Trial Outcomes & Findings for Study of Clinical Performance of Corneal Topography Measurement Using an Optical Biometer (NCT NCT01861925)
NCT ID: NCT01861925
Last Updated: 2014-06-05
Results Overview
Corneal topography is a measurement of the shape of the anterior cornea. The shape of a cornea can be fully quantified by providing a map of local power. "Diopter" is the unit of refractive power of a lens. In case of the cornea, the power K \[diopter\] is related to the radius (curvature) R \[mm\] of the best fitting sphere by the relation K=337.5/R. Here, corneal topography measurements are implemented by the Placido method, i.e. by analyzing the reflection image of a ring-shaped illumination. According to International Standards Organization (ISO) 19980-2012, repeatability of corneal topography is assessed on the central cornea: area with diameter d\<=3mm, and middle cornea: 3mm\<d\<=6mm. "power difference (2 rep. meas.)": sample mean and 95% C.I. of differences of spatial mean of corneal power between two consecutive measurements. "power difference 2 std.dev.": sample mean of twice the deviation of local power differences (two consecutive measurements).
COMPLETED
50 participants
1 day of examination
2014-06-05
Participant Flow
Participant milestones
| Measure |
Normal Eye
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Overall Study
STARTED
|
34
|
8
|
8
|
|
Overall Study
COMPLETED
|
34
|
8
|
8
|
|
Overall Study
NOT COMPLETED
|
0
|
0
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Study of Clinical Performance of Corneal Topography Measurement Using an Optical Biometer
Baseline characteristics by cohort
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
Total
n=50 Participants
Total of all reporting groups
|
|---|---|---|---|---|
|
Age, Continuous
|
38.8 years
STANDARD_DEVIATION 9.8 • n=5 Participants
|
36.4 years
STANDARD_DEVIATION 9.0 • n=7 Participants
|
32.7 years
STANDARD_DEVIATION 8.2 • n=5 Participants
|
37.4 years
STANDARD_DEVIATION 9.5 • n=4 Participants
|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
0 Participants
n=4 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
34 Participants
n=5 Participants
|
8 Participants
n=7 Participants
|
8 Participants
n=5 Participants
|
50 Participants
n=4 Participants
|
|
Age, Categorical
>=65 years
|
0 Participants
n=5 Participants
|
0 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
0 Participants
n=4 Participants
|
|
Sex: Female, Male
Female
|
13 Participants
n=5 Participants
|
2 Participants
n=7 Participants
|
0 Participants
n=5 Participants
|
15 Participants
n=4 Participants
|
|
Sex: Female, Male
Male
|
21 Participants
n=5 Participants
|
6 Participants
n=7 Participants
|
8 Participants
n=5 Participants
|
35 Participants
n=4 Participants
|
|
Region of Enrollment
Switzerland
|
34 participants
n=5 Participants
|
8 participants
n=7 Participants
|
8 participants
n=5 Participants
|
50 participants
n=4 Participants
|
PRIMARY outcome
Timeframe: 1 day of examinationCorneal topography is a measurement of the shape of the anterior cornea. The shape of a cornea can be fully quantified by providing a map of local power. "Diopter" is the unit of refractive power of a lens. In case of the cornea, the power K \[diopter\] is related to the radius (curvature) R \[mm\] of the best fitting sphere by the relation K=337.5/R. Here, corneal topography measurements are implemented by the Placido method, i.e. by analyzing the reflection image of a ring-shaped illumination. According to International Standards Organization (ISO) 19980-2012, repeatability of corneal topography is assessed on the central cornea: area with diameter d\<=3mm, and middle cornea: 3mm\<d\<=6mm. "power difference (2 rep. meas.)": sample mean and 95% C.I. of differences of spatial mean of corneal power between two consecutive measurements. "power difference 2 std.dev.": sample mean of twice the deviation of local power differences (two consecutive measurements).
