Trial Outcomes & Findings for Performance of an Investigational Blood Glucose Monitoring System (NCT NCT01824355)
NCT ID: NCT01824355
Last Updated: 2016-02-29
Results Overview
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<75mg/dL) and within +/- 20% (for reference BG results \>=75mg/dL) of the YSI reference method results.
COMPLETED
NA
113 participants
1 hour
2016-02-29
Participant Flow
Participant milestones
| Measure |
Intended Users of the Monitoring System
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population:
1. At least 60% of subjects were younger than 65 years of age.
2. At least 20% had type 1 diabetes.
3. At least 50% with type 2 diabetes were insulin users.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Overall Study
COMPLETED
|
110
|
|
Overall Study
NOT COMPLETED
|
3
|
|
Overall Study
STARTED
|
113
|
Reasons for withdrawal
| Measure |
Intended Users of the Monitoring System
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population:
1. At least 60% of subjects were younger than 65 years of age.
2. At least 20% had type 1 diabetes.
3. At least 50% with type 2 diabetes were insulin users.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Overall Study
Adverse Event
|
2
|
|
Overall Study
unevaluable reference method results per
|
1
|
Baseline Characteristics
Performance of an Investigational Blood Glucose Monitoring System
Baseline characteristics by cohort
| Measure |
Intended Users of the Monitoring System
n=113 Participants
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Age, Customized
Age of enrolled subjects
|
59.1 years
n=5 Participants
|
|
Sex: Female, Male
Female
|
63 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
50 Participants
n=5 Participants
|
|
Race (NIH/OMB)
American Indian or Alaska Native
|
1 responses
n=5 Participants
|
|
Race (NIH/OMB)
Asian
|
0 responses
n=5 Participants
|
|
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
|
0 responses
n=5 Participants
|
|
Race (NIH/OMB)
Black or African American
|
26 responses
n=5 Participants
|
|
Race (NIH/OMB)
White
|
84 responses
n=5 Participants
|
|
Race (NIH/OMB)
More than one race
|
1 responses
n=5 Participants
|
|
Race (NIH/OMB)
Unknown or Not Reported
|
3 responses
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Hispanic or Latino
|
10 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Not Hispanic or Latino
|
103 Participants
n=5 Participants
|
|
Ethnicity (NIH/OMB)
Unknown or Not Reported
|
0 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
113 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 1 hourPopulation: 220 Blood Glucose results were analyzed. Each subject who completed (110 subjects) the study provided 2 Blood Glucose test results.
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<75mg/dL) and within +/- 20% (for reference BG results \>=75mg/dL) of the YSI reference method results.
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=220 Blood Glucose Results
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method
|
99.09 percentage of BG results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 220 Blood Glucose results were analyzed. Each subject who completed (110 subjects) the study provided 2 Blood Glucose test results.
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<100mg/dL) and within +/- 15% (for reference BG results \>=100mg/dL) of the YSI reference method results.
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=220 Blood Glucose Results
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method
|
97.73 percentage of Blood Glucose Results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 220 Blood Glucose results were analyzed. Each subject who completed (110 subjects) the study provided 2 Blood Glucose test results.
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 12.5mg/dL (for reference BG results \<100mg/dL) and within +/- 12.5% (for reference BG results \>=100mg/dL) of the YSI reference method results.
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=220 Blood Glucose Results
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method
|
93.64 percentage of Blood Glucose Results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: 209 (220-11) Blood Glucose results were analyzed. 2 subjects had low blood sugar and per protocol their 4 AST BG results were not evaluable. 3 BG results were excluded as they were protocol deviations. One subject (2 BG results) did not complete testing within protocol timeframe and one subject (2 BG results) was unable to obtain AST blood.
Untrained subjects with diabetes self-tested Alternative Site (AST) Palm blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of AST BGMS results within +/- 15mg/dL (\<75mg/dL YSI capillary plasma) and within +/- 15% (\>=75mg/dL YSI capillary plasma).
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=209 Blood Glucose Results
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.
|
97.13 percentage of AST Blood Glucose Results
|
SECONDARY outcome
Timeframe: one hourPopulation: 220 Blood Glucose results were analyzed. Study Staff provided 2 Blood Glucose Test Results from each subject who completed (110 subjects) the study.
Study Staff tested subject fingerstick blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results \<75mg/dL) and within +/- 20% (for reference BG results \>=75mg/dL) of the YSI reference method results.
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=220 Blood Glucose Results
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff
|
100.0 percentage of Blood Glucose Results
|
SECONDARY outcome
Timeframe: 1 hourPopulation: Four subjects of those enrolled (113 subjects) had either missing or unevaluable questionnaire data.
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree.'
Outcome measures
| Measure |
Intended Users of the Monitoring System
n=109 Participants
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I find it easy to do a blood test with this meter
|
108 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
The meter display is easy to see and read
|
108 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
It was easy to understand my test results
|
108 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I like the overall meter design
|
106 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
I find the meter easy to use
|
106 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
The instructions (User Guide) easy to understand
|
105 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Instructions clearly explain how to run a test
|
107 participants
|
|
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Instructions clearly explain meter error messages
|
108 participants
|
Adverse Events
Intended Users of the Monitoring System
Serious adverse events
Adverse event data not reported
Other adverse events
| Measure |
Intended Users of the Monitoring System
n=113 participants at risk
Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population:
1. At least 60% of subjects were younger than 65 years of age.
2. At least 20% had type 1 diabetes.
3. At least 50% with type 2 diabetes were insulin users.
Ninja 3 PLUS Investigational BG Monitoring System: Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
|
|---|---|
|
Endocrine disorders
Hypoglycemia
|
0.88%
1/113 • Number of events 1 • 1 hour subject visit
|
|
Endocrine disorders
Dizzy
|
0.88%
1/113 • Number of events 1 • 1 hour subject visit
|
|
Endocrine disorders
Headache
|
0.88%
1/113 • Number of events 1 • 1 hour subject visit
|
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place
Restriction type: GT60