Trial Outcomes & Findings for Pilot Study on Ambulatory Intraocular Pressure and Blood Pressure Monitoring in Glaucoma (NCT NCT01912599)
NCT ID: NCT01912599
Last Updated: 2016-06-17
Results Overview
The blood pressure value shown in the table is the mean of all measurements taken over the 24 hour period, so it is the average value.
COMPLETED
NA
36 participants
24 Hours
2016-06-17
Participant Flow
Participant milestones
| Measure |
Non-Glaucomatous
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Overall Study
STARTED
|
20
|
16
|
|
Overall Study
COMPLETED
|
20
|
15
|
|
Overall Study
NOT COMPLETED
|
0
|
1
|
Reasons for withdrawal
| Measure |
Non-Glaucomatous
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Overall Study
Withdrawal by Subject
|
0
|
1
|
Baseline Characteristics
Pilot Study on Ambulatory Intraocular Pressure and Blood Pressure Monitoring in Glaucoma
Baseline characteristics by cohort
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=16 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Total
n=36 Participants
Total of all reporting groups
|
|---|---|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
18 Participants
n=5 Participants
|
8 Participants
n=7 Participants
|
26 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
2 Participants
n=5 Participants
|
8 Participants
n=7 Participants
|
10 Participants
n=5 Participants
|
|
Age, Continuous
|
50.2 years
STANDARD_DEVIATION 13.3 • n=5 Participants
|
66.9 years
STANDARD_DEVIATION 11.1 • n=7 Participants
|
57.3 years
STANDARD_DEVIATION 14.8 • n=5 Participants
|
|
Sex: Female, Male
Female
|
10 Participants
n=5 Participants
|
7 Participants
n=7 Participants
|
17 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
10 Participants
n=5 Participants
|
9 Participants
n=7 Participants
|
19 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
1 Participants
n=5 Participants
|
1 Participants
n=7 Participants
|
2 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
6 Participants
n=5 Participants
|
8 Participants
n=7 Participants
|
14 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
13 Participants
n=5 Participants
|
7 Participants
n=7 Participants
|
20 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
20 participants
n=5 Participants
|
16 participants
n=7 Participants
|
36 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 24 HoursPopulation: Data for two glaucoma patients were excluded for analysis because * In one patient the IOP recorder malfunctioned just two hours after set up, so data were not available for the entire 24 hours * In the other patient, the wireless sensor was disconnected from the recorder in the middle if the night so data could not be recorded.
The blood pressure value shown in the table is the mean of all measurements taken over the 24 hour period, so it is the average value.
Outcome measures
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=13 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Arterial Pressure
|
100.9 mmHg
Standard Deviation 23.6
|
95.5 mmHg
Standard Deviation 11.4
|
PRIMARY outcome
Timeframe: 24 HoursThe IOP provided is the average of all 144 measurements taken during the 24 hour period.
Outcome measures
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=15 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Intraocular Pressure
|
17.6 mEqv.
Standard Deviation 126.5
|
179.2 mEqv.
Standard Deviation 87.2
|
PRIMARY outcome
Timeframe: 24 HoursThe value shown in the table is the mean of all measurements taken over the 24 hour period, so it is the average value.
Outcome measures
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=15 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Mean Perfusion Pressure
|
90.8 mmHg
Standard Deviation 135.6
|
62.4 mmHg
Standard Deviation 12.7
|
PRIMARY outcome
Timeframe: 24 HoursThe value shown in the table is the mean of all measurements taken over the 24 hour period, so it is the average value.
Outcome measures
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=15 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Diastolic Pressure
|
85.7 mmHg
Standard Deviation 9.3
|
74.6 mmHg
Standard Deviation 11.6
|
PRIMARY outcome
Timeframe: 24 HoursThe value shown in the table is the mean of all measurements taken over the 24 hour period, so it is the average value.
Outcome measures
| Measure |
Non-Glaucomatous
n=20 Participants
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=15 Participants
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Systolic Pressure
|
146.4 mmHg
Standard Deviation 15.2
|
129.0 mmHg
Standard Deviation 35.2
|
Adverse Events
Non-Glaucomatous
Glaucoma
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Non-Glaucomatous
n=20 participants at risk
Patients with no history of glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
Glaucoma
n=16 participants at risk
Patients that are currently being treated for moderate to severe normal-tension glaucoma. Subjects will be fitted with Sensimed Triggerfish for 24 observation of IOP and with a blood pressure monitor for 24 hour observation of blood pressure trends.
Sensimed Triggerfish: IOP will be measured with the Sensimed Triggerfish, which is a disposable silicone contact lens embedded with a wireless sensor. The sensing resistive gauge in the device is circular around the center of the lens and is placed over a circumference of 11.5 mm in diameter. This corresponds to the corneoscleral junction position, where maximum corneal deformation occurs as a result of changes in IOP. A soft patch containing the receiving antenna will be applied around the eye and transmits the information via wire to the recorder that the patient will wear around the waist. The patient can continue to wear spectacles during monitoring. The device records a total of 144 measurements over a 24-hour period.
|
|---|---|---|
|
Eye disorders
Corneal Changes
|
30.0%
6/20 • Number of events 6
|
18.8%
3/16 • Number of events 3
|
|
Eye disorders
Conjunctival Changes
|
5.0%
1/20 • Number of events 1
|
18.8%
3/16 • Number of events 3
|
|
Skin and subcutaneous tissue disorders
Cuff Abrasion
|
0.00%
0/20
|
6.2%
1/16 • Number of events 1
|
Additional Information
Jean-Claude Mwanza, MD, MPH, PhD
University of North Carolina
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place