Trial Outcomes & Findings for Image-Guided Adaptive Radiotherapy for in Detecting Tumors During Treatment in Patients With Head and Neck Cancer (NCT NCT01843673)
NCT ID: NCT01843673
Last Updated: 2016-07-06
Results Overview
The automated patient setup procedure varies between 'OBI', 'CBCT' or 'Exactrac' imaging technologies. Each procedure gives two shifts: 'vertical' and 'lateral'. In the absence of a gold standard, our goal is to compare the shifts recommended by each pair of automated patient setup procedures. Average and Std deviation of vertical and lateral motion from the three systems were computed. P value, 1.00, refers to the test of difference of each system with OBI being more than 2 mm. Pairwise comparison for each direction between each pair of technologies were done using a t test to check if the difference in the recommended shift is more than 2 mm. The reported mean value represents the shift from planned treatment position averaged over all daily treatment setups. A negative mean vertical value indicates the patient was consistently set up posterior to plan; a negative mean lateral value indicates a set up consistently right of plan.
COMPLETED
NA
16 participants
up to 7 weeks
2016-07-06
Participant Flow
Participant milestones
| Measure |
Diagnostic (Imaging Technology)
Patients undergo FBCT once before treatment and once weekly for a total of 6-7 scans, dual CBCT up to 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian kV OBI 5 times weekly for a total of 33-35 scans, 2-D x-ray with Brain Lab ExacTrac 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian MV OBI once weekly for a total of 6-7 scans, and EPID imaging up to 5 times weekly for a total of 33-35 scans while undergoing IGART.
computed tomography: Undergo FBCT cone-beam computed tomography: Undergo dual CBCT radiography: Undergo 2-D x-ray with Varian kV OBI radiography: Undergo 2-D x-ray with Brain Lab ExacTrac radiography: Undergo 2-D x-ray with Varian MV OBI electronic portal imaging: Undergo EPID imaging image-guided adaptive radiation therapy: Undergo IGART
|
|---|---|
|
Overall Study
STARTED
|
16
|
|
Overall Study
COMPLETED
|
10
|
|
Overall Study
NOT COMPLETED
|
6
|
Reasons for withdrawal
| Measure |
Diagnostic (Imaging Technology)
Patients undergo FBCT once before treatment and once weekly for a total of 6-7 scans, dual CBCT up to 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian kV OBI 5 times weekly for a total of 33-35 scans, 2-D x-ray with Brain Lab ExacTrac 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian MV OBI once weekly for a total of 6-7 scans, and EPID imaging up to 5 times weekly for a total of 33-35 scans while undergoing IGART.
computed tomography: Undergo FBCT cone-beam computed tomography: Undergo dual CBCT radiography: Undergo 2-D x-ray with Varian kV OBI radiography: Undergo 2-D x-ray with Brain Lab ExacTrac radiography: Undergo 2-D x-ray with Varian MV OBI electronic portal imaging: Undergo EPID imaging image-guided adaptive radiation therapy: Undergo IGART
|
|---|---|
|
Overall Study
Lack of Efficacy
|
2
|
|
Overall Study
Withdrawal by Subject
|
4
|
Baseline Characteristics
Image-Guided Adaptive Radiotherapy for in Detecting Tumors During Treatment in Patients With Head and Neck Cancer
Baseline characteristics by cohort
| Measure |
Diagnostic (Imaging Technology)
n=16 Participants
Patients undergo FBCT once before treatment and once weekly for a total of 6-7 scans, dual CBCT up to 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian kV OBI 5 times weekly for a total of 33-35 scans, 2-D x-ray with Brain Lab ExacTrac 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian MV OBI once weekly for a total of 6-7 scans, and EPID imaging up to 5 times weekly for a total of 33-35 scans while undergoing IGART.
computed tomography: Undergo FBCT
cone-beam computed tomography: Undergo dual CBCT
radiography: Undergo 2-D x-ray with Varian kV OBI
radiography: Undergo 2-D x-ray with Brain Lab ExacTrac
radiography: Undergo 2-D x-ray with Varian MV OBI
electronic portal imaging: Undergo EPID imaging
image-guided adaptive radiation therapy: Undergo IGART
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
9 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
7 Participants
n=5 Participants
|
|
Age, Continuous
|
63 years
STANDARD_DEVIATION 10 • n=5 Participants
|
|
Sex: Female, Male
Female
|
0 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
16 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
0 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
16 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
2 Participants
n=5 Participants
|
|
Race (NIH/OMB)
White
|
13 Participants
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
0 Participants
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
1 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
16 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: up to 7 weeksPopulation: Ten evaluable head and neck patients treated with external beam radiation therapy received 19 to 32 fractions with all three imaging techniques. The mean daily shift of these 19 to 32 fractions were recorded for each technique. Statistical significance is determined based on 5% level of significance.
The automated patient setup procedure varies between 'OBI', 'CBCT' or 'Exactrac' imaging technologies. Each procedure gives two shifts: 'vertical' and 'lateral'. In the absence of a gold standard, our goal is to compare the shifts recommended by each pair of automated patient setup procedures. Average and Std deviation of vertical and lateral motion from the three systems were computed. P value, 1.00, refers to the test of difference of each system with OBI being more than 2 mm. Pairwise comparison for each direction between each pair of technologies were done using a t test to check if the difference in the recommended shift is more than 2 mm. The reported mean value represents the shift from planned treatment position averaged over all daily treatment setups. A negative mean vertical value indicates the patient was consistently set up posterior to plan; a negative mean lateral value indicates a set up consistently right of plan.
