Trial Outcomes & Findings for Radial Reload Open LAR Case Series (NCT NCT01717014)

NCT ID: NCT01717014

Last Updated: 2015-03-02

Results Overview

The surgeons ability to achieve a staple line at the desired level of the rectum.

Recruitment status

TERMINATED

Study phase

NA

Target enrollment

8 participants

Primary outcome timeframe

Operative

Results posted on

2015-03-02

Participant Flow

Participant milestones

Participant milestones
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Overall Study
STARTED
8
Overall Study
COMPLETED
8
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Radial Reload Open LAR Case Series

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Age, Continuous
55.6 years
STANDARD_DEVIATION 14.8 • n=5 Participants
Sex: Female, Male
Female
2 Participants
n=5 Participants
Sex: Female, Male
Male
6 Participants
n=5 Participants
Race (NIH/OMB)
American Indian or Alaska Native
0 Participants
n=5 Participants
Race (NIH/OMB)
Asian
0 Participants
n=5 Participants
Race (NIH/OMB)
Native Hawaiian or Other Pacific Islander
0 Participants
n=5 Participants
Race (NIH/OMB)
Black or African American
0 Participants
n=5 Participants
Race (NIH/OMB)
White
8 Participants
n=5 Participants
Race (NIH/OMB)
More than one race
0 Participants
n=5 Participants
Race (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Hispanic or Latino
0 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Not Hispanic or Latino
8 Participants
n=5 Participants
Ethnicity (NIH/OMB)
Unknown or Not Reported
0 Participants
n=5 Participants
Height
175.1 centimeters
STANDARD_DEVIATION 4.3 • n=5 Participants
Weight
88.7 kilograms
STANDARD_DEVIATION 18.0 • n=5 Participants
Body Mass Index (BMI)
29.0 kg/m2
STANDARD_DEVIATION 6.1 • n=5 Participants

PRIMARY outcome

Timeframe: Operative

The surgeons ability to achieve a staple line at the desired level of the rectum.

Outcome measures

Outcome measures
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Staple Line
7 participants

PRIMARY outcome

Timeframe: Operative

The ability to achieve adequate distal margins (defined as \>2cm \[or \>1cm with clear histologic evaluation\]) in the low rectum.

Outcome measures

Outcome measures
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Distal Margins
8 participants

SECONDARY outcome

Timeframe: Operatively

Visibility measured by surgeon usability questionnaire.

Outcome measures

Outcome measures
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Usability: Visibility
Visibility adequate when RR was used operatively
100 % of cases surgeon agree/strongly agree
Usability: Visibility
Did instrument impede visibility
0 % of cases surgeon agree/strongly agree

SECONDARY outcome

Timeframe: Operatively

Access measured by surgeon usability questionnaire

Outcome measures

Outcome measures
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Usability: Access
Radial Reload (RR) easy to place in pelvis
100 % of cases surgeon agree/strongly agree
Usability: Access
RR was able to get low in pelvis
100 % of cases surgeon agree/strongly agree
Usability: Access
RR was easy to place in the colon & rectum
100 % of cases surgeon agree/strongly agree
Usability: Access
Was rectum contained in the device after clamping
12.5 % of cases surgeon agree/strongly agree
Usability: Access
RR retained an adequate tissue during clamping
100 % of cases surgeon agree/strongly agree
Usability: Access
Was the rectum transected in one firing
12.5 % of cases surgeon agree/strongly agree

SECONDARY outcome

Timeframe: Operatively

Manueverability measured by surgeon usability questionnaire. Question: Maneuverability of Radial reload during the procedure was adequate

Outcome measures

Outcome measures
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 Participants
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Usability: Manueverability
87.5 % of cases surgeon agree/strongly agree

Adverse Events

Covidien Radial Reload Stapler With Tri-Staple Technology

Serious events: 4 serious events
Other events: 2 other events
Deaths: 0 deaths

Serious adverse events

Serious adverse events
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 participants at risk
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Gastrointestinal disorders
Ileus
25.0%
2/8 • Number of events 2 • Adverse events were collected through the 30 day follow-up visit
Infections and infestations
Pelvic abscess
12.5%
1/8 • Number of events 1 • Adverse events were collected through the 30 day follow-up visit
Infections and infestations
Perirectal abscess
12.5%
1/8 • Number of events 1 • Adverse events were collected through the 30 day follow-up visit
Renal and urinary disorders
Renal failure acute
25.0%
2/8 • Number of events 2 • Adverse events were collected through the 30 day follow-up visit
Renal and urinary disorders
Urinary retention
12.5%
1/8 • Number of events 1 • Adverse events were collected through the 30 day follow-up visit

Other adverse events

Other adverse events
Measure
Covidien Radial Reload Stapler With Tri-Staple Technology
n=8 participants at risk
Covidien Radial Reload Stapler with Tri-Staple Technology in open low anterior resection or anterior proctosigmoidectomy
Renal and urinary disorders
Urinary retention
12.5%
1/8 • Number of events 1 • Adverse events were collected through the 30 day follow-up visit
Injury, poisoning and procedural complications
Post-operative ileus
12.5%
1/8 • Number of events 1 • Adverse events were collected through the 30 day follow-up visit

Additional Information

Julie Doyle

Covidien

Phone: 781-839-1729

Results disclosure agreements

  • Principal investigator is a sponsor employee
  • Publication restrictions are in place