Outcome measures
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
power difference (2 rep. meas.), central cornea
|
-0.014 diopters
Interval -0.062 to 0.034
|
0.053 diopters
Interval -0.061 to 0.167
|
-0.120 diopters
Interval -0.392 to 0.153
|
|
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
power difference (2 rep. meas.), middle cornea
|
-0.029 diopters
Interval -0.09 to 0.033
|
0.015 diopters
Interval -0.149 to 0.179
|
-0.089 diopters
Interval -0.351 to 0.174
|
|
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
power difference (2 std.dev.), central cornea
|
0.77 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
0.74 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
1.36 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
|
In-vivo Repeatability of Corneal Topography Measurements With Lenstar LS 900 Topography: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences Between Two Consecutive Measurements
power difference (2 std.dev.), middle cornea
|
0.72 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
0.69 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
1.06 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
SECONDARY outcome
Timeframe: 1 day of examinationFor definition of corneal topography, corneal power in "diopter", areas of evaluation and methods, see outcome 1. This outcome measure aims at testing the equivalence of corneal topography measurement between Lenstar LS 900 Topography (Haag Streit) and Atlas 9000 (Zeiss) by analyzing differences in measurement results for the same eye between both devices. "power difference (2 devices)" is the sample mean and 95% C.I. of differences of spatial mean of corneal power between two devices. This value quantifies systematic differences between devices (e.g. calibration), ignoring local variations of the power measurement. "power difference (2 std.dev.)" is the sample mean of twice the standard deviation of local power differences between measurements with both devices. This value quantifies the spatially resolved agreement of corneal shape measurement of the two devices.
Outcome measures
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
power difference (2 devices), central cornea
|
0.044 diopters
Interval -0.023 to 0.11
|
0.012 diopters
Interval -0.039 to 0.063
|
0.160 diopters
Interval -0.57 to 0.889
|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
power difference (2 std.dev.), central cornea
|
0.79 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
0.77 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
2.49 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
power difference (2 devices), middle cornea
|
0.070 diopters
Interval -0.004 to 0.144
|
0.038 diopters
Interval -0.058 to 0.134
|
-0.083 diopters
Interval -0.701 to 0.536
|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Differences of Mean Power and Sample Mean of Std. Dev of Local Power Differences for One Measurement Per Device.
power difference (2 std.dev.), middle cornea
|
0.76 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
0.72 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
1.56 diopters
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 1 day of examinationFor definition of corneal topography, areas of evaluation and methods, see outcome 1 and 2. Corneal elevation refers to the distance between the measured corneal surface and the best fitting sphere, and is given in µm. This outcome measure aims at testing the equivalence of corneal topography measurement between Lenstar LS 900 Topography (Haag Streit) and Atlas 9000 (Zeiss) by analyzing differences in measurement results for the same eye between both devices. "elevation difference (2 std. dev.)" is the sample mean of twice the standard deviation of local corneal elevation differences between measurements with both devices. This value quantifies the spatially resolved agreement of corneal shape measurement of the two devices.
Outcome measures
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Std. Dev. of Local Corneal Elevation Differences for One Measurement Per Device.
elevation difference (2 std. dev.), middle cornea
|
2.92 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
3.08 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
12.59 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
|
Equivalence of Corneal Topography Measurement Between Lenstar LS 900 Topography and Atlas 9000: Sample Mean of Std. Dev. of Local Corneal Elevation Differences for One Measurement Per Device.
elevation difference (2 std. dev.), centr. cornea
|
2.00 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
1.84 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
5.72 micrometer
95% C.I. of Std.Dev. not required by ISO 19980-2012
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 1 day of examinationKeratometry radius refers to the corneal curvature R (see primary measure outcome). In this context, the cornea is approximated by a toric surface which can be characterized by a flat meridian (radius R1) and a steep meridian (radius R2) with an angle of 90 degrees between these meridians. This outcome measure aims at testing the equivalence of keratometry radius measurement between Lenstar LS 900 Topography and Lenstar LS 900 (both Haag Streit) by analyzing differences in measurement results for the same eye between both devices. Reported are: population mean of difference and 95% confidence interval of mean difference for radius of flat meridian (R1) and radius of steep meridian (R2).