Outcome measures
| Measure |
Orthogonal Varian kV OBI Image
n=10 Participants
Patients are first set up for treatment with surface fiducial marks and followed by an automated image guided alignment procedure to achieve better alignment with the planning image. The automated procedure OBI gives two shifts: vertical and lateral. In the absence of a gold standard, our goal is to compare the vertical and lateral shifts recommended by each pair of automated procedures.
|
Varian CBCT (Cone Beam CT) Imaging
n=10 Participants
Patients are first set up for treatment with surface fiducial marks and followed by an automated image guided alignment procedure to achieve better alignment with the planning image. The automated procedure CBCT gives two shifts: vertical and lateral. In the absence of a gold standard, our goal is to compare the vertical and lateral shifts recommended by each pair of automated procedures.
|
Oblique Brainlab ExacTrac Images
n=10 Participants
Patients are first set up for treatment with surface fiducial marks and followed by an automated image guided alignment procedure to achieve better alignment with the planning image. The automated procedure ExacTrac gives two shifts: vertical and lateral. In the absence of a gold standard, our goal is to compare the vertical and lateral shifts recommended by each pair of automated procedures.
|
|---|---|---|---|
|
Differences of Calculated Set up Errors of 2 mm Between the Different Imaging Technologies
Vertical shift
|
-1.2 mm
Standard Deviation 3.1
|
-0.8 mm
Standard Deviation 3.4
|
-1.8 mm
Standard Deviation 3.2
|
|
Differences of Calculated Set up Errors of 2 mm Between the Different Imaging Technologies
Lateral shift
|
-1.5 mm
Standard Deviation 3.3
|
-1.1 mm
Standard Deviation 3.3
|
1.3 mm
Standard Deviation 3.9
|
PRIMARY outcome
Timeframe: Up to 7 weeksPopulation: No data analyzed due to staff leaving primary institution.
Outcome measures
Outcome data not reported
PRIMARY outcome
Timeframe: Up to 7 weeksPopulation: No data analyzed due to staff leaving primary institution.
Outcome measures
Outcome data not reported
Adverse Events
Diagnostic (Imaging Technology)
Serious adverse events
| Measure |
Diagnostic (Imaging Technology)
n=16 participants at risk
Patients undergo FBCT once before treatment and once weekly for a total of 6-7 scans, dual CBCT up to 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian kV OBI 5 times weekly for a total of 33-35 scans, 2-D x-ray with Brain Lab ExacTrac 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian MV OBI once weekly for a total of 6-7 scans, and EPID imaging up to 5 times weekly for a total of 33-35 scans while undergoing IGART.
computed tomography: Undergo FBCT
cone-beam computed tomography: Undergo dual CBCT
radiography: Undergo 2-D x-ray with Varian kV OBI
radiography: Undergo 2-D x-ray with Brain Lab ExacTrac
radiography: Undergo 2-D x-ray with Varian MV OBI
electronic portal imaging: Undergo EPID imaging
image-guided adaptive radiation therapy: Undergo IGART
|
|---|---|
|
Infections and infestations
SEPSIS-ASSOCIATED ORGAN DYSFUNCTION
|
6.2%
1/16 • Number of events 1 • Adverse Event Reporting were collected up to 7 weeks while the patient was undergoing the course of radiation therapy. No adverse events occurred due to additional xray imaging from this study.
|
Other adverse events
| Measure |
Diagnostic (Imaging Technology)
n=16 participants at risk
Patients undergo FBCT once before treatment and once weekly for a total of 6-7 scans, dual CBCT up to 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian kV OBI 5 times weekly for a total of 33-35 scans, 2-D x-ray with Brain Lab ExacTrac 5 times weekly for a total of 33-35 scans, 2-D x-ray with Varian MV OBI once weekly for a total of 6-7 scans, and EPID imaging up to 5 times weekly for a total of 33-35 scans while undergoing IGART.
computed tomography: Undergo FBCT
cone-beam computed tomography: Undergo dual CBCT
radiography: Undergo 2-D x-ray with Varian kV OBI
radiography: Undergo 2-D x-ray with Brain Lab ExacTrac
radiography: Undergo 2-D x-ray with Varian MV OBI
electronic portal imaging: Undergo EPID imaging
image-guided adaptive radiation therapy: Undergo IGART
|
|---|---|
|
Blood and lymphatic system disorders
Below normal Glomerular filtration rate (GFR)
|
12.5%
2/16 • Number of events 2 • Adverse Event Reporting were collected up to 7 weeks while the patient was undergoing the course of radiation therapy. No adverse events occurred due to additional xray imaging from this study.
|
|
Blood and lymphatic system disorders
Elevated Creatinine
|
37.5%
6/16 • Number of events 7 • Adverse Event Reporting were collected up to 7 weeks while the patient was undergoing the course of radiation therapy. No adverse events occurred due to additional xray imaging from this study.
|
|
Blood and lymphatic system disorders
Elevated Blood Urea Nitrogen (BUN)
|
18.8%
3/16 • Number of events 3 • Adverse Event Reporting were collected up to 7 weeks while the patient was undergoing the course of radiation therapy. No adverse events occurred due to additional xray imaging from this study.
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place