Trial Outcomes & Findings for Mechanics Analysis of Patients Having a Sigma Mobile Bearing TKA (NCT NCT01714492)

NCT ID: NCT01714492

Last Updated: 2019-10-10

Results Overview

The values that were reported indicate the motion of the contact point from full extension to patient's maximum flexion. Throughout flexion, if the point translated forward (anteriorly) atop the tibial tray, the number was reported as positive. If the point traveled backwards (posteriorly) atop the tibial tray, the number was reported as negative.

Recruitment status

COMPLETED

Target enrollment

8 participants

Primary outcome timeframe

10 yrs post-operative

Results posted on

2019-10-10

Participant Flow

hospital, medical clinic

Participant milestones

Participant milestones
Measure
Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Overall Study
STARTED
8
Overall Study
COMPLETED
8
Overall Study
NOT COMPLETED
0

Reasons for withdrawal

Withdrawal data not reported

Baseline Characteristics

Mechanics Analysis of Patients Having a Sigma Mobile Bearing TKA

Baseline characteristics by cohort

Baseline characteristics by cohort
Measure
Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly
n=8 Participants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Age, Categorical
<=18 years
0 Participants
n=5 Participants
Age, Categorical
Between 18 and 65 years
1 Participants
n=5 Participants
Age, Categorical
>=65 years
7 Participants
n=5 Participants
Age, Continuous
74 years
STANDARD_DEVIATION 8.6 • n=5 Participants
Sex: Female, Male
Female
6 Participants
n=5 Participants
Sex: Female, Male
Male
2 Participants
n=5 Participants
Region of Enrollment
United States
8 participants
n=5 Participants

PRIMARY outcome

Timeframe: 10 yrs post-operative

Population: Subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.

The values that were reported indicate the motion of the contact point from full extension to patient's maximum flexion. Throughout flexion, if the point translated forward (anteriorly) atop the tibial tray, the number was reported as positive. If the point traveled backwards (posteriorly) atop the tibial tray, the number was reported as negative.

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
In Vivo Linear Knee Kinematics From Fluoroscopy Evaluation During Deep Knee Bend Activity
Lateral Anterior Posterior Deep Knee Bend
-4.1 mm
Standard Deviation 2.4
In Vivo Linear Knee Kinematics From Fluoroscopy Evaluation During Deep Knee Bend Activity
Medial Anterior Posterior Deep Knee Bend
1.2 mm
Standard Deviation 2.8

PRIMARY outcome

Timeframe: 10 yrs post-operative

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Femoral Axial Rotation With Respect to the Tibia During Deep Knee Bend Activity
6.1 degrees
Standard Deviation 5.8

PRIMARY outcome

Timeframe: 10 yrs post-operative

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Range of Motion During Flexion of Deep Knee Bend Activity
103 degrees
Standard Deviation 11.2

SECONDARY outcome

Timeframe: 10 yrs post-operative

Population: Subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation

The intended unit of measure is "times Body Weight" (or xBW), relating to a ratio.

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
In Vivo Knee Force Values From Fluoroscopy Evaluation During Deep Knee Bend Activity
Medial Condyle Knee Force at Maximum Flexion
1.57 times body weight
Standard Deviation 0.53
In Vivo Knee Force Values From Fluoroscopy Evaluation During Deep Knee Bend Activity
Lateral Condyle Knee Force at Maximum Flexion
1.39 times body weight
Standard Deviation 0.36

SECONDARY outcome

Timeframe: 10 yrs post-operative

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Condyle Contact Area at Maximum Flexion During Deep Knee Bend Activity
Medial
109.16 mm^2
Standard Deviation 2.82
Condyle Contact Area at Maximum Flexion During Deep Knee Bend Activity
Lateral
104.78 mm^2
Standard Deviation 21.22

SECONDARY outcome

Timeframe: 10 yrs post-operative

Outcome measures

Outcome measures
Measure
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
Condyle Contact Stress at Maximum Flexion During Deep Knee Bend Activity
Average Medial Condyle Contact Stress
10.66 MPa
Standard Deviation 1.75
Condyle Contact Stress at Maximum Flexion During Deep Knee Bend Activity
Average Lateral Condyle Contact Stress
10.05 MPa
Standard Deviation 1.45

Adverse Events

Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly

Serious events: 0 serious events
Other events: 0 other events
Deaths: 0 deaths

Serious adverse events

Adverse event data not reported

Other adverse events

Adverse event data not reported