Trial Outcomes & Findings for The Lumen Study: Comparing Devices for Patients With Symptomatic Femoro-popliteal Chronic Total Occlusion (CTO) (NCT NCT02145299)
NCT ID: NCT02145299
Last Updated: 2016-11-04
Results Overview
Technical success, defined as the ability to facilitate complete intraluminal crossing of a CTO into the true distal lumen with a TruePath or a CROSSER device and/or any subsequent conventional guidewire, as confirmed by IVUS imaging
TERMINATED
PHASE2
8 participants
Day of operation
2016-11-04
Participant Flow
Participant milestones
| Measure |
TruePath CTO Device
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Overall Study
STARTED
|
6
|
2
|
|
Overall Study
COMPLETED
|
6
|
2
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
The Lumen Study: Comparing Devices for Patients With Symptomatic Femoro-popliteal Chronic Total Occlusion (CTO)
Baseline characteristics by cohort
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
Total
n=8 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
69.0 years
STANDARD_DEVIATION 7.1 • n=5 Participants
|
64.0 years
STANDARD_DEVIATION 11.0 • n=7 Participants
|
67.8 years
STANDARD_DEVIATION 9.1 • n=5 Participants
|
|
Sex: Female, Male
Female
|
1 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
1 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
5 Participants
n=5 Participants
|
2 Participants
n=7 Participants
|
7 Participants
n=5 Participants
|
PRIMARY outcome
Timeframe: Day of operationTechnical success, defined as the ability to facilitate complete intraluminal crossing of a CTO into the true distal lumen with a TruePath or a CROSSER device and/or any subsequent conventional guidewire, as confirmed by IVUS imaging
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Technical Success
|
50 percentage of participants
|
0 percentage of participants
|
PRIMARY outcome
Timeframe: Operation through 30 day follow upIn-hospital safety, defined as a composite of all-cause death, index limb amputation above the ankle, and target lesion revascularization (TLR)
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
In-hospital Safety
|
0 percentage of participants
|
0 percentage of participants
|
SECONDARY outcome
Timeframe: Day of operationPopulation: Due to early study termination for reason unrelated to safety, rather enrollment challenges, part the Procedural Success definition required % residual stenosis which was going to be a Core Laboratory assessment to eliminate bias and maintain consistency in the analysis. With only 8 subjects enrolled, laboratory analysis was not undertaken.
Procedural success, defined as technical success and (1) residual stenosis \<50% in the treated segment (2) and improved distal flow by angiography following the procedure
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Day of operationPopulation: These data were not fully captured and summarized as the study was terminated.
Clinical Success defined as procedure success in the absence of in-hospital all-cause death, index limb amputation above the ankle, and TLR.
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Baseline, and 30 days post operationSymptomatic improvement, as assessed by change in Rutherford Class from baseline to 30 days
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Symptomatic Improvement
|
50 percentage of participants
|
50 percentage of participants
|
SECONDARY outcome
Timeframe: Baseline and 30 days post operationPopulation: These data were not fully captured and summarized as the study was terminated.
Change in walking capacity from baseline to 30 days, measured by the Walking Impairment Questionnaire. The questionaire is a subjective measure of patient-perceived walking performance developed for individuals with peripheral arterial disease. Used to evaluate the "change in walking capacity" study endpoint.
Outcome measures
Outcome data not reported
SECONDARY outcome
Timeframe: Day of Operation through 30 days post operationNeed for limb amputation
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Index Limb Amputation
|
0 participants
|
0 participants
|
SECONDARY outcome
Timeframe: 30 days post operationDescribes the percentage of patients that had stented lesions that had to be re-treated due to clinically-driven restenosis.
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Target Lesion Revascularization
|
0 percentage of participants
|
0 percentage of participants
|
SECONDARY outcome
Timeframe: baseline to 30 days post operationThe ratio of systolic blood pressure at the ankle to systolic blood pressure in the arm
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Ankle-brachial Index (ABI)
Baseline
|
0.63 ratio
Standard Deviation 0.18
|
0.75 ratio
Standard Deviation 0.19
|
|
Ankle-brachial Index (ABI)
30 days
|
0.68 ratio
Standard Deviation 0.24
|
0.93 ratio
Standard Deviation 0.07
|
SECONDARY outcome
Timeframe: 30 days post operationA repeat revascularization procedure (percutaneous or surgical) of the index procedure target vessel. TVR is classified as clinically-driven if the repeat intervention is driven by clinical findings (ischemic symptoms).
Outcome measures
| Measure |
TruePath CTO Device
n=6 Participants
The TruePath™ CTO Device (Boston Scientific Corporation, Natick, MA) is a new solution for intraluminal treatment of chronic total occlusions (CTO). It is the longest available crossing device (165 cm), and has a diamond-coated distal tip that can rotate at 13,000 rpm. Its profile is similar to a 0.018" guidewire, and includes a shapeable distal tip allowing 1:1 torque response. In addition, it provides audio and visual navigation during CTO crossing.
TruePath CTO Device: The TruePath CTO Device is composed of a 0.018" guidewire and a motor housing with a connector cable along with a sterile, disposable battery-powered Control Unit for manipulation of the device during operation. The TruePath CTO Device is indicated to facilitate the intra-luminal placement of conventional guidewires beyond peripheral artery chronic total occlusions.
|
CROSSER CTO Device
n=2 Participants
The CrosserTM CTO Recanalization Catheter ("Crosser system") (Bard Peripheral Vascular Inc. Tempe, AZ, USA), which serves as a control in this investigation, gained U.S. FDA approval for peripheral indications in 2011. The device is similar in both design and indications to the TruePath device, with the exception that the Crosser system uses vibrational angioplasty to achieve CTO crossing.
CROSSER CTO Device
|
|---|---|---|
|
Target Vessel Revascularization
|
16.7 percentage of participants
|
0 percentage of participants
|
SECONDARY outcome
Timeframe: Day of operationPopulation: These data were not fully captured and summarized as the study was terminated.
Evaluated in-hospital. The occurrence of any extravasation of contrast during the procedure (detected by the physician performing the procedure, or preferentially the Angiographic Core Laboratory) will be tabulated according to the standard Type 1-3 classification. Type 1 - Extraluminal crater without contrast extravasation * Type 2 - Perivascular blush without contrast jet extravasation * Type 3 - Contrast jet extravasation through frank (≥1 mm) perforation
Outcome measures
Outcome data not reported
Adverse Events
TruePath CTO Device
CROSSER CTO Device
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place