Trial Outcomes & Findings for Does Noninvasive Electrical Stimulation of Acupuncture Points (NESAP) Reduce Heelstick Pain in Infants? (NCT NCT01800318)

NCT ID: NCT01800318

Last Updated: 2017-01-16

Results Overview

The PIPP score includes assessment of contextual, physiological, and behavioral parameters and has been extensively validated for pain assessment in preterm and term infants. PIPP scores were given at baseline before initiation of the TENS unit,and every 30 seconds for the first two minutes of the heel stick and heel squeeze (4 times). The four PIPP scores given during heel stick and squeeze were averaged. Behavioral portion of PIPP score: facial expressions are videotaped and analyzed. Physiologic portion of PIPP score: Oxygen saturation levels and heart rates are recorded at baseline and then continuously throughout initiation of the TENS unit and the heel stick procedure. Contextual score - gestational age + sleep/wake state. Subscale scores are added for a total PIPP score. Total or composite PIPP scores are reported. Scores on the PIPP for full term infants range from 0-18, with 0 being no pain, 1-6 minimal pain, 7-12 moderate pain, 13-18 severe pain.

Recruitment status

COMPLETED

Study phase

NA

Target enrollment

162 participants

Primary outcome timeframe

Baseline and first two minutes of heel stick an squeeze. PIPP scores are given every 30 seconds for the first two minutes of the heel stick and squeeze and then averaged..

Results posted on

2017-01-16

Participant Flow

Dates of recruitment: 4/10/2013 until 12/31/2014. Types of location: Postpartal/newborn units at University of Arkansas for Medical Sciences and Lucille Packard Childrens Hospital associated with Stanford University.

All infants enrolled in the study were assigned to groups.

Participant milestones

Participant milestones
Measure
Sham NESAP With 24% Oral Sucrose
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Overall Study
STARTED
39
42
41
40
Overall Study
COMPLETED
37
37
40
39
Overall Study
NOT COMPLETED
2
5
1
1

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Does Noninvasive Electrical Stimulation of Acupuncture Points (NESAP) Reduce Heelstick Pain in Infants?

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Sham NESAP With 24% Oral Sucrose
n=39 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=42 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs. at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=41 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, small electrodes will be placed in treatment groups on the baby's legs at four specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=40 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Total
n=162 Participants
Total of all reporting groups
Age, Continuous
1 days
STANDARD_DEVIATION 0.5 • n=5 Participants
1 days
STANDARD_DEVIATION 0.5 • n=7 Participants
1 days
STANDARD_DEVIATION 0.5 • n=5 Participants
1 days
STANDARD_DEVIATION 0.5 • n=4 Participants
1 days
STANDARD_DEVIATION 0.5 • n=21 Participants
Gender
Female
15 Participants
n=5 Participants
26 Participants
n=7 Participants
16 Participants
n=5 Participants
17 Participants
n=4 Participants
74 Participants
n=21 Participants
Gender
Male
24 Participants
n=5 Participants
16 Participants
n=7 Participants
25 Participants
n=5 Participants
23 Participants
n=4 Participants
88 Participants
n=21 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
5 Participants
n=5 Participants
5 Participants
n=7 Participants
5 Participants
n=5 Participants
7 Participants
n=4 Participants
22 Participants
n=21 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
34 Participants
n=5 Participants
37 Participants
n=7 Participants
36 Participants
n=5 Participants
33 Participants
n=4 Participants
140 Participants
n=21 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
0 Participants
n=4 Participants
0 Participants
n=21 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
0 Participants
n=4 Participants
0 Participants
n=21 Participants
Race (NIH/OMB)
Asian
1 Participants
n=5 Participants
4 Participants
n=7 Participants
1 Participants
n=5 Participants
2 Participants
n=4 Participants
8 Participants
n=21 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
0 Participants
n=4 Participants
0 Participants
n=21 Participants
Race (NIH/OMB)
Black or African American
12 Participants
n=5 Participants
10 Participants
n=7 Participants
9 Participants
n=5 Participants
19 Participants
n=4 Participants
50 Participants
n=21 Participants
Race (NIH/OMB)
White
26 Participants
n=5 Participants
28 Participants
n=7 Participants
29 Participants
n=5 Participants
17 Participants
n=4 Participants
100 Participants
n=21 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
0 Participants
n=7 Participants
2 Participants
n=5 Participants
2 Participants
n=4 Participants
4 Participants
n=21 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
0 Participants
n=7 Participants
0 Participants
n=5 Participants
0 Participants
n=4 Participants
0 Participants
n=21 Participants
Participants per treatment group
39 participants
n=5 Participants
42 participants
n=7 Participants
41 participants
n=5 Participants
40 participants
n=4 Participants
162 participants
n=21 Participants

PRIMARY outcome

Timeframe: Baseline and first two minutes of heel stick an squeeze. PIPP scores are given every 30 seconds for the first two minutes of the heel stick and squeeze and then averaged..

Population: Analysis of PIPP scores assigned during heelstick procedure in newborn infants

The PIPP score includes assessment of contextual, physiological, and behavioral parameters and has been extensively validated for pain assessment in preterm and term infants. PIPP scores were given at baseline before initiation of the TENS unit,and every 30 seconds for the first two minutes of the heel stick and heel squeeze (4 times). The four PIPP scores given during heel stick and squeeze were averaged. Behavioral portion of PIPP score: facial expressions are videotaped and analyzed. Physiologic portion of PIPP score: Oxygen saturation levels and heart rates are recorded at baseline and then continuously throughout initiation of the TENS unit and the heel stick procedure. Contextual score - gestational age + sleep/wake state. Subscale scores are added for a total PIPP score. Total or composite PIPP scores are reported. Scores on the PIPP for full term infants range from 0-18, with 0 being no pain, 1-6 minimal pain, 7-12 moderate pain, 13-18 severe pain.

