Trial Outcomes & Findings for Multiple Treatment Study Using Low Field Magnetic Stimulation for Bipolar Depression (NCT NCT01929681)
NCT ID: NCT01929681
Last Updated: 2017-09-26
Results Overview
The Montgomery-Asberg Depression Rating scale assesses 10 symptom areas of depression during an interview. Each of the 10 items received a score ranging from 0 to 6. The scale has a range of 0-60 for the reported total. A higher score indicates increased depression for all items and for the total. This outcome measure is the change from pretreatment baseline acquired at the screening visit to a follow-up visit 7 days after the first treatment.
COMPLETED
NA
53 participants
Variable based on screening visit schedule, > 2 weeks.
2017-09-26
Participant Flow
Recruitment was performed through newspaper advertisement, posting, and referral
Subjects that failed exclusion criteria for variable criteria such as medications change or drug use were permitted to reapply for participation after a washout period of at least 1 month.
Participant milestones
| Measure |
LFMS - Active Treatment
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Overall Study
STARTED
|
28
|
25
|
|
Overall Study
COMPLETED
|
26
|
24
|
|
Overall Study
NOT COMPLETED
|
2
|
1
|
Reasons for withdrawal
| Measure |
LFMS - Active Treatment
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Overall Study
Withdrawal by Subject
|
1
|
0
|
|
Overall Study
Physician Decision
|
1
|
1
|
Baseline Characteristics
Multiple Treatment Study Using Low Field Magnetic Stimulation for Bipolar Depression
Baseline characteristics by cohort
| Measure |
LFMS - Active Treatment
n=28 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=25 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
Total
n=53 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Continuous
|
44.5 years
STANDARD_DEVIATION 13.5 • n=5 Participants
|
45.7 years
STANDARD_DEVIATION 10.3 • n=7 Participants
|
45.1 years
STANDARD_DEVIATION 12.0 • n=5 Participants
|
|
Sex: Female, Male
Female
|
13 Participants
n=5 Participants
|
10 Participants
n=7 Participants
|
23 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
15 Participants
n=5 Participants
|
15 Participants
n=7 Participants
|
30 Participants
n=5 Participants
|
|
Montgomery-Asberg Depression Rating Scale (MADRS)
|
28.2 units on a scale
STANDARD_DEVIATION 5.8 • n=5 Participants
|
27.4 units on a scale
STANDARD_DEVIATION 5.0 • n=7 Participants
|
27.8 units on a scale
STANDARD_DEVIATION 5.4 • n=5 Participants
|
PRIMARY outcome
Timeframe: Variable based on screening visit schedule, > 2 weeks.Population: Subjects completing 3 treatments and 1 visit
The Montgomery-Asberg Depression Rating scale assesses 10 symptom areas of depression during an interview. Each of the 10 items received a score ranging from 0 to 6. The scale has a range of 0-60 for the reported total. A higher score indicates increased depression for all items and for the total. This outcome measure is the change from pretreatment baseline acquired at the screening visit to a follow-up visit 7 days after the first treatment.
Outcome measures
| Measure |
LFMS - Active Treatment
n=26 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Long Term Change: Montgomery-Åsberg Depression Rating Scale (MADRS)
|
-14.03 units on a scale
Standard Deviation 7.85
|
-9.75 units on a scale
Standard Deviation 9.32
|
PRIMARY outcome
Timeframe: 90 minutes; change immediately (<30min) prior to the first treatment and immediately after (<30min) the first treatment.Population: subjects completing 3 treatment visits and 1 followup visit
PANAS consists of 10 positive and 10 negative valence word items. Items are rated by the participant to indicate their assessment of how they are feeling about this item "right now" on a scale of 1(slightly) or not at all through 5 (extremely). 10 of of the items form a positive affect subscale, in which a higher score indicates increased positive affect, with a subscale range of 10 to 50. 10 of the items form a negative affect subscale, in which a higher score indicates more negative affect, with a subscale range of 10 to 50. This outcome measure is change in the 10 item positive affect subscale. This outcome measure is the difference between the scale administered immediately (\<30min) prior to the first treatment and immediately after (\<30min) the first treatment.
