Trial Outcomes & Findings for Mechanics Analysis of Patients Having a Sigma Mobile Bearing TKA (NCT NCT01714492)
NCT ID: NCT01714492
Last Updated: 2019-10-10
Results Overview
The values that were reported indicate the motion of the contact point from full extension to patient's maximum flexion. Throughout flexion, if the point translated forward (anteriorly) atop the tibial tray, the number was reported as positive. If the point traveled backwards (posteriorly) atop the tibial tray, the number was reported as negative.
COMPLETED
8 participants
10 yrs post-operative
2019-10-10
Participant Flow
hospital, medical clinic
Participant milestones
| Measure |
Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Overall Study
STARTED
|
8
|
|
Overall Study
COMPLETED
|
8
|
|
Overall Study
NOT COMPLETED
|
0
|
Reasons for withdrawal
Withdrawal data not reported
Baseline Characteristics
Mechanics Analysis of Patients Having a Sigma Mobile Bearing TKA
Baseline characteristics by cohort
| Measure |
Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly
n=8 Participants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Age, Categorical
<=18 years
|
0 Participants
n=5 Participants
|
|
Age, Categorical
Between 18 and 65 years
|
1 Participants
n=5 Participants
|
|
Age, Categorical
>=65 years
|
7 Participants
n=5 Participants
|
|
Age, Continuous
|
74 years
STANDARD_DEVIATION 8.6 • n=5 Participants
|
|
Sex: Female, Male
Female
|
6 Participants
n=5 Participants
|
|
Sex: Female, Male
Male
|
2 Participants
n=5 Participants
|
|
Region of Enrollment
United States
|
8 participants
n=5 Participants
|
PRIMARY outcome
Timeframe: 10 yrs post-operativePopulation: Subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
The values that were reported indicate the motion of the contact point from full extension to patient's maximum flexion. Throughout flexion, if the point translated forward (anteriorly) atop the tibial tray, the number was reported as positive. If the point traveled backwards (posteriorly) atop the tibial tray, the number was reported as negative.
Outcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
In Vivo Linear Knee Kinematics From Fluoroscopy Evaluation During Deep Knee Bend Activity
Lateral Anterior Posterior Deep Knee Bend
|
-4.1 mm
Standard Deviation 2.4
|
|
In Vivo Linear Knee Kinematics From Fluoroscopy Evaluation During Deep Knee Bend Activity
Medial Anterior Posterior Deep Knee Bend
|
1.2 mm
Standard Deviation 2.8
|
PRIMARY outcome
Timeframe: 10 yrs post-operativeOutcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Femoral Axial Rotation With Respect to the Tibia During Deep Knee Bend Activity
|
6.1 degrees
Standard Deviation 5.8
|
PRIMARY outcome
Timeframe: 10 yrs post-operativeOutcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Range of Motion During Flexion of Deep Knee Bend Activity
|
103 degrees
Standard Deviation 11.2
|
SECONDARY outcome
Timeframe: 10 yrs post-operativePopulation: Subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation
The intended unit of measure is "times Body Weight" (or xBW), relating to a ratio.
Outcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
In Vivo Knee Force Values From Fluoroscopy Evaluation During Deep Knee Bend Activity
Medial Condyle Knee Force at Maximum Flexion
|
1.57 times body weight
Standard Deviation 0.53
|
|
In Vivo Knee Force Values From Fluoroscopy Evaluation During Deep Knee Bend Activity
Lateral Condyle Knee Force at Maximum Flexion
|
1.39 times body weight
Standard Deviation 0.36
|
SECONDARY outcome
Timeframe: 10 yrs post-operativeOutcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Condyle Contact Area at Maximum Flexion During Deep Knee Bend Activity
Medial
|
109.16 mm^2
Standard Deviation 2.82
|
|
Condyle Contact Area at Maximum Flexion During Deep Knee Bend Activity
Lateral
|
104.78 mm^2
Standard Deviation 21.22
|
SECONDARY outcome
Timeframe: 10 yrs post-operativeOutcome measures
| Measure |
In Vivo Knee Kinematics From Fluoroscopy Evaluation During Dee
n=8 Implants
subjects implanted with the Sigma Posterior Stabilizing Rotating Platform TKA including a polyethylene insert with 4 beads that allow for determination of polyethylene rotation.
|
|---|---|
|
Condyle Contact Stress at Maximum Flexion During Deep Knee Bend Activity
Average Medial Condyle Contact Stress
|
10.66 MPa
Standard Deviation 1.75
|
|
Condyle Contact Stress at Maximum Flexion During Deep Knee Bend Activity
Average Lateral Condyle Contact Stress
|
10.05 MPa
Standard Deviation 1.45
|
Adverse Events
Sigma Posterior Stabilizing Rotating Platform TKA Beaded Poly
Serious adverse events
Adverse event data not reported
Other adverse events
Adverse event data not reported
Additional Information
Results disclosure agreements
- Principal investigator is a sponsor employee
- Publication restrictions are in place