCT02204748)
NCT ID: NCT02204748
Last Updated: 2019-10-10
Results Overview
Amount of translation and lift-off for implanted knee in vivo under fluoroscopic surveillance during deep knee bend activity.
COMPLETED
5 participants
3 months post-operative
2019-10-10
Participant Flow
Participant milestones
| Measure |
DePuy Attune PS FB TKA
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Overall Study
STARTED
|
5
|
|
Overall Study
COMPLETED
|
5
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Kinematic Analysis: Posterior Stabilized, Fixed Bearing Total Knee Arthroplasty With Attune Knee System
Baseline characteristics by cohort
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
2 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
3 Participants
n=5 Participants
|
|
Age, Continuous
|
66.2 years
STANDARD_DEVIATION 8.07 • n=5 Participants
|
|
Sex: Female, Male
Female
|
4 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
1 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
5 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 3 months post-operativeAmount of translation and lift-off for implanted knee in vivo under fluoroscopic surveillance during deep knee bend activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics - Translation and Lift-off for Deep Knee Bend
Lateral Anterior Posterior Deep Knee Bend
|
-4.89 mm
Standard Deviation 3.57
|
|
Femoro-tibial Kinematics - Translation and Lift-off for Deep Knee Bend
Medial Anterior Posterior Deep Knee Bend
|
-2.58 mm
Standard Deviation 2.5
|
PRIMARY outcome
Timeframe: 3 months post-operativeAmount of translation and lift-off for implanted knee in vivo under fluoroscopic surveillance during gait activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics: Translation and Lift-off for Gait
Lateral Anterior Posterior Gait
|
-0.8 mm
Standard Deviation 1.89
|
|
Femoro-tibial Kinematics: Translation and Lift-off for Gait
Medial Anterior Posterior Gait
|
0.24 mm
Standard Deviation 0.67
|
PRIMARY outcome
Timeframe: 3 months post-operativeAmount of translation and lift-off for implanted knee in vivo under fluoroscopic surveillance during ramp down activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics: Translation and Lift-off for Ramp Down
Lateral Anterior Posterior Ramp Down
|
-1.35 mm
Standard Deviation 2.39
|
|
Femoro-tibial Kinematics: Translation and Lift-off for Ramp Down
Medial Anterior Posterior Ramp Down
|
-0.58 mm
Standard Deviation 1.72
|
PRIMARY outcome
Timeframe: 3 months post-operativeDegree of axial rotation and weight-bearing range-of-motion for implanted knee in vivo under fluoroscopic surveillance during deep knee bend activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics - Deep Knee Bend
Axial Rotation Deep Knee Bend
|
2.99 degrees
Standard Deviation 5.73
|
|
Femoro-tibial Kinematics - Deep Knee Bend
Max weight-bearing flexion Deep knee bend
|
118.2 degrees
Standard Deviation 20.96
|
PRIMARY outcome
Timeframe: 3 months post-operativeDegree of axial rotation and weight-bearing range-of-motion for implanted knee in vivo under fluoroscopic surveillance during gait activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics - Gait
Axial Rotation Gait
|
1.28 degrees
Standard Deviation 2.76
|
|
Femoro-tibial Kinematics - Gait
Max Weight-bearing flexion Gait
|
19.4 degrees
Standard Deviation 4.72
|
PRIMARY outcome
Timeframe: 3 months post-operativeDegree of axial rotation and maximum weight-bearing range-of-motion for implanted knee in vivo under fluoroscopic surveillance during ramp down activity.
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Femoro-tibial Kinematics - Ramp Down
Axial Rotation Ramp Down
|
0.97 degrees
Standard Deviation 4.32
|
|
Femoro-tibial Kinematics - Ramp Down
Max Weight-bearing flexion Ramp Down
|
28.6 degrees
Standard Deviation 5.94
|
SECONDARY outcome
Timeframe: 3 months post-operativeCollected simultaneously with fluoroscopy data, ground reaction forces were obtained using a force plate (fixed to the ground) while subject performed activity. Maximum force measured in the vertical direction measured during the described activity, then normalized with respect to participant's body weight. As such, the data are presented as the percentage of the individuals' body weight that was supported on the implanted knee using a force plate (fixed to the ground) and has been termed "maximum reaction force."
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Max Ground Reaction Force - Deep Knee Bend
|
75.3 percentage of body weight
Standard Deviation 11.2
|
SECONDARY outcome
Timeframe: 3 months post-operativeCollected simultaneously with fluoroscopy data, ground reaction forces were obtained using a force plate (fixed to the ground) while subject performed activity. Maximum force measured in the vertical direction measured during the described activity, then normalized with respect to participant's body weight. As such, the data are presented as the percentage of the individuals' body weight that was supported on the implanted knee using a force plate (fixed to the ground) and has been termed "maximum reaction force."
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Max Ground Reaction Force - Gait
|
106.0 percentage of body weight
Standard Deviation 3.0
|
SECONDARY outcome
Timeframe: 3 months post-operativeCollected simultaneously with fluoroscopy data, ground reaction forces were obtained using a force plate (fixed to the ground) while subject performed activity. Maximum force measured in the vertical direction measured during the described activity was normalized with respect to participant's body weight. As such, the data are presented as the percentage of the individuals' body weight that was supported on the implanted knee using a force plate (fixed to the ground) and has been termed "maximum reaction force."
Outcome measures
| Measure |
DePuy Attune PS FB TKA
n=5 Participants
Individuals implanted with the DePuy Attune posterior stabilizing fixed bearing knee system implanted by a single surgeon at least three months post-operative.
DePuy Attune posterior stabilizing fixed bearing knee system
|
|---|---|
|
Max Ground Reaction Force - Ramp Down
|
97.2 percentage of body weight
Standard Deviation 15.0
|
Adverse Events
DePuy Attune PS FB TKA
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place