Outcome measures

Outcome measures
Measure
Sham NESAP With 24% Oral Sucrose
n=37 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=37 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=40 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=39 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Changes From Baseline Premature Infant Pain Profile (PIPP) Score to Average PIPP Score During Heel Stick and Squeeze.
4.0 units on a scale (PIPP score)
Standard Deviation 1.8
5.0 units on a scale (PIPP score)
Standard Deviation 4.0
3.6 units on a scale (PIPP score)
Standard Deviation 1.2
4.9 units on a scale (PIPP score)
Standard Deviation 4.0

SECONDARY outcome

Timeframe: Baseline and 5±0.5 minutes after heel stick

Salivary cortisol was obtained prior to initiation of heelstick procedure, and at 5±0.5 minutes after procedure by gentle insertion in the mouth of a soft applicator (Salimetrics Infant Swab). The samples were stored at -20 degrees, and were analyzed at UAMS.

Outcome measures

Outcome measures
Measure
Sham NESAP With 24% Oral Sucrose
n=21 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=18 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=20 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=16 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Change in Salivary Cortisol After Heel Stick
36.6 ng/ml
Standard Deviation 28.9
41.7 ng/ml
Standard Deviation 26
58.5 ng/ml
Standard Deviation 22
56.3 ng/ml
Standard Deviation 35.7

SECONDARY outcome

Timeframe: Baseline, 20 minutes +/- 5 minutes

Population: Newborn infants

Changes in Heart Rate Variability (HRV) were evaluated using the DL 900 monitor with 3-channel output with 5 leads. Premature infant leads from Braemar, Incorporated, were used with the DL 900 monitor. Leads were applied to the infant's chest before initiation of the TENS unit and the heel stick. The DL300 Holter Monitor will started recording HRV within 10 minutes of TENS unit initiation and the heel stick procedure and continued recording during the procedure and for 2 minutes afterwards.

Outcome measures

Outcome measures
Measure
Sham NESAP With 24% Oral Sucrose
n=37 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=38 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=40 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=39 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Change in Heart Rate Variability During Heel Stick
1.4 LF/HF ratio
Standard Deviation 1.2
1.5 LF/HF ratio
Standard Deviation 2
1.3 LF/HF ratio
Standard Deviation 1.2
1.3 LF/HF ratio
Standard Deviation 1.4

SECONDARY outcome

Timeframe: 10 minutes

(a) Any crying after initiation of TENS unit was noted. If the PIPP scores increased by 4 points from baseline, the TENS unit would have been turned off and the infant withdrawn from the study (safety outcome).

Outcome measures

Outcome measures
Measure
Sham NESAP With 24% Oral Sucrose
n=37 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=37 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=40 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=39 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Duration of Crying After TENS Unit Was Initiated But Before Heel Stick.
0 seconds
Standard Deviation 0
0 seconds
Standard Deviation 0
0 seconds
Standard Deviation 0
0 seconds
Standard Deviation 0

SECONDARY outcome

Timeframe: 5 minutes +/- 2 minutes

Population: Newborn infants during heel stick procedure, crying timed in seconds

Any crying during the heel stick procedure was timed in seconds.

Outcome measures

Outcome measures
Measure
Sham NESAP With 24% Oral Sucrose
n=37 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP).
NESAP With Oral Water
n=37 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. 10 minutes before the heel stick, the Empi Select TENS unit will be turned on. (b) Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed in treatment groups on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. Oral water:1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick.
NESAP With 24% Oral Sucrose
n=40 Participants
(a) NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. The Empi Select TENS unit will be turned on ten minutes before the heel stick. (b) Infant will receive 1 ml of 24% oral sucrose solution given along with a pacifier two minutes before the heel stick. NESAP: Electrical stimulation will be administered via the Empi Select, a standard TENS unit. To produce analgesia, four small electrodes will be placed on the baby's legs at specific acupuncture points. A low continuous current will be provided with minimal voltage of 3.5 mA. The frequency will be delivered using a stimulation of 10 Hz for 10±1 minutes prior to the heelstick, with continued stimulation during and for 2 minutes after the heel stick. 24% oral Sucrose: One ml 24% sucrose will be given approximately two minutes before the heel stick. Sucrose will be given via oral syringe along with a pacifier.
Sham NESAP With Oral Water
n=39 Participants
1. Sham NESAP: Four Stim Care electrodes with gel backing will be placed on the infant's lower leg at acupuncture points. However, the Empi Select TENS unit will not be turned on. The TENS unit will be hidden from investigators so that they are blinded to the status of the unit. 2. Infant will receive 1 ml of sterile water given via oral syringe along with a pacifier two minutes before the heel stick. Investigators will be blinded to the solution given. Sham NESAP: Four electrodes will be placed on the infant's lower leg, but the TENS unit will not be turned on. The TENS unit will be covered and investigators will not know whether the TENS unit is turned on or not (sham NESAP). Oral water: For infants in the control group, 1 ml of water will be given via oral syringe along with a pacifier 2 minutes before the heel stick. Investigators will be blinded on whether the infants are receiving water or oral sucrose.
Duration of Crying During Heel Stick
0.43 seconds
Standard Deviation 1.7
32 seconds
Standard Deviation 59.8
7.28 seconds
Standard Deviation 26.9
23.7 seconds
Standard Deviation 57.8

Adverse Events

Sham NESAP With 24% Oral Sucrose

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

NESAP With Oral Water

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

NESAP With 24% Oral Sucrose

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Sham NESAP With Oral Water

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported