Trial Outcomes & Findings for Evaluation of an Investigational Blood Glucose Monitoring System (NCT NCT01859494)
NCT ID: NCT01859494
Last Updated: 2017-02-10
Results Overview
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma).
COMPLETED
NA
219 participants
1 hour
2017-02-10
Participant Flow
Participant milestones
| Measure |
Users of the Monitoring System
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Overall Study
STARTED
|
219
|
|
Overall Study
COMPLETED
|
218
|
|
Overall Study
NOT COMPLETED
|
1
|
Reasons for withdrawal
| Measure |
Users of the Monitoring System
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Overall Study
Adverse Event
|
1
|
Baseline Characteristics
Evaluation of an Investigational Blood Glucose Monitoring System
Baseline characteristics by cohort
| Measure |
Users of the Monitoring System
n=221 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Age, Continuous
|
51.8 years
n=5 Participants
|
|
Gender
Female
|
108 Participants
n=5 Participants
|
|
Gender
Male
|
113 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
22 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
197 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
2 Participants
n=5 Participants
|
|
Race/Ethnicity, Customized
American Indian or Alaska Native
|
4 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Asian
|
9 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Native Hawaiian or Other Pacific Islander
|
2 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Black or African American
|
10 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
White
|
194 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
More than one race
|
3 participants
n=5 Participants
|
|
Race/Ethnicity, Customized
Unknown or Not Reported
|
5 participants
n=5 Participants
|
|
Region of Enrollment
United States
|
221 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1 hourPopulation: 218 (219-1) Blood glucose results were analyzed. One subject discontinued from testing after low Blood Glucose (BG) fingerstick result (hypoglycemia AE), thus no reference result available for that subject.
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI Analyzer BG results were used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Users of the Monitoring System
n=218 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG) Results Within +/-15mg/dL (<75mg/dL) or Within +/-20% (>=75mg/dL) of Laboratory Glucose Method
|
218 Blood Glucose results within 15mg/dL/20%
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 218 (219-1) Blood glucose results were analyzed. One subject discontinued from testing after low BG fingerstick result (hypoglycemia AE), thus no reference result available for that subject.
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI BG results were used to calculate the number of BGMS results within +/- 15mg/dL (\<100mg/dL YSI capillary plasma) or +/- 15% (\>=100mg/dL YSI capillary plasma)
Outcome measures
| Measure |
Users of the Monitoring System
n=218 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG)Results Within +/- 15mg/dL (<100mg/L YSI) or Within +/- 15% (>=100mg/dL YSI) of Laboratory Glucose Method
|
215 Blood Glucose results within 15mg/dL/15%
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 218 (219-1) Blood glucose results were analyzed. One subject discontinued from testing after low BG fingerstick result (hypoglycemia AE), thus no reference result available for that subject.
Untrained subjects with diabetes self-tested fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI BG results were used to calculate the number of BGMS results within +/- 12.5mg/dL (\<100mg/dL YSI capillary plasma) or +/- 12.5% (\>=100mg/dL YSI capillary plasma)
Outcome measures
| Measure |
Users of the Monitoring System
n=218 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Self-Test Fingerstick Blood Glucose (BG)Results Within +/- 12.5mg/dL (<100mg/L YSI) or Within +/- 12.5% (>=100mg/dL YSI) of Laboratory Glucose Method
|
214 Bld Glucose results w/in 12.5mg/dL/12.5%
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 211 (219-8) Blood glucose results were analyzed. One subject discontinued from all testing after hypoglycemia Adverse Event. Three subjects with low blood sugar did not attempt palm testing per protocol. Two subjects had low blood sugar; their palm results were not evaluable per protocol. Two subjects failed to obtain palm blood for testing.
Untrained subjects with diabetes self-tested Alternative Site (AST) palm blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with subject capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma BG results were used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma)
Outcome measures
| Measure |
Users of the Monitoring System
n=211 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Glucose Results From Alternative Site Testing (AST) of the Palm Within +/- 15mg/dL (<75mg/L YSI) or Within +/- 20% (>=75mg/dL YSI) of Laboratory Glucose Method
|
209 Blood Glucose results within 15mg/dL/20%
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 218 (219-1) Blood glucose results were analyzed. One subject discontinued from testing after low BG result (hypoglycemia AE), thus no reference result available for that subject.
Study staff tested subject fingerstick blood using an investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a Yellow Springs Instrument (YSI) Analyzer. YSI capillary plasma results were used to calculate the number of BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) or +/- 20% (\>=75mg/dL YSI capillary plasma)
Outcome measures
| Measure |
Users of the Monitoring System
n=218 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Subject Fingerstick BG Results Within +/- 15mg/dL (<75mg/L YSI) or Within +/- 20% (>=75mg/dL YSI) of Laboratory Glucose Method When Tested by Study Staff
|
218 Blood Glucose results within 15mg/dL/20%
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 218 (219-1) responses were analyzed. One subject discontinued from testing after low BG result (hypoglycemia Adverse Event).
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'.
Outcome measures
| Measure |
Users of the Monitoring System
n=218 Participants
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I find it easy to do a blood test with this meter.
|
218 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
The meter display was easy to see and read.
|
216 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
It was easy to understand my test results.
|
218 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I like the overall meter design.
|
215 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I find the meter easy to use.
|
217 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
The instructions were easy to understand.
|
213 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Instructions clearly explain how to run a test.
|
216 Participant Responses
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Instructions clearly explain meter error messages.
|
217 Participant Responses
|
Adverse Events
Users of the Monitoring System
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Users of the Monitoring System
n=219 participants at risk
Untrained subjects with diabetes used the NINJA 3 Investigational Blood Glucose Monitoring System.
NINJA 3 Investigational Blood Glucose Monitoring System: Untrained subjects with diabetes self-tested capillary fingerstick and palm blood using the NINJA 3 Investigational Blood Glucose Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to reference method results obtained from subject capillary plasma.
|
|---|---|
|
Endocrine disorders
Hypoglycemia
|
3.7%
8/219 • Number of events 8 • Subject visits were approximately 1 hour.
|
Additional Information
Jane Wallace, Deputy Director for Global Clinical Affairs
Ascensia Diabetes Care
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60