Outcome measures
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Equivalence of Keratometry Radius Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Differences of Keratometry Radius Measurement Between Both Devices.
difference of flat meridian (R1)
|
2.00 micrometer
Interval -3.39 to 7.38
|
-2.59 micrometer
Interval -21.98 to 16.8
|
-28.06 micrometer
Interval -61.45 to 5.33
|
|
Equivalence of Keratometry Radius Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Differences of Keratometry Radius Measurement Between Both Devices.
difference of steep meridian (R2)
|
-5.23 micrometer
Interval -12.95 to 2.49
|
-1.63 micrometer
Interval -12.57 to 9.32
|
-59.19 micrometer
Interval -131.29 to 12.92
|
OTHER_PRE_SPECIFIED outcome
Timeframe: 1 day of examinationKeratometry axis refers to the axis of the flat meridian of the toric representation of the cornea. For additional information see outcome 4. This outcome measure aims at testing the equivalence of keratometry axis measurement between Lenstar LS 900 Topography and Lenstar LS 900 (both Haag Streit) by analyzing differences in measurement results for the same eye between both devices. Reported are: population mean of difference and 95% confidence interval of mean difference for axis of flat meridian.
Outcome measures
| Measure |
Normal Eye
n=34 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
n=8 Participants
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Equivalence of Keratometry Axis Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Differences of Keratometry Axis Measurement Between Both Devices.
|
-0.13 degree
Interval -1.85 to 1.59
|
-0.44 degree
Interval -2.02 to 1.14
|
0.25 degree
Interval -3.83 to 4.34
|
POST_HOC outcome
Timeframe: 1 day of examinationFor definition of keratometry axis see outcome measures 4 and 5. This outcome measure aims at testing the equivalence of keratometry axis measurement between Lenstar LS 900 Topography and Lenstar LS 900 (both Haag Streit) by analyzing differences in measurement results for the same eye between both devices. Reported are: population mean of normalized difference and 95% confidence interval of mean normalized difference for axis of flat meridian. Difference is normalized for each eye as function of astigmatism to a refractive error of 0.167 diopters (i.e. an axis difference of 1 normalized degree results in a refractive error of 0.167 diopter), to provide a measure which can be directly related to its impact on visual quality. Please note that for this analysis a separation in arms "normal eye" and "large regular astigmatism" is not meaningful, thus here both groups are analyzed jointly as group "regular eye"
Outcome measures
| Measure |
Normal Eye
n=42 Participants
Astigmatism smaller than 1.5 diopters
|
Large Regular Astigmatism
Astigmatism of \> 1.5 diopters, regular astigmatism.
|
Large Irregular Astigmatism
Astigmatism of \> 1.5 diopters, irregular astigmatism.
|
|---|---|---|---|
|
Equivalence of Keratometry Axis Measurement Between Lenstar LS 900 Topography and Lenstar LS 900: Population Mean of Normalized Differences of Keratometry Axis Measurement Between Both Devices.
|
-0.01 normalized degree
Interval -0.29 to 0.27
|
—
|
—
|
Adverse Events
Normal Eye
Large Regular Astigmatism
Large Irregular Astigmatism
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Dr. Lucio Robledo, Clinical Trials Biometry
Haag Streit AG
Results disclosure agreements
- Principal investigator is a sponsor employee The sponsor can review results communications prior to public release and can embargo communications regarding trial results for a period that is less than or equal to 60 days from the time submitted to the sponsor for review. The sponsor can require an extension of the embargo for 90 days. If the communication contains confidential information of the sponsor, the sponsor can require to replace this information by scientifically meaningful equivalent information.
- Publication restrictions are in place
Restriction type: OTHER