Outcome measures
| Measure |
LFMS - Active Treatment
n=26 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Change Over First Treatment: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
|
0.5 units on a scale
Standard Deviation 5.8
|
1.4 units on a scale
Standard Deviation 5.32
|
SECONDARY outcome
Timeframe: Over 3 days of treatmentPopulation: Subjects completing 3 treatments and 1 visit
PANAS consists of 10 positive and 10 negative valence word items. Items are rated by the participant to indicate their assessment of how they are feeling about this item "right now" on a scale of 1(slightly) or not at all through 5 (extremely). 10 of of the items form a positive affect subscale, in which a higher score indicates increased positive affect, with a subscale range of 10 to 50. 10 of the items form a negative affect subscale, in which a higher score indicates more negative affect, with a subscale range of 10 to 50. This outcome measure uses the 10 item positive affect subscale. This outcome measure is the least squares fit of a linear coefficient to the three post (\<30min oost treatment) minus pre (\<30min pre treatment) score differences for the three treatment days
Outcome measures
| Measure |
LFMS - Active Treatment
n=26 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Rate of of Change in Daily Improvement Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 2 mean
|
-0.63 units on a scale
Standard Error 0.68
|
0.21 units on a scale
Standard Error 0.91
|
|
Rate of of Change in Daily Improvement Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
linear coefficient
|
-0.471 units on a scale
Standard Error 0.594
|
-0.375 units on a scale
Standard Error 0.648
|
|
Rate of of Change in Daily Improvement Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 1 mean
|
0.50 units on a scale
Standard Error 1.14
|
1.40 units on a scale
Standard Error 1.09
|
|
Rate of of Change in Daily Improvement Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 3 mean
|
-0.44 units on a scale
Standard Error 0.65
|
0.65 units on a scale
Standard Error 0.77
|
SECONDARY outcome
Timeframe: Over 3 days of treatmentPopulation: Subjects completing 3 treatments and 1 visit
The Montgomery-Asberg Depression Rating scale assesses 10 symptom areas of depression during an interview. Each of the 10 items received a score ranging from 0 to 6. The scale has a range of 0-60 for the reported total. A higher score indicates increased depression for all items and for the total. This outcome measure is the least squares fit of a linear coefficient to the three pretreatment scores for the three treatment days
Outcome measures
| Measure |
LFMS - Active Treatment
n=26 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Rate of Change Over 3 Treatments: Montgomery-Asberg Depression Rating Scale (MADRS)
day 2 mean
|
16.81 units on a scale
Standard Error 1.31
|
17.83 units on a scale
Standard Error 1.45
|
|
Rate of Change Over 3 Treatments: Montgomery-Asberg Depression Rating Scale (MADRS)
linear coefficient
|
-5.42 units on a scale
Standard Error 1.07
|
-5.60 units on a scale
Standard Error 1.07
|
|
Rate of Change Over 3 Treatments: Montgomery-Asberg Depression Rating Scale (MADRS)
day 1 mean
|
26.15 units on a scale
Standard Error 1.47
|
26.13 units on a scale
Standard Error 1.44
|
|
Rate of Change Over 3 Treatments: Montgomery-Asberg Depression Rating Scale (MADRS)
day 3 mean
|
15.31 units on a scale
Standard Error 1.69
|
14.92 units on a scale
Standard Error 1.66
|
SECONDARY outcome
Timeframe: Over 3 days of treatmentPopulation: Subjects completing 3 treatments and 1 visit
PANAS consists of 10 positive and 10 negative valence word items. Items are rated by the participant to indicate their assessment of how they are feeling about this item "right now" on a scale of 1(slightly) or not at all through 5 (extremely). 10 of of the items form a positive affect subscale, in which a higher score indicates increased positive affect, with a subscale range of 10 to 50. 10 of the items form a negative affect subscale, in which a higher score indicates more negative affect, with a subscale range of 10 to 50. This outcome measure uses the 10 item positive affect subscale. This outcome measure is the least squares fit of a linear coefficient to the three pretreatment scores for the three treatment days
Outcome measures
| Measure |
LFMS - Active Treatment
n=26 Participants
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 Participants
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Rate of Change Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
linear coefficient
|
1.70 units on a scale
Standard Error 1.39
|
0.80 units on a scale
Standard Error 1.10
|
|
Rate of Change Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 2 mean
|
24.04 units on a scale
Standard Error 1.95
|
23.54 units on a scale
Standard Error 1.53
|
|
Rate of Change Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 3 mean
|
24.40 units on a scale
Standard Error 2.21
|
23.85 units on a scale
Standard Error 1.84
|
|
Rate of Change Over 3 Treatments: Positive and Negative Affect Schedule (PANAS) Positive Items Subscale
day 1 mean
|
21.00 units on a scale
Standard Error 1.82
|
22.25 units on a scale
Standard Error 1.34
|
Adverse Events
LFMS - Active Treatment
LFMS - Sham Treatment
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
LFMS - Active Treatment
n=26 participants at risk
Low Field Magnetic Stimulation (LFMS) active treatment
Active low field magnetic stimulation treatment applied with the LFMS Device; the device is on and magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
LFMS - Sham Treatment
n=24 participants at risk
Low Field Magnetic Stimulation - sham treatment
Inactive low field magnetic stimulation (no stimulation) treatment applied with the LFMS Device; the device is on, however no magnetic field stimulation is present.
Low Field Magnetic Stimulation: Low Field Magnetic Stimulation is an electromagnetic technique. It uses low strength electric fields operating at a high frequency.
|
|---|---|---|
|
Psychiatric disorders
insomnia
|
3.8%
1/26 • Number of events 1 • Adverse events data was collected over a period of 3 years.
|
0.00%
0/24 • Adverse events data was collected over a period of 3 years.
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place